Skip to content

ADVIA 2120 REFURB DAA AUTOSAMPLER (SMN 10374454) - OUS… recallZ-1399-2021 · Siemens Healthcare Diagnostics, Inc. · initiated Mar 5, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential Sample Identification (SID) Mismatch with14-Character Barcodes

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter HSW21-01.A.US to US customers on March 5, 2021 via FedEx overnight delivery. An Urgent Field Safety Notice (UFSN) HSW21-01.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities. Letter states reason for recall, health risk and action to take: with your Medical Director. If you are using 14-character Sample Identification (SID) barcodes: o Ensure that the Barcode Encoding Selectivity feature is enabled. Open the Sample Identification screen to determine if the checkbox to turn the Barcode Encoded Selectivity is on o If Barcode Encoding Selectivity is disabled, please perform the following steps to enable Barcode Encoding Selectivity: Open the Sample Identification screen Select the checkbox to turn the Barcode Encoded Selectivity on Whether or not you have the Barcode Encode Selectivity feature disabled, please complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Recall details

Recall number
Z-1399-2021
Status
Open, Classified
Initiated by the company
Mar 5, 2021
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
43 units
Product code
GKZ
FDA submissions
K162977, K102644
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bosnia Herzeg., Botswana, Brazil, Bulgaria, Burkina, Faso, Cambodia, Canada, Chile, Colombia, Croatia, Cura¿ao Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Fren.Polynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Mali Malta, Martinique, Mexico, Namibia, Netherlands, Norway, Oman, P.R. China, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad,Tobago, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Uruguay, Vatikancity, Vietnam.

Product description

ADVIA 2120 REFURB DAA AUTOSAMPLER (SMN 10374454) - OUS, automated hematology analyzer Software Versions 6.10 and 6.11

Code information (models, lots, serial numbers)

Software Versions 6.10 and 6.11

Other devices under the same FDA submission

6 devices cleared under K162977, K102644 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 87488).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page