RINGLOC BI-POLAR HIP SYSTEM recallZ-0097-2021 · Biomet, Inc. · initiated Sep 3, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package integrity for more than one-time sterilization.
FDA root cause category: Process control
What the company is doing
All affected customers were sent a letter on 09/03/2020 stating the following : Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package integrity for more than one-time sterilization. Zimmer Biomet will follow up w/ consignees that have not responded after the initial notification. Additional notifications via telephone, fax, email, or traceable carrier will be sent to those who have not completed the required certification form or provided surgeon and hospital information/confirmation. 100% of the delivery notifications will be accounted for. An inventory return certification form will be returned from each distributor that certifies that all available products from their territory have been returned or accounted for. Affected products that are not returned or accounted for will be considered as "not in inventory and unavailable for return." Zimmer Biomet will deem accounts unresponsive after 3 effectiveness checks if there continues to be no response are making reasonable efforts. Product that has not been recovered rom unresponsive consignees will be reported as unaccounted for. Product returned from the field will be scrapped.
Recall details
- Recall number
- Z-0097-2021
- Status
- Terminated
- Initiated by the company
- Sep 3, 2020
- Posted by the FDA
- Oct 9, 2020
- Recalling firm
- Biomet, Inc.
- Quantity
- 48 devices
- Product code
- JDI
- FDA submissions
- K051569
- Distribution
- Devices were distributed nationwide throughout the United States and to locations OUS. There has been one unit returned prior to the recall and approximately 14 units that have been previously implanted, leaving 33 units in the field subject to removal.
Product description
(1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458; Lot No. 710930; UDI No. (01) 00880304568747 (17) 230614 (10) 710930 (2) RingLoc Hip system Acetabular Bi-Polar Cup: Acetabular Cup, ArCom, 41 MM OD, 28 MM ID: Item No. 11-165206, Lot No. 649750; UDI No. (01) 00880304001923 (17) 230614 (10) 649750 (3) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Cup, ArCom, 52 MM OD, 28 MM ID: Item No. 11-165228; Lot No. 433290; UDI No. (01) 00880304001930 (17) 230612 (10) 433290
Code information (models, lots, serial numbers)
(1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458; Lot No. 710930; UDI No. (01) 00880304568747 (17) 230614 (10) 710930; EXP Date June 14th 2023 (2) RingLoc Hip system Acetabular Bi-Polar Cup: Acetabular Cup, ArCom, 41 MM OD, 28 MM ID: Item No. 11-165206, Lot No. 649750; UDI No. (01) 00880304001923 (17) 230614 (10) 649750; EXP Date June 14th 2023 (3) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Cup, ArCom, 52 MM OD, 28 MM ID: Item No. 11-165228; Lot No. 433290; UDI No. (01) 00880304001930 (17) 230612 (10) 433290; EXP Date June 12th 2023
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- RINGLOC BI-POLAR HIP SYSTEM 11-165206Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 recalls3 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
Product lines
Other devices under the same FDA submission
14 devices cleared under K051569 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RINGLOC BI-POLAR HIP SYSTEM 11-165218Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 recall2 deaths9 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165212Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 death12 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165226Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 death10 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165214Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 death9 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165232Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED8 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165224Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented7 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165222Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death6 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165208Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED6 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165216Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate5 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165210Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate4 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165230Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate3 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165236Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.