Baxter SIGMA Spectrum Infusion Pump with Master Drug… recallZ-0122-2021 · Baxter Healthcare Corporation · initiated Jun 30, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Baxter is updating the device IFU for safety communication regarding cleaning practices.
FDA root cause category: Labeling design
What the company is doing
A Safety Alert communication was sent to affected customers on 04/01/2020 via USPS First Class Mail. A follow-up Urgent Device Correction communication was sent to all affected customers on 8/28/2020 via USPS first class mail. The customer communications reiterate instructions for use listed in the Operator's Manual. Baxter is asking customers to: 1) Operators may continue to safely use the infusion pumps while following the instructions for cleaning provided in the Operator's Manual. An electronic copy of the Operator's Manual can be accessed at https://service.baxter.com. In addition, a full list of approved cleaning agents can be accessed at www.spectrumIQ.com/resources.html. 2) Inspect all pumps at their facility to assess the electrical pins on the pump rear case and the battery electrical contacts for residue buildup or corrosion. If corrosion or residue buildup is identified, please contact Baxter to service the device. 3) Baxter will be updating the Instructions for Use to recommend a routine inspection process for the electrical pins on the pump rear case and the battery electrical contacts. Once the updated Instructions for Use are made available, Baxter will issue a written notification to inform customers of the availability of the updated IFU. The updated IFU may be accessed online at the Baxter Global Technical E Service Center at URL: https://service.baxter.com. 4) Clinicians should ensure backup devices are readily available when infusing critical medications where interruptions could cause serious injury or death. 5) To maintain the battery properly, please ensure the pumps are plugged into AC power when possible to prevent battery depletion. 6) Complete the enclosed customer reply form and return it to Baxter by email. On 9/22/2021, the firm issued an Urgent Device Correction letter dated 9/22/2021 in follow-up to their communication of the 4/1/2020 safety alert and 8/28/2020 Urgent Device Correction. This was issued via first class USPS ma
Recall details
- Recall number
- Z-0122-2021
- Status
- Terminated
- Initiated by the company
- Jun 30, 2020
- Posted by the FDA
- Oct 22, 2020
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 152,539 units
- Product code
- FRN
- FDA submissions
- K133801
- Distribution
- Distribution throughout the United States, St. Thomas USVI, and Puerto Rico. Foreign distribution including Canada, Caribbean (Barbados, Trinidad and Tobago, Jamaica, Bahamas, Bermuda, and Guyana)
Product description
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8)
Code information (models, lots, serial numbers)
Product Code: 35700BAX2, All Serial Numbers; GTIN 00085412498683
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SIGMA Spectrum Infusion Pump 35700BAX2Baxter Healthcare Corporation · Infusion safety management software9 recalls12 deaths11,585 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 86062).
- Z-0120-2021 · TerminatedBaxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6)
- Z-0121-2021 · TerminatedBaxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8)
- Z-0123-2021 · TerminatedSpectrum IQ Infusion System with Dose IQ Safety Software
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page