Medical device recalls: October 2020
The FDA posted 21 medical device recall records in October 2020, from 10 recall events by 9 companies. 12 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Healthcare Diagnostics, Inc. · 2 events, 3 records
- Baxter Healthcare Corporation · 1 event, 4 records, 4 Class I
- Vascular Solutions, Inc. · 1 event, 4 records, 4 Class I
- Medtronic Vascular · 1 event, 3 records, 3 Class I
- Pentax of America Inc · 1 event, 2 records
- RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) · 1 event, 2 records
- Biomet, Inc. · 1 event, 1 record
- Encore Medical, LP · 1 event, 1 record
FDA root cause
- Device Design · 7
- Process control · 5
- Labeling design · 4
- Under Investigation by firm · 2
- Software design · 2
- Process change control · 1
Class I recalls (12)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0120-2021 · Baxter Healthcare Corporation · posted Oct 22, 2020 · Terminated
Baxter is updating the device IFU for safety communication regarding cleaning practices.
- Z-0121-2021 · Baxter Healthcare Corporation · posted Oct 22, 2020 · Terminated
Baxter is updating the device IFU for safety communication regarding cleaning practices.
- Z-0122-2021 · Baxter Healthcare Corporation · posted Oct 22, 2020 · Terminated
Baxter is updating the device IFU for safety communication regarding cleaning practices.
- Z-0123-2021 · Baxter Healthcare Corporation · posted Oct 22, 2020 · Terminated
Baxter is updating the device IFU for safety communication regarding cleaning practices.
- Z-0114-2021 · Stryker Neurovascular · posted Oct 21, 2020 · Terminated
Increase in fracture complaints of the flexible, tapered core wire, resulting in stent retriever separation from the core wire during use. The stent retriever can be left behind in the vasculature.
- Langston Dual Lumen CatheterClass IZ-0110-2021 · Vascular Solutions, Inc. · posted Oct 20, 2020 · Terminated
The inner lumen of the Langston catheter may separate from the hub during pressure injection.
- Z-0111-2021 · Vascular Solutions, Inc. · posted Oct 20, 2020 · Terminated
The inner lumen of the Langston catheter may separate from the hub during pressure injection.
- Z-0112-2021 · Vascular Solutions, Inc. · posted Oct 20, 2020 · Terminated
The inner lumen of the Langston catheter may separate from the hub during pressure injection.
- Z-0113-2021 · Vascular Solutions, Inc. · posted Oct 20, 2020 · Terminated
The inner lumen of the Langston catheter may separate from the hub during pressure injection.
- Z-0087-2021 · Medtronic Vascular · posted Oct 16, 2020 · Terminated
Subsequent failure of catheters in the field and failed quality testing
- Z-0088-2021 · Medtronic Vascular · posted Oct 16, 2020 · Terminated
Subsequent failure of catheters in the field and failed quality testing
- Z-0089-2021 · Medtronic Vascular · posted Oct 16, 2020 · Terminated
Subsequent failure of catheters in the field and failed quality testing
Class II recalls (9)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0299-2021 · RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) · posted Oct 27, 2020 · Terminated
The Streamline Navigation System taps have experienced binding with other devices.
- Z-0300-2021 · RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) · posted Oct 27, 2020 · Terminated
The Streamline Navigation System taps have experienced binding with other devices.
- Z-0292-2021 · Pentax of America Inc · posted Oct 23, 2020 · Terminated
There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is…
- Z-0293-2021 · Pentax of America Inc · posted Oct 23, 2020 · Terminated
There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is…
- Z-0134-2021 · Siemens Healthcare Diagnostics, Inc. · posted Oct 20, 2020 · Terminated
Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the…
- Z-0101-2021 · Siemens Healthcare Diagnostics, Inc. · posted Oct 13, 2020 · Open, Classified
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes…
- Z-0102-2021 · Siemens Healthcare Diagnostics, Inc. · posted Oct 13, 2020 · Open, Classified
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes…
- Z-0097-2021 · Biomet, Inc. · posted Oct 9, 2020 · Terminated
Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package…
- Z-0060-2021 · Encore Medical, LP · posted Oct 7, 2020 · Open, Classified
There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.