Skip to content

PC Unit Front Case with Keypad Replacement Kits recallZ-2910-2020 · CareFusion 303, Inc. · initiated Aug 4, 2020

Class I Open, Classified All recalls posted in September 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.

FDA root cause category: Reprocessing Controls

What the company is doing

On 08/04/20, Urgent Medical Device Recall notices were mailed to Clinical Users: If the PC unit keypad becomes unresponsive or a key gets stuck, clinicians should remove the device from service and send to Biomedical Engineering. If administering a critical medication, continue the infusion until it is safe to replace the PC unit. Cleaning personnel can minimize the potential for fluid ingress during cleaning by 1) not using a cloth that drips, 2) wringing out the cleaning cloth to squeeze out excess fluid, and 3) not spraying fluids directly onto the device. Biomedical Engineering: If an affected PC unit has an unresponsive keypad or stuck key, remove the unit from service and call the recalling firm for information replacement parts. Until 510(k) clearance, the affected PC unit keypad design will continue to be provided for replacement or repair. Inspect all PC unit keypads for bubbling, cracks, damage, layer separations, and holes. Replace the front case keypad if it fails inspection. Return the completed Customer Response Form. Provide a copy of this letter and enclosures to customers who are currently renting affected devices. Firm is currently working on a design change for the PC unit keypad and will notify customers when replacement keypads are available. Additional Information: Customer Advocacy, 888-812-3266, [email protected]; Training Resources www.bd.com/alaris-system-hardware-recall; Recall Support Center, 888-562-6018, [email protected]; Technical Support, 888-812-3229, [email protected]; Customer Order Management, 1-800-482-4822, [email protected] On 12/16/22, Medical Device Notifications were sent to customers informing them that they may have received impacted parts for repair purposes or received product with impacted parts. Customers were provided with links to the original recall notices and asked to review the information and complete the customer response form.

Recall details

Recall number
Z-2910-2020
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Aug 4, 2020
Posted by the FDA
Sep 4, 2020
Recalling firm
CareFusion 303, Inc.
Quantity
28346
Product code
FRN
FDA submissions
K133532
Distribution
U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA

Product description

PC Unit Front Case with Keypad Replacement Kits: TC10008389 ASSY CASE FRONT W/KEYPAD 8015LS, TC10010217 ASSY FRT CASE W/ KEYPAD 8015 M2, TC10012515 ASSY FR CASE W/ KEYPAD 8015 M2, TC10013702 ASSY, CASE, FRONT W/KEYPAD, 8015LS, TC10013664 ASSY FR CASE W/ KEYPAD 8015 M2

Code information (models, lots, serial numbers)

All kits manufactured from April 07, 2017 to Present.

Other devices under the same FDA submission

4 devices cleared under K133532 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85900).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page