Signia recallZ-1385-2021 · Covidien, LP · initiated Mar 4, 2021
Class II Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
Covidien/Medtronic issued Urgent Medical Device Recall via Federal Express and/or certified mail beginning March 4, 2021, states reason for recall, health risk and action to take: 1. Please immediately quarantine and discontinue use of affected item codes with associated lot numbers listed on Attachment A. 2. Please return affected product as indicated below. All unused products from the affected item codes and associated lot numbers must be returned. 3. If you have distributed the Covidien Signia" small diameter curved tip intelligent reload listed on Attachment A, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Recall details
- Recall number
- Z-1385-2021
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 4, 2021
- Recalling firm
- Covidien, LP
- Quantity
- 15298 units WW: 9204 US; 6094 OUS
- Product code
- CDW
- FDA submissions
- K191070
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Product description
Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short Item Code:SIGSDS30CTVT
Code information (models, lots, serial numbers)
Lot Numbers: N0A0274UY N0D0037UY N0G0491UY N0K1016UY N0A0897UY N0D0391UY N0G0736UY N0L0077UY N0B0640UY N0E0175UY N0J0510UY N0M0063UY N0B0937UY N0E0802UY N0J0680UY N0M0259UY N0C0144UY N0F1055UY N0J0999UY N9K1094UY N0C0486UY N0G0185UY N0K0584UY N9L0795UY N0C1115UY N0G0250UY N0K0805UY UDI: 10884521741874 20884521741871
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Signia SIGSDS30CTVCovidien LP · Staple, implantable1 recall1 death80 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K191070 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Signia SIGSDS30CTVTCovidien LP · Staple, implantable2 recalls1 death152 reports
- Signia SIGSDL45CTVTCovidien LP · Staple, implantable2 recalls27 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87475).
- Z-1383-2021 · TerminatedCovidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT
- Z-1384-2021 · TerminatedCovidien Signia Grey Small Diameter Curved Tip Intelligent Reload 30 mm Vascular 8 mm - Short Item Code: SIGSDS30CTV
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page