Penumbra System recallZ-0850-2021 · Penumbra Inc. · initiated Dec 15, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Penumbra Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
FDA root cause category: Process design
What the company is doing
On 12/15/2020, the firm posted a press release (https://www.penumbrainc.com/wp-content/uploads/2020/12/JET-7XF-15Dec20.pdf) and issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" letter via FedEx informing customers that all configurations of their reperfusion catheters may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization or contrast injection may result in potential vessel damage, and subsequent patient injury or death. Customers are instructed to: 1. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 2. Immediately review your inventory for the specific catalog numbers listed above. 3. Remove all units from inventory and place them in quarantine for return to Penumbra, Inc. 4. Complete and return the attached product identification / return form within 3 business days. 5. Continue to report adverse events and quality problems experienced with the use of this product. Adverse events and quality problems may be reported to FDA s MedWatch Adverse Event Reporting Program either online, by regular mail, or by fax. For more information, please see https://www.accessdata.fda.gov/scripts/medwatch/index.cfm. The Recalling Firm will contact customers directly to arrange return of affected units and replace/refund returned product. If customers have any questions or concerns, they can contact the Recalling Firm's Customer Service ([email protected] or 1.888.272.4606), available Monday - Friday 7:30 AM to 4:00 PM PST, or their sales representative.
Recall details
- Recall number
- Z-0850-2021
- Status
- Terminated
- Initiated by the company
- Dec 15, 2020
- Posted by the FDA
- Jan 18, 2021
- Recalling firm
- Penumbra Inc.
- Quantity
- 0 catheters
- Product code
- NRY
- FDA submissions
- K190010, K202251, K191946
- Distribution
- Worldwide distribution: U.S. (nationwide) including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO , MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. And O.U.S (countries) of: Canada, Chile, Colombia, Costa Rica, Dominican Republic, Hong Kong, Israel, Japan, Korea, Lebanon, Malaysia, Nepal, New Zealand, Oman, Panama, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, United Arab Emirates, and Vietnam.
Product description
Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7L138KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Code information (models, lots, serial numbers)
Product Name: Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing Catalog Number: 5MAXJET7L138KIT UDI Code: 00815948021594 Lot Number Range: F97123 thru F98981 Note - none were distributed
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Penumbra System 5MAXJET7L138KIT-BPenumbra, Inc. · Catheter, Thrombus Retriever1 recall0 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K190010, K202251, K191946 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Penumbra System J7XKIT132-APenumbra, Inc. · Catheter, Thrombus Retriever1 recall3 reports
- RED 72 RED72SDKIT-APenumbra, Inc. · Catheter, Thrombus Retriever3 deaths29 reports
- Penumbra System J7XBKIT132-APenumbra, Inc. · Catheter, Thrombus Retriever1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 86989).
- Z-0848-2021 · TerminatedPenumbra JET 7 Reperfusion Catheter REF 5MAXJET7 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7…
- Z-0849-2021 · TerminatedPenumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) i…
- Z-0851-2021 · TerminatedPenumbra JET 7MAX KIT Penumbra JET 7 Reperfusion Catheter+ MAX Delivery Device + Hi-Flow Aspiration Tubing REF J7XKIT132 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page