uMI 550 PET/CT Diagnostic Imaging System - Product Usage:… recallZ-2800-2020 · Shanghai United Imaging Healthcare Co., Ltd. · initiated Jul 16, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Shanghai United Imaging Healthcare Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions (N/A for Serial No. 200036)
FDA root cause category: Device Design
What the company is doing
Urgent Medical Device Correction Letters regarding the uEXPLORER and uMI 550 were sent to customers on July 16, 2020 via email. Customers were informed that the firm had received complaints on these products and advised customers to take the following actions if problems were observed: 1. if image artifacts occur, reboot the system and continue the reconstruction. If necessary, repeat the scan which will result in an additional dose of radiation and the possible need for additional contrast medium. 2. If scout scanning is unexpectedly interrupted, try repeating the scout. This will result in an additional dose of radiation. 3: Be careful to avoid the sharp edge when the patient raises their arms during movement of the Ml patient table. Customers with questions may contact their United Imaging Service Engineer or call the United Imaging Customer Contact Center at 24-hour hotline (855) 221-1552 (toll free) or (832) 699-2799 (direct). Customers were advised to report adverse reactions or quality problems experienced with the use of these products to the FDA's MedWatch Adverse Event Reporting program either online, by regulatory mail or by fax. Customers were requested to in-turn notify all users of the affected products via the Customer Notification Letter. Customers were informed that United Imaging is updating software to solve these issues, and that United Imaging will replace the mylar strip with an improved one on the affected products via on-site update. Customers are requested to sign the Receipt of Customer Notification Form indicating they received the notice and that they will extend the notification to the final customer. Signed forms are to be returned to the recalling firm.
Recall details
- Recall number
- Z-2800-2020
- Status
- Terminated
- Initiated by the company
- Jul 16, 2020
- Posted by the FDA
- Aug 13, 2020
- Recalling firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Quantity
- 4 units
- Product code
- JAK
- FDA submissions
- K182237
- Distribution
- US Nationwide distribution including in the state of Texas.
Product description
uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Code information (models, lots, serial numbers)
Serial No.s 200017, 200023, 200024 and 200036.
Other devices under the same FDA submission
2 devices cleared under K182237 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- uMI uMI 550Shanghai United Imaging Healthcare Co., Ltd. · System, Tomography, Computed, Emission6 recalls0 reports
- uMI uMI 550Shanghai United Imaging Healthcare Co., Ltd. · System, Tomography, Computed, Emission2 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 85948).
- Z-2799-2020 · TerminateduEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, d…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page