Medical device recalls: June 2020
The FDA posted 57 medical device recall records in June 2020, from 21 recall events by 19 companies. 11 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Healthcare Diagnostics, Inc. · 2 events, 8 records
- GE Healthcare, LLC · 2 events, 3 records
- Stryker GmbH · 1 event, 15 records
- Arrow International Inc · 1 event, 10 records, 10 Class I
- Beckman Coulter Inc. · 1 event, 6 records
- Cook Inc. · 1 event, 2 records
- Abbott Laboratories · 1 event, 1 record
- Codan US Corporation · 1 event, 1 record
FDA root cause
- Process control · 19
- Component design/selection · 11
- Device Design · 10
- Other · 8
- Software Design Change · 2
- Employee error · 2
- Under Investigation by firm · 1
- Software design · 1
Class I recalls (11)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2408-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2409-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2410-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2411-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2412-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2413-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2414-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2415-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA, Product Code IAP-0700 ↗Class IZ-2416-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701 ↗Class IZ-2417-2020 · Arrow International Inc · posted Jun 26, 2020 · Terminated
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector…
- Z-2263-2020 · Endologix, Inc. · posted Jun 11, 2020 · Terminated
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic…
Class II recalls (46)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2433-2020 · Codan US Corporation · posted Jun 26, 2020 · Terminated
Individual packages of connectors may have incomplete or bad seals which would compromise sterility.
- Z-2424-2020 · GE Healthcare, LLC · posted Jun 24, 2020 · Open, Classified
The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform…
- Z-2418-2020 · Hocoma AG · posted Jun 23, 2020 · Open, Classified
The motor controller may fail, which can potentially lead to an error in functionality of the Body Weights Support Rope, which means that the rope can go up or down in an uncontrolled way.
- Z-2404-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 19, 2020 · Open, Classified
Insufficient sample may not be detected when using Tube Top Sample Cup (TTSC) or V-shaped Sample Cup sample containers with the Atellica IM1300 and IM1600 analyzers.
- Z-2405-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 19, 2020 · Open, Classified
Insufficient sample may not be detected when using Tube Top Sample Cup (TTSC) or V-shaped Sample Cup sample containers with the Atellica IM1300 and Atellica IM1600 analyzers.
- Z-2373-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2374-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2375-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2376-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2377-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2378-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2379-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2380-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2381-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2382-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2383-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2384-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2385-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2386-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2387-2020 · Stryker GmbH · posted Jun 16, 2020 · Terminated
Seal integrity of the blister pack may be compromised and sterility cannot be assured
- Z-2388-2020 · Siemens Medical Solutions USA, Inc · posted Jun 16, 2020 · Terminated
If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to difficulties in…
- HardyCHROM MRSA,REFClass IIZ-2298-2020 · Hardy Diagnostics · posted Jun 9, 2020 · Terminated
False Positive results due to the MSSA (methicillin-susceptible) strain Staphylococcus aureus (ATCC 29213) breakthrough when this strain should have been inhibited.
- Z-2330-2020 · Medtronic Navigation, Inc. · posted Jun 9, 2020 · Open, Classified
There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]
- Z-2297-2020 · Elekta Limited · posted Jun 8, 2020 · Terminated
In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure.
- Z-2280-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 5, 2020 · Terminated
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Z-2281-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 5, 2020 · Terminated
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Z-2282-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 5, 2020 · Terminated
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Z-2283-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 5, 2020 · Terminated
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Z-2284-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 5, 2020 · Terminated
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Z-2285-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jun 5, 2020 · Terminated
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Z-2289-2020 · Elekta, Inc. · posted Jun 5, 2020 · Terminated
potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.
- Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001 ↗Class IIZ-2206-2020 · Medtronic Inc. · posted Jun 3, 2020 · Terminated
Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock,…
- Z-2251-2020 · Abbott Laboratories · posted Jun 3, 2020 · Terminated
In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.
- Z-2258-2020 · GE Healthcare, LLC · posted Jun 3, 2020 · Open, Classified
GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used;…
- Z-2259-2020 · GE Healthcare, LLC · posted Jun 3, 2020 · Open, Classified
GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used;…
- da Vinci SP surgical system ↗Class IIZ-2260-2020 · Intuitive Surgical, Inc. · posted Jun 3, 2020 · Terminated
Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions:…
- Z-2203-2020 · Cook Inc. · posted Jun 2, 2020 · Terminated
The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occlusion.
- Z-2204-2020 · Cook Inc. · posted Jun 2, 2020 · Terminated
The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occlusion.
- Z-2237-2020 · Beckman Coulter Inc. · posted Jun 2, 2020 · Terminated
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and…
- Z-2238-2020 · Beckman Coulter Inc. · posted Jun 2, 2020 · Terminated
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and…
- Z-2239-2020 · Beckman Coulter Inc. · posted Jun 2, 2020 · Terminated
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and…
- Z-2240-2020 · Beckman Coulter Inc. · posted Jun 2, 2020 · Terminated
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860,…
- Z-2241-2020 · Beckman Coulter Inc. · posted Jun 2, 2020 · Terminated
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and…
- Z-2242-2020 · Beckman Coulter Inc. · posted Jun 2, 2020 · Terminated
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and…
- Z-2243-2020 · Med Tec Inc · posted Jun 2, 2020 · Terminated
Potential for disengagement of the Varian Style Type-S Extension during patient repositioning. The risk associated with the disengagement of the extension could be a patient fall if the patient is not supported by clinical staff.
- Z-2113-2020 · Stryker Sustainability Solutions · posted Jun 1, 2020 · Terminated
A subset of temporary intracardiac ultrasound catheters are labeled with the incorrect expiration date. They are mislabeled as having a shelf life of two years, while the actual shelf life is one year. The earliest affected product will expire in May 2020.…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.