Skip to content

Covidien TA Auto Suture Vascular Loading Unit with DST… recallZ-0748-2021 · Covidien Llc · initiated Nov 23, 2020

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis

FDA root cause category: Process control

What the company is doing

Medtronic issued Urgent Medical Device Recall letter by letter delivered via Federal Express and/or certified mail beginning November 23, 2020. The letter states reason for recall, health risk and action to take: 1. Please immediately quarantine and discontinue use of affected item codes with associated lot numbers listed above. 2. Please return affected product as indicated below. All unused products from the affected item codes and associated lot numbers must be returned. 3. If you have distributed the Covidien TA" Auto Suture" Vascular Stapler and Loading Unit with DST Series" Technology listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory We request that you contact Medtronic if you experienced a quality problem or adverse event. Email Medtronic Post Market Vigilance at: [email protected] If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.

Recall details

Recall number
Z-0748-2021
Status
Completed
Initiated by the company
Nov 23, 2020
Recalling firm
Covidien Llc
Quantity
5316 units US; 1209 units OUS
Product code
GDW
FDA submissions
K111825
Distribution
Nationwide Foreign: Canada, Albania Armenia Australia Austria Belarus Belgium Bosnia And Herzegovina Cyprus Czech Republic Egypt Estonia France Germany Greece Hungary Iceland Israel Italy Japan Jordan Korea Kuwait Latvia Libyan Arab Jamahiriya Lithuania Malaysia Martinique Morocco Myanmar, Netherlands Norway Pakistan Poland Russian Federation Serbia Singapore Slovakia Spain Sweden Switzerland Taiwan Turkey United Arab Emirates United Kingdom Viet Nam

Product description

Covidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue. Item Number: TA30V3L

Code information (models, lots, serial numbers)

Lot Numbers: P0B1335MY, P0D0170MY, P0D1445M, 0F0377MY, P0B1336MY,P0D1444MY, P0E1211MY, P0F0606MY.  UDI: 10884521043152, 20884521043159

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

76 devices cleared under K111825 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 86886).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page