Medical device recalls: September 2020
The FDA posted 19 medical device recall records in September 2020, from 9 recall events by 7 companies. 5 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- CareFusion 303, Inc. · 2 events, 3 records, 3 Class I
- MicroPort Orthopedics Inc. · 2 events, 3 records, 2 Class I
- Stimwave Technologies Inc · 1 event, 6 records
- Integra LifeSciences Corp. · 1 event, 3 records
- Tosoh Bioscience Inc · 1 event, 2 records
- Novarad Corporation · 1 event, 1 record
- Siemens Medical Solutions USA, Inc · 1 event, 1 record
FDA root cause
- Under Investigation by firm · 8
- Mixed-up of materials/components · 3
- Reprocessing Controls · 2
- Device Design · 2
- Software design · 1
- Software Design Change · 1
- Labeling Change Control · 1
- Component design/selection · 1
Class I recalls (5)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2941-2020 · MicroPort Orthopedics Inc. · posted Sep 18, 2020 · Terminated
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among…
- Z-2942-2020 · MicroPort Orthopedics Inc. · posted Sep 18, 2020 · Terminated
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among…
- BD Alaris SystemClass IZ-2882-2020 · CareFusion 303, Inc. · posted Sep 4, 2020 · Open, Classified
The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion.
- Z-2909-2020 · CareFusion 303, Inc. · posted Sep 4, 2020 · Open, Classified
The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations…
- Z-2910-2020 · CareFusion 303, Inc. · posted Sep 4, 2020 · Open, Classified
The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations…
Class II recalls (14)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-3025-2020 · Siemens Medical Solutions USA, Inc · posted Sep 28, 2020 · Terminated
Sporadic problems with the current software may result in scanning workflow interruptions and unexpected user notifications causing delay in diagnosis or patient rescan
- Z-2967-2020 · Tosoh Bioscience Inc · posted Sep 14, 2020 · Terminated
Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
- Z-2968-2020 · Tosoh Bioscience Inc · posted Sep 14, 2020 · Terminated
Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
- Z-2953-2020 · Integra LifeSciences Corp. · posted Sep 12, 2020 · Terminated
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
- Z-2954-2020 · Integra LifeSciences Corp. · posted Sep 12, 2020 · Terminated
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
- Z-2955-2020 · Integra LifeSciences Corp. · posted Sep 12, 2020 · Terminated
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
- Z-2956-2020 · Novarad Corporation · posted Sep 12, 2020 · Terminated
The firm received a report of an atypical dataset being generated. When using the cross-localization feature and images from a modality that generates asymmetrical images, the cross-localization reference line may be inaccurately placed on any of the…
- Z-2948-2020 · MicroPort Orthopedics Inc. · posted Sep 9, 2020 · Open, Classified
MicroPort Orthopedics Inc. is voluntarily recalling any existing inventory of PROFEMUR Titanium and Cobalt Chrome modular necks with previous package insert versions to replace the PROFEMUR Hip System Package Insert with the most recent revision (150803-8).
- Z-2928-2020 · Stimwave Technologies Inc · posted Sep 2, 2020 · Open, Classified
The product contains a non-functional component not referenced in product labeling.
- Z-2929-2020 · Stimwave Technologies Inc · posted Sep 2, 2020 · Open, Classified
The product contains a non-functional component not referenced in product labeling.
- Z-2930-2020 · Stimwave Technologies Inc · posted Sep 2, 2020 · Open, Classified
The product contains a non-functional component not referenced in product labeling.
- Z-2931-2020 · Stimwave Technologies Inc · posted Sep 2, 2020 · Open, Classified
The product contains a non-functional component not referenced in product labeling.
- Z-2932-2020 · Stimwave Technologies Inc · posted Sep 2, 2020 · Open, Classified
The product contains a non-functional component not referenced in product labeling.
- Z-2933-2020 · Stimwave Technologies Inc · posted Sep 2, 2020 · Open, Classified
The product contains a non-functional component not referenced in product labeling.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.