Skip to content

CERTAIN® BELLATEK® IEDAN5Recalls & reported problems

ZimVie US Corp LLC · CERTAIN® BELLATEK® TIN ABUTMENT 5.0MM · Class 2 · Implant

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Aug 31, 2020Terminated

    Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

    Z-0076-2021 · Root cause: Device Design · FDA record ↗

Other ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices

All ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.