BioPlex 2200Recalls & reported problems across all models

Bio-Rad Laboratories, Inc. · 8 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

9 open recalls name at least one BioPlex 2200 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls
9

We found 8 models of BioPlex 2200 by Bio-Rad Laboratories, Inc. in the FDA's device registry (GUDID). Together they are named in 8 adverse event reports in the FDA's MAUDE database, received between Dec 13, 2016 and Oct 21, 2025 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (3 reports).

9 FDA recall records name at least one of these models by its device identifier, 9 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

BioPlex 2200 recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 28, 2023Open, Classified

    APLS IgM reagent kits were packaged with the incorrect conjugate, which could lead to an increase in false-positive and false-negative results.

    Z-1572-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 21, 2023Open, Classified

    Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.

    Z-1378-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Feb 21, 2023Open, Classified

    Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.

    Z-1377-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Apr 14, 2022Open, Classified

    Performance concerns for the CMV IgM analyte. Analyte has demonstrated variable elevated background depending on the handling of the reagent pack.

    Z-1096-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Feb 8, 2022Open, Classified

    Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet stability specifications

    Z-0807-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. Feb 8, 2022Open, Classified

    Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet stability specifications

    Z-0806-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  7. Sep 17, 2021Open, Classified

    Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

    Z-0236-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  8. Sep 17, 2021Open, Classified

    Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

    Z-0235-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  9. Apr 21, 2020Open, Classified

    The increased level of RPR reactivity in the BioPlex 2200 Syphilis Total & RPR kit material number 12000650.

    Z-2982-2020 · Root cause: Unknown/Undetermined by firm · names 2 of these models · Recall details · FDA record ↗

Most reported BioPlex 2200 problems

Device problems

  • False Positive Result3
  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • False Negative Result1
  • Human Factors Issue1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Exposure to Body Fluids1
  • Injury1
  • No Consequences Or Impact To Patient1

BioPlex 2200 models

8 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent BioPlex 2200 adverse event reports

The 8 most recent reports across all models, in the reporter's own words as published by the FDA.