Skip to content

Vanguard Knee System recallZ-2154-2020 · Biomet, Inc. · initiated Apr 17, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential presence of elevated endotoxin levels that exceed the specification limit

FDA root cause category: Process control

What the company is doing

Zimmer Biomet issued Urgent Medical Device recall letter dated April 17, 2002 to distributors with direct sales and/or field inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory, as well as hospitals that have consumed products will be notified via FedEx. Surgeons that have consumed products will be notified via FedEx. Action to take: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. This form must be returned even if you do not have affected products available to return in your territory. b. For international returns, request an IRA by emailing [email protected] c. For each return, send a copy of Attachment 1 to [email protected]. CF04107 Rev. 4, Eff. Date: 16 Jul 2019 Ref. CP04102 Field Action Activities d. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. e. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or i

Recall details

Recall number
Z-2154-2020
Status
Terminated
Initiated by the company
Apr 17, 2020
Recalling firm
Biomet, Inc.
Quantity
91 units
Product code
JWH
FDA submissions
K040770
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, AHMADABAD, GUJARAT, BANGKOK, BELROSE NSW, BUENOS AIRES, CHILE, COSTA RICA, HIRATSUKA CITY, HUECHURABA SANTIAGO, INDIA, JAPAN, MEXICO DF, NETHERLANDS, PICHINCHA, EC, SAN JOSE, SINGAPORE, TAIWAN, THAILAND, YONGSAN-GU SEOUL.

Product description

Knee Products: 184762 Knees Vanguard Knee System, Series-A Standard Patella, 28 MM; 184764 Knees Vanguard Knee System, Series-A Standard Patella, 31 MM Product Usage: Knee prosthesis

Code information (models, lots, serial numbers)

Lot Numbers: UDI Number 508390 Vanguard Knee System, Series-A Standard Patella, 28 MM (01)00880304431638(17)211218(10)508390; 309810 Vanguard Knee System, Series-A Standard Patella, 31 MM (01)00880304431317(17)211228(10)309810

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

15 devices cleared under K040770 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85524).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page