Skip to content

Captivator Single-Use Polypectomy Snares, Captiflex… recallZ-1168-2021 · Boston Scientific Corporation · initiated Jan 19, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.

FDA root cause category: Process change control

What the company is doing

On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.

Recall details

Recall number
Z-1168-2021
Status
Open, Classified
Initiated by the company
Jan 19, 2021
Recalling firm
Boston Scientific Corporation
Quantity
335,788 units total (Domestic: 222,882, Foreign: 112,906)
Product code
FDI
FDA submissions
K202478
Distribution
US nationwide distribution.

Product description

Captivator Single-Use Polypectomy Snares, Captiflex Single-Use Polypectomy Snares. Outer box UPN, Inner Pouch UPN: M00561311 (Box 10), M00561310 M00562301 (Box 10), M00562300; M00562321 (Box 10), M00562320; M00562341 (Box 10), M00562340; M00562451 (Box 10), M00562450; M00562401 (Box 10), M00562400; M00562402 (Box 40), M00562400; M00562422 (Box 40), M00562420; M00562471 (Box 10), M00562470.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Outer box GTIN: 08714729019251 Inner pouch GTIN: 08714729747680 Lot/Batch#: 24725319, 24782695,   Outer box GTIN: 08714729019312 Inner pouch GTIN: 08714729747833 Lot/Batch#: 24842597, 24812577,   Outer box GTIN: 08714729019336 Inner pouch GTIN: 08714729747857 Lot/Batch#: 24889757, 24719202, 24709536, 24773287,   Outer box GTIN: 08714729019350 Inner pouch GTIN: 08714729747871 Lot/Batch#: 24782693, 24732964, 24889755, 24780939,   Outer box GTIN: 08714729071068 Inner pouch GTIN: 08714729747970 Lot/Batch#: 24882357, 24782689, 24782687, 24729758,   Outer box GTIN: 08714729019411 Inner pouch GTIN:08714729747932 Lot/Batch#: 24961643  Outer box GTIN:08714729501640 Inner pouch GTIN: 08714729747932 Lot/Batch#: 24961645, 24989260, 24844193,   Outer box GTIN:08714729501657 Inner pouch GTIN:08714729747956 Lot/Batch#: 25010054,   Outer box GTIN:08714729019459 Inner pouch GTIN:08714729747994 Lot/Batch#: 25016793.

Devices named in this recall

8 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 87221).