Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I… recallZ-2263-2020 · Endologix, Inc. · initiated May 6, 2020
Class I Terminated All recalls posted in June 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Endologix, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
FDA root cause category: Process control
What the company is doing
On 05/06/20, "Urgent Medical Device Correction notices" were distributed to customers via person/mail/email. These notices reaffirmed treatment recommendations for patients provided in the previously distributed Urgent Field Safety Notice, dated 09/26/2018 that were mailed on 09/27/18. Safety Update: Treatment of a Patient with Polymer Leak - Patient Reaction During the polymer injection step of the procedure, systemic hypotension may indicate that a polymer leak is occurring. Blood pressure monitoring during polymer fill may assist in early identification of a potential polymer leak. In the absence of other obvious diagnoses causing sudden hypotension during polymer fill, firm recommends that a hypersensitivity reaction (an anaphylactoid response) to intravascular polymer leak be considered a probable diagnosis. Patients with a polymer leak should undergo immediate treatment for a potential severe hypersensitivity response in accordance with institutional protocols (e.g., intravascular fluids, antihistamines, corticosteroids, epinephrine). In addition to systemic hypotension, device related findings that are indicative of a polymer leak include rapid emptying of the fill polymer syringe, and incomplete filling of the polymer channels. Safety Update: Treatment of a Patient with Polymer Leak - Aneurysm Management Aneurysm related complications that may occur due to polymer leak should be treated with standard endovascular techniques at the physician's discretion, utilizing the ancillary equipment listed in the Instructions for Use (IFU), or an open surgical approach. The specific treatment will be dependent on the extent and location of incomplete filling of the polymer rings and the associated clinical findings. In respect of intra-operative Type 1a endoleaks resulting from polymer leak, there were two main treatment strategies: (1) conservative management (in the cases of small endoleaks expected to resolve spontaneously) or (2) the use of balloon expand
Recall details
- Recall number
- Z-2263-2020
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- May 6, 2020
- Posted by the FDA
- Jun 11, 2020
- Recalling firm
- Endologix, Inc.
- Quantity
- 9156
- Product code
- MIH
- FDA submissions
- P120006
- Distribution
- Worldwide DIstribution: U.S (nationwide).: PA, AZ, NY, FL, HI, CA, OR, IL, MS, AR, WI, KY, TX, TN, MO, NJ, MA, MT, VA, MI, CO, WV, NC, NH, NV, CT, ME, WY, OH, OK, IN, UT, AL, WA, GA, SC, KS, LA, NM, MD, DC, NE, SD, RI, IA, VT, ND. and Countries of:: Argentina, Australia, Chile, Hong Kong, Japan, Malaysia, Mexico, New Zealand, Taiwan, Thailand, Canada, Austria, Belgium, Cyprus, Czech Republic, Germany, Hungary, Ireland, Italy, Switzerland, Luxembourg, Netherlands, Poland, Portugal, Romania, Spain, Sweden, and United Kingdom
Product description
Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
Code information (models, lots, serial numbers)
All Lot and Serial Numbers not yet implanted
Other devices under the same FDA submission
42 devices cleared under P120006 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment11 deaths105 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths87 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths72 reports
- Alto™ Abdominal Stent Graft System TV-AB2980-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death65 reports
- Alto™ Abdominal Stent Graft System TV-AB2680-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment5 deaths56 reports
- Alto™ Abdominal Stent Graft System TV-AB3480-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment6 deaths56 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths54 reports
- Alto™ Abdominal Stent Graft System TV-AB2380-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths48 reports
- Alto™ Abdominal Stent Graft System TV-AB2080-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment20 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment20 reports
- Alto™ Abdominal Stent Graft System TV-AB2680-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment8 reports
- Alto™ Abdominal Stent Graft System TV-AB3480-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death8 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page