Skip to content

Giraffe Shuttle. Upgrade Kits with part number 2109672-001… recallZ-2111-2020 · GE Healthcare, LLC · initiated Apr 2, 2020

Class II Terminated All recalls posted in May 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.

FDA root cause category: Device Design

What the company is doing

GE Healthcare notified customers on 04/02/2020 via "Urgent Medical Device Correction" letter. The recall letter identified the affected product and serial numbers. The firm instructed the users to discontinue use of the product if at any time they smell a gas odor. GE Healthcare will correct all affected products at a later date when the correction is available at no cost.

Recall details

Recall number
Z-2111-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 2, 2020
Posted by the FDA
May 22, 2020
Recalling firm
GE Healthcare, LLC
Quantity
204 units.
Product code
FMZ
FDA submissions
K101788
Distribution
US Nationwide distribution in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, LA, MD, MI, MS, NC, ND, NM, NV, NY, OH, Ok, OR, PA, SC, TN, TX, VA, WA, WI and WV.

Product description

Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.

Code information (models, lots, serial numbers)

Serial numbers: GSHW60086, GSHX60001, GSHX60014, GSHY70002-GSHY70195, GSHZ70101-GSHZ70108, MAAP50034, MAAQ50170, MAAR50021, MAAT50013, MAAT60045, MAAU61029, MAAU61040 and MAAV6059.

Other devices under the same FDA submission

1 device cleared under K101788 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page