Medical device recalls: January 2021
The FDA posted 67 medical device recall records in January 2021, from 14 recall events by 12 companies. 12 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Fresenius Medical Care Holdings, Inc. · 2 events, 30 records
- Boston Scientific Corporation · 2 events, 7 records, 7 Class I
- Respironics California, LLC · 1 event, 16 records
- Penumbra Inc. · 1 event, 4 records, 4 Class I
- Ortho-Clinical Diagnostics, Inc. · 1 event, 2 records
- Roche Diagnostics Operations, Inc. · 1 event, 2 records
- Biomet, Inc. · 1 event, 1 record
- Hill-Rom, Inc. · 1 event, 1 record, 1 Class I
FDA root cause
- Under Investigation by firm · 31
- Device Design · 28
- Process design · 4
- Software design · 2
- Radiation Control for Health and Safety Act · 1
- Employee error · 1
Class I recalls (12)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0925-2021 · Hill-Rom, Inc. · posted Jan 29, 2021 · Terminated
Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the rail system, potentially resulting in the fall of both the lift motor and the patient
- Z-0879-2021 · Boston Scientific Corporation · posted Jan 19, 2021 · Terminated
There is potential for fractures which results in the inability to delivery therapy.
- Z-0848-2021 · Penumbra Inc. · posted Jan 18, 2021 · Terminated
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
- Z-0849-2021 · Penumbra Inc. · posted Jan 18, 2021 · Terminated
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
- Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7L138KITClass IZ-0850-2021 · Penumbra Inc. · posted Jan 18, 2021 · Terminated
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
- Z-0851-2021 · Penumbra Inc. · posted Jan 18, 2021 · Terminated
The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.
- Z-0728-2021 · Boston Scientific Corporation · posted Jan 9, 2021 · Terminated
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
- Z-0729-2021 · Boston Scientific Corporation · posted Jan 9, 2021 · Terminated
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
- Z-0730-2021 · Boston Scientific Corporation · posted Jan 9, 2021 · Terminated
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
- Z-0734-2021 · Boston Scientific Corporation · posted Jan 9, 2021 · Terminated
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
- Z-0735-2021 · Boston Scientific Corporation · posted Jan 9, 2021 · Terminated
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
- Z-0736-2021 · Boston Scientific Corporation · posted Jan 9, 2021 · Terminated
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
Class II recalls (55)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0727-2021 · Philips Home Healthcare Solution · posted Jan 28, 2021 · Terminated
After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.
- Yukon Polyaxial Screw Size 4.0x26mm (catalog number 7601-04026), component of YUKON Spinal SystemClass IIZ-0931-2021 · K2M, Inc · posted Jan 27, 2021 · Terminated
The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).
- Z-0902-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0903-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0904-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0905-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0906-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0907-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0908-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0909-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0910-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0911-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0912-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0913-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0914-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0915-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0916-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0917-2021 · Respironics California, LLC · posted Jan 22, 2021 · Open, Classified
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting…
- Z-0882-2021 · Biomet, Inc. · posted Jan 21, 2021 · Terminated
The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
- Z-0865-2021 · Nihon Kohden America Inc · posted Jan 14, 2021 · Open, Classified
Medical device non-conformance to electrical safety standard (IEC 60601-1).
- Z-0811-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0812-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0813-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0814-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0815-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0816-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0817-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0818-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0819-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0820-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0821-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0822-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0823-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0824-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0825-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0826-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0827-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0828-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0829-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
- Z-0831-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0832-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0833-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0834-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0835-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0836-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0837-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0838-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0839-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0840-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0841-2021 · Fresenius Medical Care Holdings, Inc. · posted Jan 11, 2021 · Open, Classified
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Z-0800-2021 · Ortho-Clinical Diagnostics, Inc. · posted Jan 7, 2021 · Terminated
A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.
- Z-0801-2021 · Ortho-Clinical Diagnostics, Inc. · posted Jan 7, 2021 · Terminated
A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.
- Z-0803-2021 · Roche Diagnostics Operations, Inc. · posted Jan 7, 2021 · Terminated
Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
- Z-0804-2021 · Roche Diagnostics Operations, Inc. · posted Jan 7, 2021 · Terminated
Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
- Z-0770-2021 · Ortho Clinical Diagnostics Inc · posted Jan 5, 2021 · Open, Classified
VITROS Chemistry Products CKMB Slides Potential for Falsely Elevated Results Due to CK-MM Interference