MAGNUS operating table system recallZ-1296-2021 · Getinge Group Logistics America, LLC · initiated Feb 11, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Getinge Group Logistics America, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.
FDA root cause category: Device Design
What the company is doing
In February, 2021, Maquet GmbH/Getinge, will issue an Urgent Medical Device Correction notice to customers via FedEx 2 Day Delivery for the Magnus 1180 Operating Table due to a potential for hydraulic leakage to occur at the inclination function which may result in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. If this were to occur, it may lead to the patient having to be transferred to another OR table causing a delay in the procedure.
Recall details
- Recall number
- Z-1296-2021
- Status
- Open, Classified
- Initiated by the company
- Feb 11, 2021
- Recalling firm
- Getinge Group Logistics America, LLC
- Quantity
- 4288 units
- Product code
- FQO
- Distribution
- Domestic: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. Foreign: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Luxembourg, Lybia, Macedonia, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom, Uzbekistan, Venezuela, Vietnam and Yemen.
Product description
Operating Table System
Code information (models, lots, serial numbers)
Part Number: 118001A0, 118001B0, 118001B1, 118001B2, 118001B3, 118001B4, 118001C0, 118001D0. UDI code: 04046768031500, 04046768031531, 04046768040700, 04046768088351, 04046768088061, 04046768124752, 04046768031555, 04046768031579. Serial Number: 1 3402, 1 1028, 1 1264, 1 401, 3 74, 1 81, 2 1159, 2 - 762.
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- MAGNUS operating table system 118001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall17 reports
- MAGNUS operating table system 118001A0MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall6 reports
- MAGNUS operating table system 118001C0MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall5 reports
- MAGNUS operating table system 118001B1MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall4 reports
- MAGNUS operating table system 118001B4MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall4 reports
- MAGNUS operating table system 118001D0MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall2 reports
- MAGNUS operating table system 118001B0MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall1 report
- MAGNUS operating table system 118001B3MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page