BioPlex 2200 recallZ-2982-2020 · Bio-Rad Laboratories, Inc. · initiated Apr 21, 2020
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bio-Rad Laboratories, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The increased level of RPR reactivity in the BioPlex 2200 Syphilis Total & RPR kit material number 12000650.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
Consignees that have purchased active lots of the BioPlex 2200 Syphilis Total and RPR kit have been notified of the Field Correction via a customer letter sent via Fed Ex overnight ; date issued 04/21/2020. The communication was to inform that the RPR portion of the BioPlex 2200 Syphilis Total & RPR assay will be temporarily disabled. The RPR assay will be disabled beginning with reagent lot 301142 of the BioPlex 2200 Syphilis Total & RPR assay (Catalog No. 12000650, UDI 03610520653012) and calibrator lot 52216 (Catalog No. 12000651, UDI 03610520652770). The reporting of the RPR assay results will be disabled by the installation of a new assay protocol file (APF). The communication includes that there are no modifications to the chemistry of the assay reagents or processing of the assay by the BioPlex 2200 System. The RPR assay will remain disabled until further notice. Consignees are asked to complete and return the Customer Response form (fax: 1-888-228-0688/email: [email protected]) to confirm receipt of the communication and to receive instructions for installation of the new APF; return the form within 10 days of receiving the notice. Additionally, consignees are asked to contact their local technical support if there are any questions about the notification.
Recall details
- Recall number
- Z-2982-2020
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2020
- Recalling firm
- Bio-Rad Laboratories, Inc.
- Quantity
- 12000 kits
- Product code
- LIP
- FDA submissions
- K170413
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL,GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TX, UT, VA, WA, WI, and WV. The countries of Canada (CA), Switzerland (CH), France (FR), Italy (IT), Spain (ES), and Romania (RO), and United Kingdom (GB).
Product description
BioPlex 2200 Syphilis Total & RPR Linked Immunoabsorption Assay, Treponema pallidum:UDI:03610520653012 - Product Usage: intended for the qualitative detection of total (IgG/IgM) antibodies to Treponema pallidum and the qualitative detection and/or titer determination of non-treponemal reagin antibodies in human serum or plasma.
Code information (models, lots, serial numbers)
The RPR assay will be disabled beginning with reagent lot 301142 and future manufactured lots of the BioPlex 2200 Syphilis Total & RPR assay (Catalog No. 12000650, UDI 03610520653012) and calibrator lot 52216 (Catalog No. 12000651, UDI 03610520652770). Product Code 1: ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM (LIP), Product Code 2: ANTIGENS, NONTREPONEMAL, ALL (GMQ), Product Code 3: CALIBRATOR, SECONDARY (JIT), Product Code 4:SINGLE (SPECIFIED) ANALYTE CONTROLS (ASSAYED AND UNASSAYED) (JJX).
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- BioPlex 2200 12000650Bio-Rad Laboratories, Inc. · ANTIGENS, NONTREPONEMAL, ALL3 recalls1 report
- BioPlex 2200 12000651Bio-Rad Laboratories, Inc. · Calibrator, secondary3 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.