RINGLOC BI-POLAR HIP SYSTEM recallZ-2145-2020 · Biomet, Inc. · initiated Apr 17, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential presence of elevated endotoxin levels that exceed the specification limit
FDA root cause category: Process control
What the company is doing
Zimmer Biomet issued Urgent Medical Device recall letter dated April 17, 2002 to distributors with direct sales and/or field inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory, as well as hospitals that have consumed products will be notified via FedEx. Surgeons that have consumed products will be notified via FedEx. Action to take: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. This form must be returned even if you do not have affected products available to return in your territory. b. For international returns, request an IRA by emailing [email protected] c. For each return, send a copy of Attachment 1 to [email protected]. CF04107 Rev. 4, Eff. Date: 16 Jul 2019 Ref. CP04102 Field Action Activities d. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. e. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or i
Recall details
- Recall number
- Z-2145-2020
- Status
- Terminated
- Initiated by the company
- Apr 17, 2020
- Recalling firm
- Biomet, Inc.
- Quantity
- 58 units
- Product code
- LZO
- FDA submissions
- K051411
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, AHMADABAD, GUJARAT, BANGKOK, BELROSE NSW, BUENOS AIRES, CHILE, COSTA RICA, HIRATSUKA CITY, HUECHURABA SANTIAGO, INDIA, JAPAN, MEXICO DF, NETHERLANDS, PICHINCHA, EC, SAN JOSE, SINGAPORE, TAIWAN, THAILAND, YONGSAN-GU SEOUL.
Product description
Hip Products Item Number: 1) 110010462 RingLoc Hip System, Acetabular Bi-Polar Cup, 28 MM X 51 MM 2) 11-165218 RingLoc Bi-Polar Hip System, Acetabular Cup, 28 MM X 47 MM 3) XL-105916 588500 RingLoc Hip System, Acetabular Liner, 36 MM, Size 26 Product Usage: Hemi hip arthroplasty.
Code information (models, lots, serial numbers)
Lot Numbers: UDI Number 1) 752470 (01)00880304568785(17)220118(10)752470 854070 (01)00880304568785(17)220119(10)854070; 2) 094360 (01)00880304209688(17)220106(10)094360; 3) 588500 (01)00880304209688(17)220106(10)094360
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RINGLOC BI-POLAR HIP SYSTEM 11-165218Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 recall2 deaths9 reports
Product lines
Other devices under the same FDA submission
12 devices cleared under K051411 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BIOLOX® Delta Hip System 12-115109Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- RingLoc® Hip System XL-105924Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 report
- RingLoc® Hip System XL-105833Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 report
- BIOLOX DELTA HIP SYSTEM 12-115116Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- BIOLOX DELTA HIP SYSTEM 12-115115Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- RingLoc® Hip System XL-105916Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- RingLoc® Hip System XL-105917Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- BIOLOX® Delta Hip System 12-115110Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- BIOLOX® Delta Hip System 12-115111Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- BIOLOX® Delta Hip System 12-115114Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- BIOLOX® Delta Hip System 12-115116Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- BIOLOX® Delta Hip System 12-115117Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 85524).
- Z-2144-2020 · TerminatedExtremities Item Number: 1)113954 Hybrid Glenoid Glenoid Base, 4 MM 2) XL-115366 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 41 MM, Standard 3) XL-115364 Comp…
- Z-2146-2020 · TerminatedKnee Products: 141356 Regenerex Series-A Patella 3 Peg, 31 MM 141358 Regenerex Series-A Patella 3 Peg, 37 MM Product Usage: Knee prosthesis
- Z-2147-2020 · TerminatedKnee Products: 150414 Orthopedic Salvage System (OSS) Tibial Bearing, 20 MM Standard - Product Usage: Knee prosthesis
- Z-2148-2020 · TerminatedKnee Products: 154335 Oxford Partial Knee System, Fixed Lateral Tibial Construct, Left, Cemented, D3 154336 Oxford Partial Knee System, Fixed Lateral Tibial Construct, Left, Cement…
- Z-2149-2020 · TerminatedKnee Products: 155308 Knees AGC Knee System PS Molded Tibial Component, 10 MM X 65 MM 155326 Knees AGC Knee System PS Molded Tibial Component, 10 MM X 70 MM 155330 Knees AGC Knee S…
- Z-2150-2020 · TerminatedKnee Products: 183622 Vanguard Knee System PS Tibial Bearing, 12 MM X 63/67 MM 183620 Vanguard Knee System PS Tibial Bearing, 10 MM X 63/67 MM 189048 Vanguard Knee System, AS Tibia…
- Z-2151-2020 · TerminatedKnee Products: 189260 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 71 MM 189720 Vanguard Knee System, CR Mono Lock Tibial Bearing, 10 MM X 83 MM 189320 Vanguard Kne…
- Z-2152-2020 · TerminatedKnee Products: 189420 Vanguard Knee System, PS Mono Lock Tibial Bearing, 10 MM X 63 MM 189422 Vanguard Knee System, PS Mono Lock Tibial Bearing, 12 MM X 63 MM 189442 Vanguard Knee …
- Z-2153-2020 · TerminatedKnee Products: 183742 Vanguard Knee System, PS+ Tibial Bearing, 12 MM, 71/75 MM EP-183608 Vanguard Knee System, PS Tibial Bearing, E1 Infused, 18 MM X 59 MM EP-183608 Vang…
- Z-2154-2020 · TerminatedKnee Products: 184762 Knees Vanguard Knee System, Series-A Standard Patella, 28 MM; 184764 Knees Vanguard Knee System, Series-A Standard Patella, 31 MM Product Usage: Knee prosth…
- Z-2155-2020 · TerminatedKnee Products: US154709 Knees Vanguard M Partial Knee System MonoBlock Tibial Tray, Right Medial/Left Lateral, B8 US154722 Knees Vanguard M Partial Knee System MonoBlock Tibial Tra…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page