Alaris Infusion Pump 8100 Bezel (plastic piece only), Part:… recallZ-1332-2021 · Tenacore LLC · initiated Feb 25, 2021
Class I Terminated All recalls posted in April 2021
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Tenacore LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
FDA root cause category: Component design/selection
What the company is doing
On 02/25/2021, Field Safety Corrective Action notifications were mailed to customers; customers were also notified via phone. The recalling firm provided the following services to customers: serviced infusion pumps, sold infusion pumps, and sold bezel repair parts. Customers were asked to do the following: 1. Review the serialized list of potentially impacted pumps and remove these devices from clinical use and immediately quarantine them until they can be inspected, and impacted parts are replaced, if necessary. Part customers were told to quarantine affected parts. 2. Inspect the potentially affected devices to determine if the impacted bezel repair part is installed. Refer to inspection instructions and a) identify the impacted Alaris 8100 Infusion Pump Module by serial number; b) Look from the top down for the ejector pin marks, c) If NO ejector pin marks are identified, this module requires bezel. Customers who were sold pumps can identify affected pumps through serial number alone. 3. Dealers and Distributors only: If potentially impacted pumps were sold, immediately identify, inspect and recall, if necessary, the potentially affected pumps. An electronic copy of a proposed field safety corrective action notice will be provided that can to be used for the facilities that may be using the potentially affected pumps. 4. Contact the recalling firm to schedule bezel repair part replacement. The impacted bezel repair part must be replaced before the pump can be returned to service. 5. Please immediately share this field safety corrective action notification with all necessary parties within your organization to ensure that they are also aware of this action. Notify end user customers. 6. Complete and return the customer response form. Contact Quality Assurance at 714-371-9919 or [email protected] to schedule replacement of the impacted part.
Recall details
- Recall number
- Z-1332-2021
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Feb 25, 2021
- Posted by the FDA
- Apr 2, 2021
- Recalling firm
- Tenacore LLC
- Quantity
- 1494
- Product code
- FRN
- FDA submissions
- K133532
- Distribution
- US distribution to states of: WA, AZ, NY, TX, LA, CA, NV, ME, MN, DE, NC, FL, MO, KY, TN, and PA
Product description
Alaris Infusion Pump 8100 Bezel (plastic piece only), Part: TIPA-8100-4410
Code information (models, lots, serial numbers)
Parts distributed between July 10, 2020 and January 22, 2021. Time stamp at bezel rear side shows 5 or 6 SO Numbers: 271111, 263873, 264442, 264890, 268606, 269930, 270760, 270936, 271211, 272860, 267603, 268936, 269258, 269291, 272496, 272225, 271408, 267392
Other devices under the same FDA submission
4 devices cleared under K133532 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion1 recall2 deaths5,636 reports
- Alaris 8975Carefusion 303, Inc. · Infusion safety management software169 reports
- Alaris 9601Carefusion 303, Inc. · Pump, infusion1 recall0 reports
- Alaris™ Syringe Module, Model 8110 8100Carefusion 303, Inc. · Pump, infusion14 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 87470).
- Z-1331-2021 · TerminatedAlaris Infusion Pump Module, Model 8100, serviced/repaired by Tenacor with an impacted bezel repair part
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.