MAGNUS operating table system 118001B0Recalls & reported problems
MAQUET GmbH · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046768031531), received between Nov 16, 2023 and Nov 16, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Positioning Failure” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Feb 11, 2021Open, Classified
There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.
Z-1296-2021 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 16, 2023Positioning Failure; Positioning Failure
ON 7TH NOVEMBER 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR COLUMNS - 118001B0 - MAGNUS OP TABLE COLUMN - SURFACE MOUNTED. AS IT WAS STATED, THE TABLE COULD NO LONGER MAKE CERTAIN MOVEMENTS DURING THE PROCEDURE DUE TO MALFUNCTION OF HYDRAULIC PUMP LEADING TO TERMINATION OF THE CURRENT OPERATION AND CANCELLATION OF THE REST OF THE SURGICAL PROGRAM. THE TABLE WAS MALFUNCTIONING DURING "UP" MOVEMENT AND IN THE TRANSLATION POSITION ON THE LEG SIDE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY INABILITY TO POSITION PATIENT AS REQUIRED DUE TO MALFUNCTION OF HYDRAULIC PUMP LEADING TO TERMINATION OF THE SURGERY, WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18148987 ↗
Other TABLE, OPERATING-ROOM, AC-POWERED devices
- ORT300 TableIMRIS Imaging, Inc. · Table, Operating-Room, Ac-Powered2 recalls21 reports
- ORT200 TableIMRIS Imaging, Inc. · Table, Operating-Room, Ac-Powered2 recalls12 reports
- MEERA mobile operating table 720001F2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- MEERA mobile operating table 720001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall19 reports
- MAGNUS operating table system 118001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.