Skip to content

Signia SIGSDS30CTVTRecalls & reported problems

Covidien LP · Small Diameter Curved Tip Intelligent Reload Vascular/Thin; For use with 8 mm Cannula or larger · Class 2 · Implant

All models of Signia (16)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Covidien Signia stapler reload is a cartridge used in surgical stapling. Its 151 adverse event reports include 31 injuries and 1 death; many describe staplers that fail to fire or release tissue, with some bleeding. Of its 2 recalls, 1 is open and concerns reloads that can move uncontrollably and disrupt the staple line; the other, closed, concerns a safety feature.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
150
Deaths reported
1
Injuries reported
32
Recalls naming it
2

Summary

The FDA's MAUDE database contains 150 adverse event reports that list this exact device (DI 10884521741881), received between Jun 9, 2020 and Aug 7, 2026 — classified as 1 death, 32 injuries, 117 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (36 reports).

2 FDA recall records name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K191070, K253657); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 6 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 8 reportsJul 2024: 6 reportsAug 2024: 0 reportsSep 2024: 6 reportsOct 2024: 4 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 5 reportsJul 2025: 8 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 14, 2024Class IIOpen, Classified

    Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

    Z-1457-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Mar 4, 2021Class IITerminated

    Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

    Z-1384-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem54
  • Failure to Fire36
  • Entrapment of Device27
  • Failure to Form Staple24
  • Misfire10
  • Difficult to Open or Close9
  • Failure to Cut8
  • Mechanics Altered6

Patient problems

  • No Clinical Signs, Symptoms or Conditions75
  • Hemorrhage/Blood Loss/Bleeding43
  • Unspecified Tissue Injury17
  • Tissue Breakdown15
  • Insufficient Information10
  • Laceration(s)7
  • No Consequences Or Impact To Patient6
  • Blood Loss5

Recent adverse event reports

The 25 most recent of 150 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 7, 2026Adverse Event Without Identified Device or Use Problem

    According to the reporter, during a video-assisted thoracoscopic surgery (VATS), bleeding occurred from a pulmonary artery (PA) branch. The reload stapled completely across the pulmonary artery, resulting in additional bleeding, conversion to open surgery, and the need for a large incision.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Laceration(s)

    MAUDE report 50169111 ↗
  • MalfunctionMay 21, 2026Entrapment of Device

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC APPENDECTOMY, FOLLOWING A RELOAD, STAPLER FIRED STAPLES BUT WOULD NOT RELEASE THE TISSUE. FOLLOWING SEVERAL ATTEMPTS, THE SURGEON WAS UNABLE TO GET THE STAPLER TO RELEASE THE TISSUE. MANUAL RELEASE KEY WAS USED TO OPEN THE JAWS. STAPLE LINE APPEARED TO BE INTACT FOLLOWING MANUAL RELEASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25238832 ↗
  • MalfunctionMay 5, 2026Difficult to Open or Close

    ACCORDING TO THE REPORTER, DURING THE PROCEDURE, WHEN SETTING UP, WHILE CHECKING THE OPENING AND CLOSING DURING CONNECTION, THERE WAS DIFFICULTY OPENING THE JAW AND THE JAW WOULD NOT OPEN AT ALL. THE PRODUCT WAS REPLACED WITH ANOTHER NEW SIMILAR CARTRIDGE. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT FOUND THAT THERE WAS STAPLING - INCOMPLETE RETRACTION. THE JAWS OF THE RELOAD WERE CLAMPED. THE I BEAM WAS SLIGHTLY ADVANCED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25086745 ↗
  • MalfunctionMay 4, 2026

    FOLLOWING A RELOAD, STAPLER FIRED STAPLES BUT WOULD NOT RELEASE THE TISSUE. FOLLOWING SEVERAL ATTEMPTS, SURGEON WAS UNABLE TO GET THE STAPLER TO RELEASE THE TISSUE. MANUAL RELEASE KEY WAS USED TO OPEN THE JAWS. STAPLE LINE APPEARED TO BE INTACT FOLLOWING MANUAL RELEASE.

    Patient problems: Insufficient Information

    MAUDE report 25067404 ↗
  • MalfunctionApr 1, 2026Retraction Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY TO CUT THE APPENDIX, THE RELOAD FIRED COMPLETELY BUT WOULD NOT RETRACT THE KNIFE BLADE. STAFF REPORTEDLY CRANKED THE SCREWDRIVER THE WRONG WAY FOR A LONG TIME WHILE TRYING TO DETERMINE WHAT TO DO, AND THE PROCEDURE WAS EXTENDED BY ABOUT 30 MINUTES. A MANUAL RETRACTION TOOL WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24760827 ↗
  • MalfunctionMar 19, 2026Entrapment of Device

    IT WAS REPORTED THAT DURING THE THORACIC SEGMENTECTOMY, AFTER FIRING ON THE LUNG ARTERY, THE STAPLER LOCKED ON TISSUE AND COULD NOT BE REMOVED. THE SURGEON WAS ABLE TO OPEN THE STAPLER AFTER STRUGGLING, AND THE HANDLE WAS REPLACED WITH A NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24644411 ↗
  • MalfunctionFeb 18, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A PROCEDURE, WHILE PREPARING THE RELOAD, IT WOULD NOT FIT AND KEPT COMING OFF. TO RESOLVE THE ISSUE, A HANDLE AND WHITE RELOAD WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24378928 ↗
  • MalfunctionJan 29, 2026Entrapment of Device

    ACCORDING TO THE REPORTER, DURING THE VIDEO ASSISTED THORACOSCOPIC SURGERY, THE JAWS DID NOT OPEN AFTER STAPLING THE BLOOD VESSEL. A DDITIONALLY, AFTER WAITING FOR A SHORT TIME, THE JAW EVENTUALLY OPENED SO NO TREATMENT INTERVENTION WAS PERFORMED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24218208 ↗
  • MalfunctionJan 16, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A PULMONARY LOBECTOMY USING THE POWERED STAPLER, AFTER PRESSING THE GREEN BUTTON, THE DEVICE DID NOT FIRE DESPITE ALL PREPARATIONS BEING COMPLETED. THE RELOAD WAS REPLACED WITH ANOTHER SMALL DIAMETER RELOAD, WHICH WAS FIRED WITHOUT ANY ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24094902 ↗
  • MalfunctionJan 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A LIVING TRANSPLANT PROCEDURE¿SPECIFICALLY DURING THE RESECTION OF THE DONOR'S RENAL ARTERY¿THE SUTURE LINE BLED, REQUIRING THE APPLICATION OF ADDITIONAL REFILLS AND CLIPS. PROLONGED SURGICAL TIME WAS NOTED BUT NOT MORE THAN 30 MINUTES. ADDITIONALLY, THE HANDLE MADE A NOISE WHEN ROTATED. TO ADDRESS THE SITUATION, THE SURGICAL TEAM PERFORMED RESECTION AND RE-STAPLING.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown

    MAUDE report 24034799 ↗
  • MalfunctionDec 30, 2025Difficult to Open or Close; Physical Resistance/Sticking; Difficult to Open or Close; Physical Resistance/Sticking

    ACCORDING TO THE REPORTER, DURING A VIDEO-ASSISTED THORACOSCOPIC SURGERY (VATS) LOBECTOMY, WHILE ON BLOOD VESSEL, THE ANGLE OF THE JAW WAS ADJUSTED, BUT IT WAS NOT AT THE DESIRED ANGLE, SO IT WAS REMOVED ONCE. THE ANGLE OF THE JAW WAS ADJUSTED OUTSIDE THE SURGICAL FIELD, AND AN ATTEMPT WAS MADE TO CLOSE THE JAW, BUT IT DID NOT CLOSE. A NEW RELOAD WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23926209 ↗
  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE RELOAD WAS USED ON ONE OF THE BRANCHES OF THE PULMONARY ARTERY. AFTER FIRING, STAPLE LINE BLEEDING WAS OBSERVED, AND TWO ADDITIONAL METAL CLIPS WERE REQUIRED TO ACHIEVE HEMOSTASIS. THE SURGICAL TIME WAS EXTENDED FOR ABOUT 30 TO 45 MINUTES, AND THE PATIENT REQUIRED REOPERATION FOR HEMOSTASIS. THERE WAS BLOOD LOSS OF LESS THAN 500CC.

    Patient problems: Tissue Breakdown

    MAUDE report 23691847 ↗
  • InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE POSTOPERATIVE VIDEO-ASSISTED THORACOSCOPIC SURGERY, EXCESSIVE AND SEVERE STAPLE LINE BLEEDING OCCURRED AT THE ARTERIAL BRANCH OF SEGMENT 6 OF THE LEFT LUNG. THE ISSUE WAS IDENTIFIED WHILE THE PATIENT WAS STILL IN THE OPERATING ROOM DURING THE EXTUBATION. THE PATIENT BEGAN WAKING UP QUICKLY, PROMPTING THE SURGICAL TEAM TO PUT THE PATIENT BACK TO SLEEP TO ADDRESS THE BLEEDING. THE PROCEDURE WAS THEN CONVERTED FROM THORACOSCOPIC TO OPEN SURGERY. AN ADDITIONAL SURGICAL OPERATION WAS REQUIRED TO CONTROL THE BLEEDING, DURING WHICH THE SURGEON MANUALLY SUTURED AND CLOSE THE AFFECTED ARTERY. AS A RESULT, THE SURGICAL TIME WAS EXTENDED BY ONE HOUR. IT WAS NOTED THAT THE PATIENT HAD BLOOD LOSS OF 2000-2500CC AND REQUIRED BLOOD TRANSFUSION.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Laceration(s)

    MAUDE report 23344427 ↗
  • MalfunctionSep 19, 2025Unintended System Motion; Unintended System Motion

    IT WAS REPORTED THAT DURING THE ROBOTIC LAPAROSCOPIC THORACIC WEDGE RESECTION, WHILE STAPLING THE PULMONARY ARTERY, THE RELOAD ARTICULATED/STRAIGHTENED DURING FIRING. IT WAS NOTED THAT THE RELOAD ARTICULATED IN ONE DIRECTION WHILE THE KNIFE BLADE WAS ON RETRACTION. NO INTERVENTION WAS DONE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23096274 ↗
  • MalfunctionAug 27, 2025Failure to Fire; Difficult to Open or Close; Expulsion; Failure to Fire; Difficult to Open or Close; Expulsion

    ACCORDING TO THE REPORTER, DURING AN UPPER LOBE RESECTION, THERE WAS NO ISSUE DURING THE FIRST AND SECOND FIRINGS, BUT AT THE TIME OF LOADING FOR THE THIRD FIRING, THE SLIGHT OPENING AND CLOSING DID NOT OPERATE ACCURATELY, AND THE OPENING AND CLOSING OPERATION WAS STIFF. AFTER COMPLETING THE OPENING AND CLOSING CHECK OF THE JAWS, THE BLOOD VESSEL WAS APPROACHED AND FIRING WAS ATTEMPTED, BUT FIRING COULD NOT BE PERFORMED. THE TIP SIDE OF THE CARTRIDGE WAS OPENED AND COLLECTED; THE HANDLE AND CARTRIDGE WERE BOTH REPLACED, AND WHEN AN APPROACH WAS MADE AGAIN, THE RESECTION WAS PERFORMED WITHOUT ANY PROBLEM. AFTER THE SURGERY, THE CARTRIDGE WAS CHECKED, AND A SLIGHT/FEW PRE-FIRINGS WERE FOUND AT THE PROXIMAL SIDE OF THE CHANNEL JAWS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903464 ↗
  • MalfunctionJul 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER DURING A RIGHT UPPER SEGMENTECTOMY, THE POWERED STAPLER WAS FIRED AND RETRACED WITHOUT AN ISSUE. HOWEVER, THE STAPLE LINE HAD A SMALL SPURTING BLEED. THE BLEEDING WAS RESOLVED BY USING A HORIZONTAL CLIP.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22636740 ↗
  • MalfunctionJul 21, 2025Entrapment of Device; Entrapment of Device

    ACCORDING TO THE REPORTER, DURING A LIVING DONOR KIDNEY TRANSPLANTATION, AT THE TIME OF RENAL ARTERY RESECTION, AFTER THE FIRING WAS COMPLETED, THE JAW DID NOT OPEN. MOREOVER, THE JAW NATURALLY OPENED, AND IT WAS ABLE TO BE REMOVED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22555707 ↗
  • MalfunctionJul 17, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A VIDEO-ASSISTED THORACOSCOPIC SURGERY (VATS) LOBECTOMY CASE, THE RELOAD COULD NOT BE FIRED BY THE MANUAL STAPLER. THE SURGEON WAS ABLE TO PRESS THE GREEN SAFETY BUTTON BUT WAS NOT ABLE TO SQUEEZE THE HANDLE. THE ISSUE WAS RESOLVED BY REPLACING THE RELOAD WITH A NEW ONE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22532683 ↗
  • MalfunctionJul 16, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING THE DONOR NEPHRECTOMY, THE RELOAD WAS POSITIONED ONTO THE VEIN, CLAMPED, AND MOST OF THE STAPLES HAVE BEEN DEPLOYED AND RELEASED FROM THE STAPLER, HOWEVER, IN THE SECTION WHERE THE VEIN WAS, THE STAPLES WERE STILL IN THE JAWS OF THE STAPLE, FORMED BUT NOT RELEASED. THIS RESULTED IN THE VEIN BEGINNING TO SPURT BLOOD AS IT WAS NOT SEALED, AND THE DISTAL STAPLE LINE WAS INCOMPLETE. THE PROCEDURE WAS EXTENDED BY FIVE MINUTES. TO MANAGE THE BLEEDING, TWO HEMOLOCKS WERE PLACED ON THE VEIN TO SEAL IT AND ADDITIONAL RELOAD WAS USED TO COMPLETE THE CASE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown

    MAUDE report 22527316 ↗
  • MalfunctionJul 11, 2025Failure to Fire; Failure to Fire

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE FIRING COULD NOT BE PERFORMED. ANOTHER RELOAD WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22477346 ↗
  • DeathJul 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT WHEN CLAMPING A SEGMENT OF THE LEFT PULMONARY ARTERY (A1/3), BLEEDING OCCURRED AFTER REMOVAL OF THE MAGAZINE AT THE PROXIMAL CLAMP LINE, WHICH RAPIDLY INCREASED AND NECESSITATED AN EMERGENCY THORACOTOMY. DUE TO MASSIVE BLEEDING WITH CARDIAC ARREST, THE PATIENT DIED.

    Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding; Laceration(s)

    MAUDE report 22466837 ↗
  • MalfunctionJul 8, 2025Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT DURING A DONOR NEPHRECTOMY USING A POWERED STAPLER, THE RELOAD WAS ATTACHED BY THE NURSING TEAM, AND THE SURGEON PLACED IT ON THE ARTERY AND CLOSED THE JAWS AND WHEN ATTEMPTING TO REOPEN THEM FOR REPOSITIONING, THE JAWS WOULD NOT OPEN, CAUSING THE VESSEL TO BE STUCK AND DELAYING THE PROCEDURE BY FIVE MINUTES. THE VESSEL WAS EVENTUALLY RELEASED, AND A STAPLER FROM ANOTHER COMPANY WAS USED TO COMPLETE THE FIRING. A DIFFERENT RELOAD WAS LATER USED SUCCESSFULLY, AND NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22446635 ↗
  • MalfunctionJul 2, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT THE DURING A PROCEDURE, USING A CURVE TIP RELOAD, THE MANUAL HANDLE DID NOT FIRE. THE HANDLE WAS ABLE TO BE SQUEEZED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22390766 ↗
  • MalfunctionJun 24, 2025Failure to Cut; Failure to Cut

    ACCORDING TO THE REPORTER, DURING A RIGHT THORACOSCOPIC UPPER LOBE RESECTION, TWO CARTRIDGES MALFUNCTIONED. THE FIRST UNIT SUCCESSFULLY STAPLED THE PULMONARY BLOOD VESSEL BUT FAILED TO CUT, RESULTING IN INCOMPLETE RESECTION. THE SECOND UNIT ALLOWED THE KNIFE TO DEPLOY WITHOUT PROPER STAPLING, CAUSING INTRAOPERATIVE BLEEDING. HEMOSTASIS WAS ACHIEVED USING A HEMOSTATIC SHEET.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22318997 ↗
  • MalfunctionJun 24, 2025Mechanics Altered; Mechanics Altered

    ACCORDING TO THE REPORTER DURING A RIGHT THORACOSCOPIC UPPER LOBE RESECTION, TWO CARTRIDGES MALFUNCTIONED. THE FIRST UNIT SUCCESSFULLY STAPLED THE PULMONARY BLOOD VESSEL BUT FAILED TO CUT, RESULTING IN INCOMPLETE RESECTION. THE SECOND UNIT ALLOWED THE KNIFE TO DEPLOY WITHOUT PROPER STAPLING, CAUSING INTRAOPERATIVE BLEEDING. HEMOSTASIS WAS ACHIEVED USING A HEMOSTATIC SHEET.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22318380 ↗

FDA clearance and classification

Cleared through 510(k) K253657 on Apr 10, 2026 (decision: Substantially Equivalent). Submitted by Covidien (Part of Medtronic) as “Tri-staple 2.0™ Reloads; Endo GIA™ Reloads with Tri-Staple™ Technology; Endo GIA™ Gray Articulating Reloads; Signia™ Small Diameter Reloads”. The FDA review took 141 days. FDA record ↗

Cleared through 510(k) K191070 on Sep 4, 2019 (decision: Substantially Equivalent). Submitted by Covidien as “Signia Small Diameter Reloads”. The FDA review took 135 days. FDA record ↗

The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K191070, K253657. One submission often covers many models, so check the product and lot details in each record.

  • Z-0485-2026 · Sep 25, 2025 · Open, Classified
    Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
  • Z-0486-2026 · Sep 25, 2025 · Open, Classified
    Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
  • Z-1458-2024 · Feb 14, 2024 · Open, Classified
    Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
  • Z-1383-2021 · Mar 4, 2021 · Terminated
    Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT
  • Z-1385-2021 · Mar 4, 2021 · Terminated
    Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short Item Code:SIGSDS30CTVT

Common questions about the Signia SIGSDS30CTVT

Has the Signia SIGSDS30CTVT been recalled?

Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-1457-2024 (Class II), began on Feb 14, 2024: Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Signia SIGSDS30CTVT?

The FDA's MAUDE database lists 150 adverse event reports for this device, received between Jun 9, 2020 and Aug 7, 2026. In those reports, the most common specific device problems are “Failure to Fire” (36), “Entrapment of Device” (27) and “Failure to Form Staple” (24); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (43), “Unspecified Tissue Injury” (17) and “Tissue Breakdown” (15). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Signia SIGSDS30CTVT?

1 report is classified as a death, 32 as injuries and 117 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Signia SIGSDS30CTVT FDA approved?

It was cleared through 510(k) K191070 on Sep 4, 2019; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Signia SIGSDS30CTVT MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Signia SIGSDS30CTVT, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand Signia, model SIGSDS30CTVT). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Staple, implantable devices

All Staple, implantable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page