Vital recallZ-2214-2020 · Zimmer Biomet, Inc. · initiated May 5, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
FDA root cause category: Under Investigation by firm
What the company is doing
URGENT MEDICAL DEVICE RECALL notification letters dated 5/5/20 were sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 800-447-3625 between 8:00 am and 6:00pm MST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. Patients should continue with the post-operative care as scheduled by their surgeon who typically assess the construct as part of their evaluation(s). During these visits, surgeons should compare follow up radiographs to the immediate post-op radiographs to determine if the Closure Tops have loosened. Additionally, consider doing dynamic imaging such as flexion/extension to detect implant movement when appropriate. If Closure Top loosening and/or migration is noted, surgeon may consider surgical intervention. During this surgical intervention, consideration should be made regarding replacement of the closure tops as well as the screws. 3. Complete Attachment 1 Certificate of Ackn
Recall details
- Recall number
- Z-2214-2020
- Status
- Terminated
- Initiated by the company
- May 5, 2020
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 3235 total
- Product code
- NKB
- FDA submissions
- K183550
- Distribution
- US nationwide distribution.
Product description
Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X35MM, Item Number 810M5535 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.
Code information (models, lots, serial numbers)
UDI: (01)00887868324420; Lot Numbers: C19H0007 C19J0004 C19K0001 C19K0003 C20A0002 SBM108403 SBM118985 SBM122395
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vital™ 810M5535Biomet Spine LLC · Thoracolumbosacral pedicle screw system1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 85613).
- Z-2210-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 4.5X30MM, Item Number 810M4530 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2211-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 4.5X35MM, Item Number 810M4535 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2212-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 4.5X40MM, Item Number 810M4540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2213-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 4.5X45MM, Item Number 810M4545 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2215-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X40MM, Item Number 810M5540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2216-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X45MM, Item Number 810M5545 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2217-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X50MM, Item Number 810M5550 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2218-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X55MM, Item Number 810M5555 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2219-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X35MM, Item Number 810M6535 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2220-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X45MM, Item Number 810M6545 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2221-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X50MM, Item Number 810M6550 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2222-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X55MM, Item Number 810M6555 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2223-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 7.5X35MM, Item Number 810M7535 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2224-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X40MM, Item Number 810M6540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2225-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 7.5X45MM, Item Number 810M7545 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2226-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 7.5X50MM, Item Number 810M7550 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2227-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 7.5X55MM, Item Number 810M7555 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
- Z-2228-2020 · TerminatedVital MIS Spinal Fixation system implants CANN BL POLY EXT 8.5X40MM, Item Number 811M8540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle s…
- Z-2229-2020 · TerminatedVital MIS Spinal Fixation system implants CANN BL POLY EXT 8.5X45MM, Item Number 811M8545 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle s…
- Z-2230-2020 · TerminatedVital MIS Spinal Fixation system implants CANN BL POLY EXT 8.5X50MM, Item Number 811M8550 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle s…
- Z-2231-2020 · TerminatedVital MIS Spinal Fixation system implants CANN BL POLY EXT 8.5X55MM, Item Number 811M8555 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle s…
- Z-2232-2020 · TerminatedVital MIS Spinal Fixation system implants CANN BL POLY EXT 8.5X60MM, Item Number 811M8560 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle s…
- Z-2233-2020 · TerminatedVital MIS Spinal Fixation system implants VITAL MIS TAB BREAKER, Item Number 850M0055 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw…
- Z-2234-2020 · TerminatedVital MIS Spinal Fixation system implants CANN POLY EXT TAB 7.5X40MM, Item Number 810M7540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page