ADVIA 2120i Refurb Single Aspirate (SMN 11314044) - US… recallZ-1392-2021 · Siemens Healthcare Diagnostics, Inc. · initiated Mar 5, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential Sample Identification (SID) Mismatch with14-Character Barcodes
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter HSW21-01.A.US to US customers on March 5, 2021 via FedEx overnight delivery. An Urgent Field Safety Notice (UFSN) HSW21-01.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities. Letter states reason for recall, health risk and action to take: with your Medical Director. If you are using 14-character Sample Identification (SID) barcodes: o Ensure that the Barcode Encoding Selectivity feature is enabled. Open the Sample Identification screen to determine if the checkbox to turn the Barcode Encoded Selectivity is on o If Barcode Encoding Selectivity is disabled, please perform the following steps to enable Barcode Encoding Selectivity: Open the Sample Identification screen Select the checkbox to turn the Barcode Encoded Selectivity on Whether or not you have the Barcode Encode Selectivity feature disabled, please complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-1392-2021
- Status
- Open, Classified
- Initiated by the company
- Mar 5, 2021
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 1 unit
- Product code
- GKZ
- FDA submissions
- K162977, K102644
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bosnia Herzeg., Botswana, Brazil, Bulgaria, Burkina, Faso, Cambodia, Canada, Chile, Colombia, Croatia, Cura¿ao Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Fren.Polynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Mali Malta, Martinique, Mexico, Namibia, Netherlands, Norway, Oman, P.R. China, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad,Tobago, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Uruguay, Vatikancity, Vietnam.
Product description
ADVIA 2120i Refurb Single Aspirate (SMN 11314044) - US, automated hematology analyzer Software Versions 6.10 and 6.11
Code information (models, lots, serial numbers)
Software Versions 6.10 and 6.11 UDI: 00630414237879
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- C-SAA ANALYT MODULE ASSY 2120I 11314044Siemens Healthcare Diagnostics Inc. · Counter, differential cell1 recall0 reports
Other devices under the same FDA submission
5 devices cleared under K162977, K102644 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA® 2120i Hematology System 10285573Siemens Healthcare Diagnostics Inc. · Counter, differential cell1 recall28 reports
- ADVIA® 2120i Hematology System 10488923Siemens Healthcare Diagnostics Inc. · Counter, differential cell1 recall12 reports
- ADVIA 2120i with DAA Autosampler 11219529Siemens Healthcare Diagnostics Inc. · Counter, differential cell1 recall10 reports
- ADVIA 2120i with SAA Autosampler 11219530Siemens Healthcare Diagnostics Inc. · Counter, differential cell1 recall7 reports
- ADVIA 2120i Refurbished System Dual Aspirate 11314045Siemens Healthcare Diagnostics Inc. · Counter, differential cell1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87488).
- Z-1386-2021 · Open, ClassifiedADVIA 2120 with Single Aspirate Autosampler (SMN 10316162) - US, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1387-2021 · Open, ClassifiedADVIA 2120 Dual Aspirate Autosampler (SMN 10313419) - US, automated hematology analyzer -automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1388-2021 · Open, ClassifiedADVIA 2120 REFURB SAA AUTOSAMPLER (SMN 10374453) - US, automated hematology analyzer -automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1389-2021 · Open, ClassifiedADVIA 2120 REFURB DAA AUTOSAMPLER (SMN 10374454) - US, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1390-2021 · Open, ClassifiedADVIA 2120i with Single Aspirate Autosampler (SMN 10488923) - US, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1391-2021 · Open, ClassifiedADVIA 2120i Dual Aspirate Autosampler (SMN 10285573) - US, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1393-2021 · Open, ClassifiedADVIA 2120i Refurb Dual Aspirate (SMN 11314045) - US, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1394-2021 · Open, ClassifiedADVIA 2120i (RoHS) with Single Aspirate Autosampler (SMN 11219529) - US, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1395-2021 · Open, ClassifiedADVIA 2120i (RoHS) Dual Aspirate Autosampler (SMN 11219530) - US, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1396-2021 · Open, ClassifiedADVIA 2120 with Single Aspirate Autosampler (SMN 10316162) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1397-2021 · Open, ClassifiedADVIA 2120 Dual Aspirate Autosampler (SMN 10313419) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1398-2021 · Open, ClassifiedADVIA 2120 REFURB SAA AUTOSAMPLER (SMN 10374453) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1399-2021 · Open, ClassifiedADVIA 2120 REFURB DAA AUTOSAMPLER (SMN 10374454) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1400-2021 · Open, ClassifiedADVIA 2120i with Single Aspirate Autosampler (SMN 10488923) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1401-2021 · Open, ClassifiedADVIA 2120i Dual Aspirate Autosampler (SMN 10285573) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1402-2021 · Open, ClassifiedADVIA 2120i Refurb Single Aspirate (SMN 11314044) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1403-2021 · Open, ClassifiedADVIA 2120i Refurb Dual Aspirate (SMN 11314045) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1404-2021 · Open, ClassifiedADVIA 2120i (RoHS) with Single Aspirate Autosampler (SMN 11219529) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
- Z-1405-2021 · Open, ClassifiedADVIA 2120i (RoHS) Dual Aspirate Autosampler (SMN 11219530) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page