iQ200 Series Urine Microscopy Analyzer- All part numbers recallZ-1945-2020 · Beckman Coulter Inc. · initiated Mar 30, 2020
Class II Terminated All recalls posted in May 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 03/25/2020 Beckman Coulter Inc. sent out Urgent Medical Device Alert identifying the problem, the circumstances in which it can occur, the actions that should be taken by the customer/user in order to prevent risks to patients or users, and the actions planned by Beckman Coulter to correct the problem. PRODUCTS: Products installed between July of 2012 and December of 2014 are in scope of this field corrective action and can be identified by the name of the Product (all part numbers) that were installed in the time frames noted. ACTION TO BE TAKEN BY CUSTOMER/USER: If flames and or smoke be detected within the computer the customers/users should; follow laboratorys safety guidelines, contact Beckman Coulter by phone: call 800-526-7694 in the United States and Canada or if outside the United States and Canada, contact Beckman Coulter hotline or the local Beckman Coulter representative, and Follow Standard Operating Procedures (SOP) for Instrument backup testing methodology. This information should be communicated to all laboratory staff and retained as part of lab quality system documentation. Customers are asked to complete the Notification Response Form either electronically or manually depending on the method the notice was received. ACTIONS PLANNED BY BECKMAN COULTER: A Beckman Coulter service provider will proactively schedule a service visit to inspect the PC6 computer and replace the affected cable, if necessary. Additional potentially affected customers will be notified of the issue via Updated Urgent Medical Device Recall letter that will be sent during the week of 10/26/2020 via mail and/or e-mail. Note: Customers that received the initial letter but have any of the additional potentially affected devices will be re-notified of the issue.
Recall details
- Recall number
- Z-1945-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 30, 2020
- Posted by the FDA
- May 9, 2020
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 3576 units
- Product code
- KQO
- FDA submissions
- K022774
- Distribution
- US Distribution: Nationwide Foreign Distribution: Worldwide
Product description
iQ200 Series Urine Microscopy Analyzer- All part numbers
Code information (models, lots, serial numbers)
Catalog Numbers: 00-3322, 700-3325, 700-3326, 700-3345, 700-3347, 700-3370, 700-3375, 700-3300RFB, B48999, C10683, C10684., All part numbers
Other devices under the same FDA submission
4 devices cleared under K022774 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- iQ200 Elite Plus (US) (2008 Version) (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls2 reports
- iQ200 Sprint Pro (NA)/SPRINT w load/unload (NA)/SPRINT Pro w load/unload (NA)/Sprint PBeckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls1 report
- iQ200 Select/Elite Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls0 reports
- iQ200 Select Plus Kit 2008 Analyzer (NA)/ Select Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System2 recalls4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page