Skip to content

CERTAIN BELLATEK recallZ-0076-2021 · Biomet 3i, LLC · initiated Aug 31, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet 3i, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

FDA root cause category: Device Design

What the company is doing

Zimmer Biomet issued Urgent Medical Device Recall letter dated 8/31/2020 to Dental Labs and Clinicians stating reason for recall, health risk and action to take: Quarantine any affected products (definitive BellaTek abutments/3D printed models) from Attachment 1 in your possession for return to Zimmer Biomet Dental. 2. Identify any clinicians who have received an affected BellaTek Definitive Abutment as identified in Attachment 1. 3. For clinicians that have received an affected BellaTek Definitive Abutment from you, provide a copy of this notice to them along with Attachment 3 for them to complete. B. Certificate of Acknowledgement 1. Complete Attachment 2 Certificate of Acknowledgement and email to [email protected]. Return of product 1. Include a copy of Attachment 2 Certificate of Acknowledgement with any product tobe returned. Using the return shipping label included with this notice, please returnaffected product to the address below. Customer Service will provide additionalshipping labels if needed. Field Action PM Regulatory Compliance Zimmer Biomet 4555 Riverside Dr. Palm Beach Gardens, FL 33410 US D. Retain a copy of the Acknowledgement Form with your recall records in the event of a compliance audit of your facilitys documentation. E. If you have any questions after reviewing this notice, please call or email the Customer Service team at: 1-888-800-8045 or [email protected].

Recall details

Recall number
Z-0076-2021
Status
Terminated
Initiated by the company
Aug 31, 2020
Recalling firm
Biomet 3i, LLC
Quantity
4 units
Product code
NHA
FDA submissions
K052648
Distribution
FL, GA, ID, IL, MI, NV, NY, OH, SC, TN, VA

Product description

CERTAIN BELLATEK TIN ABUTMENT 5.0MM Reference Number: IEDAN5

Code information (models, lots, serial numbers)

Lot Numbers: 8560155-1 8565718-1 8565734-1 8569780-1  UDI:  (01)00844868031154(10)8560155-1 (01)00844868031154(10)8565718-1  (01)00844868031154(10)8565734-1  (01)00844868031154(10)8569780-1

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 86402).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page