CARESCAPE R860 Ventilator recallZ-2883-2020 · GE Healthcare, LLC · initiated Jul 14, 2020

Class II Terminated All recalls posted in August 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The oxygen sensor for certain CARESCAPE R860 devices has a potential issue that may result in an inaccurate display of FiO2 value from what is being delivered.

FDA root cause category: Process control

What the company is doing

On July 14th, 2020 the firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated 7/14/2020 to its customers with the following instructions: Safety Instructions You may continue to use the CARESCAPE R860 Ventilator. Ensure that the FiO2 alarm limit is set to the factory default limit +/- 6 % so that the clinician is aware if the measured value differs from set value. When making an extubation decision, use the set FiO2 value instead of the FiO2 measured by the oxygen sensor. The user can use other methods to measure FiO2 or switch to another device and call GE Healthcare. Product Correction: GE Healthcare will correct all affected units at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. Contact Information: If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-2883-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 14, 2020
Posted by the FDA
Aug 24, 2020
Recalling firm
GE Healthcare, LLC
Quantity
2277 units
Product code
CBK
FDA submissions
K142679
Distribution
Domestic Distribution: USA International Distribution: Australia, Bahrain, Bermuda, Canada, Egypt, France, Germany, Honduras, India, Iraq, Italy, Lebanon, Mexico, Panama, Peru, Russia, Saudi Arabia, Singapore, South Africa, Spain, Taiwan and United Arab Emirates.

Product description

CARESCAPE R860 Ventilator The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and Volume.

Code information (models, lots, serial numbers)

Show the full code information (22k characters)

CARESCAPE R860: 1506-8600-000 Spare parts: CARESCAPE R860: 1505-3215-000, 2071409-S, 2071410-001 Engstrom: M1115997-S, M1081816-S  Identifier # CBRZ03152, CBRZ03159, CBRZ03171, CBRZ03173, CBRZ03174, CBRZ03175, CBRZ03176, CBRZ03177, CBRZ03180, CBRZ03182, CBRZ03867, CBRZ03869, CBRZ03870, CBRZ03873, CBRZ03856, CBRZ03259, CBRZ03261, CBRZ03263, CBRZ03264 CBRZ03265, CBRZ03267, CBRZ03274, CBRZ03275, CBRZ03278, CBRZ03281, CBRZ03928, CBRZ03946, CBRZ03989, CBRZ03990, CBRZ03994, CBRZ03845, CBRZ03857, CBRZ03858, CBRZ03861, CBRZ03868, CBRZ03872, CBRZ03879, CBRZ03880, CBRZ03881, CBRZ03883, CBRZ03889, CBRZ03890, CBRZ03891, CBRZ03484, CBRZ03485, CBRZ03487, CBRZ03488, CBRZ03490, CBRZ03502, CBRZ03793, CBRZ03842, CBRZ03843, CBRZ03849, CBRZ03850, CBRZ03851, CBRZ03853, CBRZ03864, CBRZ03865, CBRZ03866, CBRZ03438, CBRZ03501, CBRZ03503, CBRZ03506, CBRZ03508, CBRZ03609, CBRZ03610, CBRZ03620, CBRZ03621, CBRZ03625, CBRZ03342, CBRZ03437, CBRZ03451, CBRZ03724, CBRZ03730, CBRZ03731, CBRZ03732, CBRZ03733, CBRZ03734, CBRZ03759, CBRZ03767, CBRZ03768, CBRZ03769, CBRZ03771, CBRZ03772, CBRZ03773, CBRZ03774, CBRZ03775, CBRZ03776, CBRZ03777, CBRZ03778, CBRZ03779, CBRZ03780, CBRZ03781, CBRZ03782, CBRZ03783, CBRZ03784, CBRZ03785, CBRZ03786, CBRZ03787, CBRZ03788, CBRZ03789, CBRZ03791, CBRZ03792, CBRZ03794, CBRZ03795, CBRZ03796, CBRZ03797, CBRZ03798, CBRZ03799, CBRZ03800, CBRZ03801, CBRZ03802, CBRZ03803, CBRZ03804, CBRZ03805, CBRZ03806, CBRZ03807, CBRZ03808, CBRZ03809, CBRZ03810, CBRZ03812, CBRZ03813, CBRZ03815, CBRZ03816, CBRZ03817, CBRZ03819, CBRZ03821, CBRZ03822, CBRZ03823, CBRZ03824, CBRZ03825, CBRZ03826, CBRZ03827, CBRZ03828, CBRZ03829, CBRZ03830, CBRZ03831, CBRZ03832, CBRZ03834, CBRZ03835, CBRZ03836, CBRZ03837, CBRZ03838, CBRZ03839, CBRZ03841, CBRZ03877, CBRZ03913, CBRZ03925, CBRZ03949, CBRZ03951, CBRZ03952, CBRZ03954, CBRZ03955, CBRZ03956, CBRZ03957, CBRZ03958, CBRZ03959, CBRZ03961, CBRZ03962, CBRZ03979, CBRZ03984, CBRZ03985, CBRZ03997, CBRZ03998, CBRZ04001, CBRZ04003, CBRZ04006, CBRZ04008, CBRZ04009, CBRZ04011, CBRZ04020, CBRZ04025, CBRZ04036, CBRZ04038, CBRZ04039, CBRZ04040, CBRZ04041, CBRZ04042, CBRZ04043, CBRZ04046, CBRZ04047, CBRZ04050, CBRZ04051, CBRZ04052, CBRZ04053, CBRZ04054, CBRZ04057, CBRZ04058, CBRZ04059, CBRZ04061, CBRZ04062, CBRZ04063, CBRZ04064, CBRZ04065, CBRZ04066, CBRZ04067, CBRZ04068, CBRZ04069, CBRZ04070, CBRZ04071, CBRZ04072, CBRZ04073, CBRZ04074, CBRZ04075, CBRZ04076, CBRZ04078, CBRZ04080, CBRZ04081, CBRZ04082, CBRZ04083, CBRZ04084, CBRZ04085, CBRZ04086, CBRZ04087, CBRZ04088, CBRZ04089, CBRZ04090, CBRZ04091, CBRZ04092, CBRZ04093, CBRZ04094, CBRZ04095, CBRZ04096, CBRZ04097, CBRZ04098, CBRZ04099, CBRZ04100, CBRZ04101, CBRZ04102, CBRZ04103, CBRZ04104, CBRZ04105, CBRZ04106, CBRZ04107, CBRZ04108, CBRZ04109, CBRZ04110, CBRZ04111, CBRZ04112, CBRZ04113, CBRZ04115, CBRZ04116, CBRZ04117, CBRZ04118, CBRZ04119, CBRZ04120, CBRZ04121, CBRZ04122, CBRZ04123, CBRZ04124, CBRZ04130, CBRZ04137, CBRZ04139, CBRZ04196, CBRZ04212, CBRZ04269, CBRZ04273, CBRZ04330, CBRZ04336, CBRZ04373, CBRZ04374, CBRZ04376, CBRZ04377, CBRZ04378, CBRZ04379, CBRZ04380, CBRZ04387, CBRZ04389, CBRZ04438, CBRZ04446, CBRZ04447, CBRZ04449, CBRZ04450, CBRZ04485, CBRZ04491, CBRZ04506, CBRZ04511, CBRZ04524, CBRZ04649, CBRZ04684, CBRZ04692, CBRZ04693, CBRZ04694, CBRZ04695, CBRZ04696, CBRZ04698, CBRZ04699, CBRZ04700, CBRZ04701, CBRZ04702, CBRZ04703, CBRZ04704, CBRZ04705, CBRZ04706, CBRZ04707, CBRZ04708, CBRZ04709, CBRZ04710, CBRZ04711, CBRZ04712, CBRZ04713, CBRZ04714, CBRZ04715, CBRZ04716, CBRZ04717, CBRZ04718, CBRZ04719, CBRZ04720, CBRZ04721, CBRZ04722, CBRZ04723, CBRZ04724, CBRZ04725, CBRZ04726, CBRZ04727, CBRZ04728, CBRZ04729, CBRZ04730, CBRZ04731, CBRZ04733, CBRZ04734, CBRZ04735, CBRZ04736, CBRZ04737, CBRZ04738, CBRZ04739, CBRZ04740, CBRZ04741, CBRZ04742, CBRZ04743, CBRZ04744, CBRZ04745, CBRZ04746, CBRZ04747, CBRZ04748, CBRZ04749, CBRZ04750, CBRZ04751, CBRZ04752, CBRZ04753, CBRZ04754, CBRZ04755, CBRZ04759, CBRZ04762, CBRZ04770, CBRZ04772, CBRZ04778, CBRZ04779, CBRZ04788, CBRZ04790, CBRZ04791, CBRZ04800, CBRZ04813, CBRZ04814, CBRZ04815, CBRZ04816, CBRZ04817, CBRZ04818, CBRZ04819, CBRZ04820, CBRZ04821, CBRZ04822, CBRZ04823, CBRZ04824, CBRZ04825, CBRZ04826, CBRZ04827, CBRZ04828, CBRZ04829, CBRZ04830, CBRZ04831, CBRZ04832, CBRZ04833, CBRZ04834, CBRZ04837,CBRZ04838, CBRZ04839, CBRZ04840, CBRZ04841, CBRZ04842, CBRZ04843, CBRZ04847, CBRZ04848, CBRZ04849, CBRZ04850, CBRZ04851, CBRZ04852, CBRZ04853, CBRZ04854, CBRZ04855, CBRZ04856, CBRZ04857, CBRZ04859, CBRZ04860, CBRZ04861, CBRZ04862, CBRZ04863, CBRZ04864, CBRZ04865, CBRZ04866, CBRZ04867, CBRZ04868, CBRZ04869, CBRZ04870, CBRZ04871, CBRZ04872, CBRZ04873, CBRZ04874, CBRZ04875, CBRZ04876, CBRZ04877, CBRZ04878, CBRZ04879, CBRZ04880, CBRZ04881, CBRZ04882, CBRZ04883, CBRZ04884, CBRZ04885, CBRZ04886, CBRZ04887, CBRZ04888, CBRZ04889, CBRZ04890, CBRZ04891, CBRZ04892, CBRZ04893, CBRZ04894, CBRZ04895, CBRZ04897, CBRZ04898, CBRZ04899, CBRZ04900, CBRZ04901, CBRZ04902, CBRZ04903, CBRZ04904, CBRZ04905, CBRZ04906, CBRZ04907, CBRZ04908, CBRZ04909, CBRZ04910, CBRZ04911, CBRZ04912, CBRZ04913, CBRZ04914, CBRZ04915, CBRZ04917, CBRZ04918, CBRZ04919, CBRZ04920, CBRZ04921, CBRZ04922, CBRZ04923, CBRZ04924, CBRZ04925, CBRZ04927, CBRZ04928, CBRZ04929, CBRZ04930, CBRZ04931, CBRZ04932, CBRZ04933, CBRZ04934, CBRZ04935, CBRZ04936, CBRZ04947, CBRZ04951, CBRZ05012, CBRZ05013, CBRZ05014, CBRZ05015, CBRZ05016, CBRZ05017, CBRZ05018, CBRZ05019, CBRZ05020, CBRZ05021, CBRZ05022, CBRZ05023, CBRZ05024, CBRZ05025, CBRZ05026, CBRZ05027, CBRZ05028, CBRZ05029, CBRZ05030, CBRZ05031, CBRZ05032, CBRZ05033, CBRZ05034, CBRZ05035, CBRZ05036, CBRZ05037, CBRZ05038, CBRZ05039, CBRZ05040, CBRZ05041, CBRZ05042, CBRZ05043, CBRZ05044, CBRZ05045, CBRZ05046, CBRZ05047, CBRZ05048, CBRZ05049, CBRZ05050, CBRZ05051,CBRZ05052, CBRZ05053, CBRZ05054, CBRZ05055, CBRZ05056, CBRZ05057, CBRZ05058, CBRZ05059, CBRZ05060, CBRZ05061, CBRZ05062, CBRZ05063, CBRZ05064, CBRZ05065, CBRZ05066, CBRZ05067, CBRZ05068, CBRZ05069, CBRZ05070, CBRZ05071, CBRZ05072, CBRZ05073, CBRZ05074, CBRZ05075, CBRZ05076, CBRZ05077, CBRZ05078, CBRZ05079, CBRZ05080, CBRZ05081, CBRZ05082, CBRZ05083, CBRZ05084, CBRZ05085, CBRZ05086, CBRZ05087, CBRZ05088, CBRZ05089, CBRZ05090, CBRZ05091, CBRZ05092, CBRZ05093, CBRZ05094, CBRZ05095, CBRZ05096, CBRZ05097, CBRZ05098, CBRZ05099, CBRZ05100, CBRZ05101, CBRZ05102, CBRZ05103, CBRZ05106, CBRZ05107, CBRZ05108, CBRZ05109, CBRZ05110, CBRZ05111, CBRZ05112, CBRZ05113 CBRZ05114, CBRZ05115, CBRZ05116, CBRZ05117, CBRZ05118, CBRZ05119, CBRZ05120, CBRZ05121, CBRZ05122, CBRZ05123, CBRZ05124, CBRZ05125, CBRZ05126, CBRZ05127, CBRZ05128, CBRZ05129, CBRZ05130, CBRZ05131, CBRZ05132, CBRZ05133, CBRZ05134, CBRZ05135, CBRZ05136, CBRZ05137, CBRZ05138, CBRZ05139, CBRZ05140, CBRZ05141, CBRZ05142, CBRZ05143, CBRZ05144, CBRZ05145, CBRZ05146, CBRZ05147, CBRZ05148, CBRZ05149, CBRZ05150, CBRZ05151, CBRZ05152, CBRZ05154, CBRZ05153, CBRZ05155, CBRZ05156, CBRZ05157, CBRZ05158, CBRZ05159, CBRZ05160, CBRZ05161, CBRZ05162, CBRZ05163, CBRZ05164, CBRZ05165, CBRZ05166, CBRZ05167, CBRZ05168, CBRZ05169, CBRZ05170, CBRZ05171, CBRZ05172, CBRZ05173, CBRZ05174, CBRZ05175, CBRZ05176, CBRZ05177, CBRZ05178 CBRZ05179, CBRZ05180, CBRZ05181, CBRZ05182, CBRZ05183, CBRZ05184, CBRZ05185, CBRZ05186, CBRZ05187, CBRZ05188, CBRZ05189, CBRZ05190, CBRZ05191, CBRZ05192, CBRZ05193, CBRZ05194, CBRZ05195, CBRZ05196, CBRZ05197, CBRZ05198, CBRZ05199, CBRZ05200, CBRZ05201, CBRZ05202, CBRZ05203, CBRZ05204, CBRZ05205, CBRZ05206, CBRZ05207, CBRZ05208, CBRZ05209, CBRZ05210, CBRZ05211, CBRZ05212, CBRZ05213, CBRZ05214, CBRZ05215, CBRZ05216, CBRZ05217, CBRZ05218, CBRZ05219, CBRZ05220, CBRZ05221, CBRZ05222, CBRZ05223, CBRZ05224, CBRZ05225, CBRZ05226, CBRZ05227, CBRZ05228, CBRZ05229, CBRZ05230, CBRZ05231, CBRZ05232, CBRZ05233, CBRZ05234, CBRZ05235, CBRZ05236, CBRZ05237, CBRZ05238, CBRZ05239, CBRZ05241, CBRZ05242, CBRZ05244, CBRZ05245, CBRZ05247, CBRZ05248, CBRZ05249, CBRZ05252, CBRZ05255, CBRZ05256, CBRZ05257, CBRZ05258, CBRZ05259, CBRZ05260, CBRZ05261, CBRZ05262, CBRZ05263, CBRZ05264, CBRZ05265, CBRZ05266, CBRZ05267, CBRZ05268, CBRZ05273, CBRZ05276, CBRZ05277, CBRZ05278, CBRZ05279, CBRZ05280, CBRZ05281, CBRZ05282, CBRZ05283, CBRZ05284, CBRZ05285, CBRZ05286, CBRZ05287, CBRZ05288, CBRZ05289, CBRZ05290, CBRZ05291, CBRZ05292, CBRZ05294, CBRZ05295, CBRZ05296, CBRZ05302, CBRZ05303, CBRZ05304, CBRZ05305, CBRZ05306, CBRZ05309, CBRZ05311, CBRZ05312, CBRZ05313, CBRZ05314 CBRZ05315, CBRZ05317, CBRZ05318, CBRZ05319, CBRZ05320, CBRZ05321, CBRZ05322, CBRZ05323, CBRZ05324, CBRZ05326, CBRZ05327, CBRZ05328, CBRZ05329, CBRZ05330 CBRZ05331, CBRZ05332, CBRZ05333, CBRZ05334, CBRZ05335, CBRZ05340, CBRZ40010, CBRZ40038, CBRZ40039, CBRZ40040, CBRZ03469, CBRZ03504, CBRZ03505, CBRZ03510, CBRZ03513, CBRZ03514, CBRZ03515, CBRZ03516, CBRZ03517, CBRZ03518, CBRZ03522, CBRZ03523, CBRZ03526, CBRZ03531, CBRZ03532, CBRZ03533, CBRZ03534, CBRZ03535, CBRZ03536, CBRZ03537, CBRZ03538, CBRZ03539, CBRZ03540, CBRZ03541, CBRZ03543, CBRZ03544, CBRZ03548, CBRZ03550, CBRZ03552, CBRZ02183, CBRZ03110, CBRZ03128, CBRZ03130, CBRZ03131, CBRZ03132, CBRZ03133, CBRZ03134, CBRZ03135, CBRZ03136, CBRZ03137, CBRZ03138, CBRZ03139, CBRZ03140, CBRZ03141, CBRZ03142, CBRZ03143, CBRZ03144, CBRZ03145, CBRZ03146, CBRZ03147, CBRZ03148, CBRZ03149, CBRZ03150, CBRZ03151, CBRZ03153, CBRZ03154, CBRZ03156, CBRZ03158, CBRZ03160, CBRZ03161, CBRZ03163, CBRZ03164, CBRZ03165, CBRZ03167, CBRZ03170, CBRZ03185, CBRZ03188, CBRZ03202, CBRZ03206, CBRZ03208, CBRZ03209, CBRZ03210, CBRZ03211, CBRZ03212, CBRZ03213, CBRZ03214, CBRZ03215, CBRZ03216, CBRZ03217, CBRZ03218, CBRZ03219, CBRZ03220, CBRZ03221, CBRZ03222, CBRZ03223, CBRZ03226, CBRZ03227, CBRZ03228, CBRZ03229, CBRZ03230, CBRZ03232, CBRZ03234, CBRZ03236, CBRZ03237, CBRZ03238, CBRZ03239, CBRZ03240, CBRZ03241, CBRZ03242, CBRZ03243, CBRZ03244, CBRZ03245, CBRZ03246, CBRZ03247, CBRZ03248, CBRZ03249, CBRZ03250, CBRZ03251, CBRZ03252, CBRZ03253, CBRZ03254, CBRZ03255, CBRZ03256, CBRZ03258, CBRZ03262, CBRZ03269, CBRZ03270, CBRZ03271, CBRZ03272, CBRZ03273, CBRZ03277, CBRZ03279, CBRZ03280, CBRZ03283, CBRZ03286, CBRZ03287, CBRZ03288, CBRZ03289, CBRZ03290, CBRZ03291, CBRZ03292, CBRZ03293, CBRZ03294, CBRZ03295, CBRZ03296, CBRZ03297, CBRZ03298, CBRZ03299, CBRZ03300,CBRZ03301, CBRZ03302, CBRZ03303, CBRZ03304, CBRZ03305, CBRZ03306, CBRZ03307, CBRZ03308, CBRZ03309, CBRZ03310, CBRZ03311, CBRZ03315, CBRZ03317, CBRZ03318, CBRZ03319, CBRZ03320, CBRZ03321, CBRZ03322, CBRZ03323, CBRZ03324, CBRZ03325, CBRZ03326 ,CBRZ03327, CBRZ03329, CBRZ03330, CBRZ03331, CBRZ03332, CBRZ03333, CBRZ03334, CBRZ03335, CBRZ03336, CBRZ03337, CBRZ03338, CBRZ03339, CBRZ03340, CBRZ03341, CBRZ03343, CBRZ03344, CBRZ03345, CBRZ03346, CBRZ03347, CBRZ03348, CBRZ03349, CBRZ03350, CBRZ03351, CBRZ03352, CBRZ03353, CBRZ03354, CBRZ03355, CBRZ03356, CBRZ03357, CBRZ03358, CBRZ03359, CBRZ03360, CBRZ03361, CBRZ03362, CBRZ03363, CBRZ03364, CBRZ03365, CBRZ03366, CBRZ03367, CBRZ03368, CBRZ03369, CBRZ03370, CBRZ03371, CBRZ03372, CBRZ03373, CBRZ03374, CBRZ03375, CBRZ03376, CBRZ03377, CBRZ03378, CBRZ03379, CBRZ03380, CBRZ03381, CBRZ03382, CBRZ03383, CBRZ03384, CBRZ03385, CBRZ03386, CBRZ03387, CBRZ03388, CBRZ03389, CBRZ03390, CBRZ03391, CBRZ03392, CBRZ03393, CBRZ03394, CBRZ03395, CBRZ03396, CBRZ03397, CBRZ03398, CBRZ03399, CBRZ03400, CBRZ03401, CBRZ03402, CBRZ03403, CBRZ03404, CBRZ03405, CBRZ03406, CBRZ03407, CBRZ03408, CBRZ03409, CBRZ03410, CBRZ03411, CBRZ03412, CBRZ03413, CBRZ03414, CBRZ03415, CBRZ03416, CBRZ03417, CBRZ03418, CBRZ03419, CBRZ03420, CBRZ03421, CBRZ03422, CBRZ03423, CBRZ03424, CBRZ03425, CBRZ03426, CBRZ03428, CBRZ03429, CBRZ03430, CBRZ03431, CBRZ03432, CBRZ03433, CBRZ03434, CBRZ03435, CBRZ03436, CBRZ03439, CBRZ03440, CBRZ03441, CBRZ03442, CBRZ03443, CBRZ03444, CBRZ03445, CBRZ03446, CBRZ03447, CBRZ03448, CBRZ03449, CBRZ03450, CBRZ03452, CBRZ03453, CBRZ03454, CBRZ03455, CBRZ03456, CBRZ03457, CBRZ03458, CBRZ03459, CBRZ03460, CBRZ03461, CBRZ03462, CBRZ03463, CBRZ03464, CBRZ03465, CBRZ03466, CBRZ03467, CBRZ03468 CBRZ03470, CBRZ03471, CBRZ03473, CBRZ03474, CBRZ03475, CBRZ03476, CBRZ03479, CBRZ03480, CBRZ03481, CBRZ03486, CBRZ03507, CBRZ03512, CBRZ03519, CBRZ03520, CBRZ03524, CBRZ03525, CBRZ03527, CBRZ03529, CBRZ03530, CBRZ03542, CBRZ03545, CBRZ03546, CBRZ03547, CBRZ03549, CBRZ03551, CBRZ03553, CBRZ03554, CBRZ03555, CBRZ03556, CBRZ03557, CBRZ03558, CBRZ03559, CBRZ03560, CBRZ03561, CBRZ03562, CBRZ03563, CBRZ03564, CBRZ03565, CBRZ03566, CBRZ03567, CBRZ03568, CBRZ03569, CBRZ03570, CBRZ03571, CBRZ03572, CBRZ03573, CBRZ03574, CBRZ03576, CBRZ03577, CBRZ03578, CBRZ03579, CBRZ03580, CBRZ03581, CBRZ03582, CBRZ03583, CBRZ03584, CBRZ03585, CBRZ03586, CBRZ03587, CBRZ03588, CBRZ03589, CBRZ03590, CBRZ03591, CBRZ03592, CBRZ03594, CBRZ03595, CBRZ03599, CBRZ03601, CBRZ03604, CBRZ03605, CBRZ03606, CBRZ03607, CBRZ03608, CBRZ03611, CBRZ03612, CBRZ03613, CBRZ03614, CBRZ03615, CBRZ03617, CBRZ03618, CBRZ03619, CBRZ03626, CBRZ03628, CBRZ03631, CBRZ03633, CBRZ03634, CBRZ03635, CBRZ03636, CBRZ03637, CBRZ03638, CBRZ03639, CBRZ03640, CBRZ03641, CBRZ03643, CBRZ03644, CBRZ03645, CBRZ03648, CBRZ03650, CBRZ03651, CBRZ03652, CBRZ03653, CBRZ03654, CBRZ03655, CBRZ03656, CBRZ03657, CBRZ03658, CBRZ03659, CBRZ03660, CBRZ03661, CBRZ03662, CBRZ03663, CBRZ03664, CBRZ03665, CBRZ03667, CBRZ03668, CBRZ03670, CBRZ03689, CBRZ03695, CBRZ03707, CBRZ03712, CBRZ03714, CBRZ03716, CBRZ03717, CBRZ03718, CBRZ03719, CBRZ03720, CBRZ03721, CBRZ03722, CBRZ03723, CBRZ03725, CBRZ03726, CBRZ03727, CBRZ03728, CBRZ03729, CBRZ03736, CBRZ03738, CBRZ03739, CBRZ03740, CBRZ03741, CBRZ03745, CBRZ03747, CBRZ03748, CBRZ03749, CBRZ03750, CBRZ03751, CBRZ03752, CBRZ03753, CBRZ03754, CBRZ03755, CBRZ03756, CBRZ03757, CBRZ03758, CBRZ03760, CBRZ03761, CBRZ03762, CBRZ03763, CBRZ03764, CBRZ03765, CBRZ03766, CBRZ03770, CBRZ03903, CBRZ03932, CBRZ03943, CBRZ03944, CBRZ03945, CBRZ03953, CBRZ03960, CBRZ03963, CBRZ03964, CBRZ03965, CBRZ03966, CBRZ03967, CBRZ03968, CBRZ03969, CBRZ03970, CBRZ03971, CBRZ03972, CBRZ03973, CBRZ03974, CBRZ03975, CBRZ03976, CBRZ03977, CBRZ03978, CBRZ03981, CBRZ03982, CBRZ03983, CBRZ03991, CBRZ03992, CBRZ03993, CBRZ03995, CBRZ03996, CBRZ03999, CBRZ04000, CBRZ04002, CBRZ04004, CBRZ04005, CBRZ04007, CBRZ04010, CBRZ04012, CBRZ04013, CBRZ04016, CBRZ04017, CBRZ04018, CBRZ04019, CBRZ04027, CBRZ04029, CBRZ04030, CBRZ04032, CBRZ04033, CBRZ04034, CBRZ04035, CBRZ04037, CBRZ04044, CBRZ04045, CBRZ04048, CBRZ04049, CBRZ04055, CBRZ04056, CBRZ04060, CBRZ04077, CBRZ04079, CBRZ04114, CBRZ04126, CBRZ04127, CBRZ04128, CBRZ04129, CBRZ04131, CBRZ04132, CBRZ04133, CBRZ04134, CBRZ04135, CBRZ04136, CBRZ04138, CBRZ04140, CBRZ04141, CBRZ04142, CBRZ04143, CBRZ04144, CBRZ04145, CBRZ04146, CBRZ04147, CBRZ04148, CBRZ04149, CBRZ04150, CBRZ04151,CBRZ04152, CBRZ04153, CBRZ04154, CBRZ04155, CBRZ04156, CBRZ04157, CBRZ04158, CBRZ04159, CBRZ04160, CBRZ04161, CBRZ04162, CBRZ04163, CBRZ04164, CBRZ04165, CBRZ04166, CBRZ04167, CBRZ04168, CBRZ04169, CBRZ04170, CBRZ04171, CBRZ04173, CBRZ04174, CBRZ04175, CBRZ04177, CBRZ04178, CBRZ04179, CBRZ04180, CBRZ04181, CBRZ04182, CBRZ04183, CBRZ04184, CBRZ04185, CBRZ04186, CBRZ04187, CBRZ04188, CBRZ04189, CBRZ04190, CBRZ04191, CBRZ04192, CBRZ04193, CBRZ04194, CBRZ04195, CBRZ04197, CBRZ04198, CBRZ04199, CBRZ04200, CBRZ04201, CBRZ04202, CBRZ04203, CBRZ04204, CBRZ04205, CBRZ04206, CBRZ04207, CBRZ04208, CBRZ04209, CBRZ04210, CBRZ04211, CBRZ04213, CBRZ04214, CBRZ04215, CBRZ04216, CBRZ04217, CBRZ04218, CBRZ04219, CBRZ04220, CBRZ04221, CBRZ04222, CBRZ04223, CBRZ04224, CBRZ04225, CBRZ04226, CBRZ04227, CBRZ04228, CBRZ04229, CBRZ04230, CBRZ04231, CBRZ04232, CBRZ04233, CBRZ04234, CBRZ04235, CBRZ04236, CBRZ04237, CBRZ04238, CBRZ04239, CBRZ04240, CBRZ04241, CBRZ04242, CBRZ04243, CBRZ04244, CBRZ04245, CBRZ04246, CBRZ04247, CBRZ04248, CBRZ04249, CBRZ04250, CBRZ04251, CBRZ04252, CBRZ04253, CBRZ04254, CBRZ04255, CBRZ04256, CBRZ04257, CBRZ04258, CBRZ04259, CBRZ04260, CBRZ04261, CBRZ04262, CBRZ04263, CBRZ04264, CBRZ04265, CBRZ04266, CBRZ04267, CBRZ04268, CBRZ04270, CBRZ04271, CBRZ04272, CBRZ04274, CBRZ04275, CBRZ04276, CBRZ04277, CBRZ04278, CBRZ04279, CBRZ04280, CBRZ04281, CBRZ04282, CBRZ04283, CBRZ04284, CBRZ04285, CBRZ04286, CBRZ04287, CBRZ04288, CBRZ04289, CBRZ04290, CBRZ04291, CBRZ04292, CBRZ04293, CBRZ04294, CBRZ04295, CBRZ04296, CBRZ04297, CBRZ04298, CBRZ04299, CBRZ04300, CBRZ04301, CBRZ04302, CBRZ04303, CBRZ04304, CBRZ04305, CBRZ04306, CBRZ04307, CBRZ04308, CBRZ04309, CBRZ04310, CBRZ04311, CBRZ04312, CBRZ04313, CBRZ04314, CBRZ04315, CBRZ04316, CBRZ04317, CBRZ04318, CBRZ04319, CBRZ04320, CBRZ04321, CBRZ04322, CBRZ04323, CBRZ04324, CBRZ04325, CBRZ04326, CBRZ04327, CBRZ04328, CBRZ04329, CBRZ04331, CBRZ04332, CBRZ04333, CBRZ04334, CBRZ04335, CBRZ04337, CBRZ04338, CBRZ04339, CBRZ04340, CBRZ04341, CBRZ04343, CBRZ04344, CBRZ04345, CBRZ04346, CBRZ04347, CBRZ04348, CBRZ04349, CBRZ04350, CBRZ04351, CBRZ04352, CBRZ04353, CBRZ04354, CBRZ04355, CBRZ04356, CBRZ04357, CBRZ04358, CBRZ04359, CBRZ04365, CBRZ04366, CBRZ04367, CBRZ04368, CBRZ04369, CBRZ04370, CBRZ04371, CBRZ04372, CBRZ04375, CBRZ04381, CBRZ04382, CBRZ04383, CBRZ04384, CBRZ04385, CBRZ04386, CBRZ04388, CBRZ04390, CBRZ04391, CBRZ04392, CBRZ04393, CBRZ04394, CBRZ04395, CBRZ04396, CBRZ04397, CBRZ04398, CBRZ04399, CBRZ04400, CBRZ04401, CBRZ04402, CBRZ04403, CBRZ04404, CBRZ04405, CBRZ04406, CBRZ04407, CBRZ04408, CBRZ04409, CBRZ04410, CBRZ04411, CBRZ04412, CBRZ04413, CBRZ04414, CBRZ04415, CBRZ04416, CBRZ04417, CBRZ04418, CBRZ04419, CBRZ04420, CBRZ04421, CBRZ04422, CBRZ04423, CBRZ04424, CBRZ04425, CBRZ04426, CBRZ04427, CBRZ04428, CBRZ04429, CBRZ04430, CBRZ04431, CBRZ04432, CBRZ04433, CBRZ04434, CBRZ04435, CBRZ04436, CBRZ04437, CBRZ04439, CBRZ04440, CBRZ04441, CBRZ04442, CBRZ04443, CBRZ04444, CBRZ04445, CBRZ04448, CBRZ04451, CBRZ04452, CBRZ04453, CBRZ04454, CBRZ04455, CBRZ04456, CBRZ04457, CBRZ04458, CBRZ04459, CBRZ04460, CBRZ04461, CBRZ04462, CBRZ04463, CBRZ04464, CBRZ04465, CBRZ04466, CBRZ04467, CBRZ04468, CBRZ04469, CBRZ04470, CBRZ04471, CBRZ04472, CBRZ04473, CBRZ04474, CBRZ04475, CBRZ04476, CBRZ04477, CBRZ04478, CBRZ04479, CBRZ04480, CBRZ04481, CBRZ04482, CBRZ04483, CBRZ04484, CBRZ04486, CBRZ04487, CBRZ04488, CBRZ04489, CBRZ04490, CBRZ04492, CBRZ04493, CBRZ04494, CBRZ04495, CBRZ04496, CBRZ04497, CBRZ04498, CBRZ04499, CBRZ04500, CBRZ04501, CBRZ04502, CBRZ04503, CBRZ04504, CBRZ04505, CBRZ04507, CBRZ04508, CBRZ04509, CBRZ04510, CBRZ04512, CBRZ04513, CBRZ04514, CBRZ04515, CBRZ04516, CBRZ04517, CBRZ04518, CBRZ04519, CBRZ04520, CBRZ04521, CBRZ04522, CBRZ04523, CBRZ04525, CBRZ04526, CBRZ04527, CBRZ04528, CBRZ04529, CBRZ04530, CBRZ04531, CBRZ04532, CBRZ04533, CBRZ04534, CBRZ04535, CBRZ04536, CBRZ04537, CBRZ04538, CBRZ04539, CBRZ04540, CBRZ04541, CBRZ04542, CBRZ04543, CBRZ04544, CBRZ04545, CBRZ04546, CBRZ04547, CBRZ04548, CBRZ04549, CBRZ04550, CBRZ04551, CBRZ04552, CBRZ04553, CBRZ04554, CBRZ04555, CBRZ04556, CBRZ04557, CBRZ04558, CBRZ04559, CBRZ04560, CBRZ04561, CBRZ04562, CBRZ04563, CBRZ04564, CBRZ04565, CBRZ04566, CBRZ04567, CBRZ04568, CBRZ04569, CBRZ04570, CBRZ04571, CBRZ04572, CBRZ04573, CBRZ04574, CBRZ04575, CBRZ04576, CBRZ04577, CBRZ04578, CBRZ04579, CBRZ04580, CBRZ04581, CBRZ04582, CBRZ04583, CBRZ04584, CBRZ04585, CBRZ04586, CBRZ04587, CBRZ04588, CBRZ04589, CBRZ04590, CBRZ04591, CBRZ04592, CBRZ04593, CBRZ04594, CBRZ04595, CBRZ04596, CBRZ04597, CBRZ04598, CBRZ04599, CBRZ04600, CBRZ04601, CBRZ04602, CBRZ04603, CBRZ04604, CBRZ04605, CBRZ04606, CBRZ04607, CBRZ04608, CBRZ04609, CBRZ04610, CBRZ04611, CBRZ04612, CBRZ04613, CBRZ04614, CBRZ04615, CBRZ04616, CBRZ04617, CBRZ04618, CBRZ04619, CBRZ04620, CBRZ04621, CBRZ04622, CBRZ04623, CBRZ04624, CBRZ04625, CBRZ04626, CBRZ04627, CBRZ04628, CBRZ04629, CBRZ04630, CBRZ04631, CBRZ04632, CBRZ04633, CBRZ04634, CBRZ04635, CBRZ04636, CBRZ04637, CBRZ04638, CBRZ04639, CBRZ04640, CBRZ04641, CBRZ04642, CBRZ04643, CBRZ04644, CBRZ04645, CBRZ04646, CBRZ04647, CBRZ04648, CBRZ04650, CBRZ04651, CBRZ04652, CBRZ04653, CBRZ04654, CBRZ04655, CBRZ04656, CBRZ04657, CBRZ04658, CBRZ04659, CBRZ04660, CBRZ04661, CBRZ04662, CBRZ04663, CBRZ04664, CBRZ04665, CBRZ04666, CBRZ04667, CBRZ04668, CBRZ04669, CBRZ04670, CBRZ04671, CBRZ04672, CBRZ04673, CBRZ04674, CBRZ04675, CBRZ04676, CBRZ04677, CBRZ04678, CBRZ04679, CBRZ04680, CBRZ04681, CBRZ04682, CBRZ04683, CBRZ04685, CBRZ04686, CBRZ04687, CBRZ04688, CBRZ04689, CBRZ04690, CBRZ04691, CBRZ04697, CBRZ04732, CBRZ04756, CBRZ04757, CBRZ04758, CBRZ04760, CBRZ04761, CBRZ04763, CBRZ04764, CBRZ04765, CBRZ04766, CBRZ04767, CBRZ04768, CBRZ04769, CBRZ04771, CBRZ04773, CBRZ04774, CBRZ04775, CBRZ04776, CBRZ04777, CBRZ04780, CBRZ04781, CBRZ04782, CBRZ04783, CBRZ04784, CBRZ04785, CBRZ04786, CBRZ04787, CBRZ04789, CBRZ04792, CBRZ04793, CBRZ04794, CBRZ04795, CBRZ04796, CBRZ04797, CBRZ04798, CBRZ04799, CBRZ04801, CBRZ04802, CBRZ04803, CBRZ04804, CBRZ04805, CBRZ04806, CBRZ04807, CBRZ04808, CBRZ04809, CBRZ04810, CBRZ04811, CBRZ04812, CBRZ04835, CBRZ04836, CBRZ04844, CBRZ04845, CBRZ04846, CBRZ04858, CBRZ04896, CBRZ04916, CBRZ04926, CBRZ04937, CBRZ04938, CBRZ04939, CBRZ04940, CBRZ04941, CBRZ04942, CBRZ04943, CBRZ04944, CBRZ04945, CBRZ04946, CBRZ04948, CBRZ04949, CBRZ04950, CBRZ04952, CBRZ04953, CBRZ05010, CBRZ05011,CBRZ05243, CBRZ05246, CBRZ05250, CBRZ05251, CBRZ05253, CBRZ05254, CBRZ05269, CBRZ05270, CBRZ05271, CBRZ05272, CBRZ05274, CBRZ05275, CBRZ05297, CBRZ05298, CBRZ05299, CBRZ05300, CBRZ05301, CBRZ05308, CBRZ05316, CBRZ05339, CBRZ40011, CBRZ40012, CBRZ40013, CBRZ40014, CBRZ40015, CBRZ40016, CBRZ40017, CBRZ40018, CBRZ40019, CBRZ40020, CBRZ40021, CBRZ40022, CBRZ40023, CBRZ40024, CBRZ40025, CBRZ40026, CBRZ40027, CBRZ40028, CBRZ40029, CBRZ40030, CBRZ40031, CBRZ40032, CBRZ40033, CBRZ40034, CBRZ40035, CBRZ40036, CBRZ40037, CBRZ03627, CBRZ03632, CBRZ03642, CBRZ03491, CBRZ03818, CBRZ03840, CBRZ03844, CBRZ03846, CBRZ03847, CBRZ03478, CBRZ03509, CBRZ03511, CBRZ03521, CBRZ03528, CBRZ03172, CBRZ03183, CBRZ03189, CBRZ03190, CBRZ03192, CBRZ03477, CBRZ03500, CBRZ03472, CBRZ03489, CBRZ03493, CBRZ03312, CBRZ03313, CBRZ03314, CBRZ03316, CBRZ03328, 70187546-2, 70187546-3, 70187742-5, 70187742-6, 70187742-7, 70187742-8, 70187742-9, 70189854-4, 70189854-5, CBRZ03181, CBRZ03184, CBRZ03186, CBRZ03191, CBRZ03194, CBRZ03933, CBRZ03986, CBRZ03988, CBRZ03947, CBRZ03950, CBRZ03980, CBRZ03987, CBRZ03622, CBRZ03623, CBRZ03624, CBRZ03629, CBRZ03630, CBRZ03427, CBRZ03482, CBRZ03483, CBRZ03492, CBRZ03494, CBRZ03742, CBRZ03705, CBRZ03708, CBRZ03710, CBRZ03715, CBRZ03073, CBRZ03074, CBRZ03076, CBRZ03080, CBRZ03082.