Medical device recalls: April 2021
The FDA posted 26 medical device recall records in April 2021, from 11 recall events by 11 companies. 2 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Smiths Medical ASD Inc. · 1 event, 12 records
- Nuvasive Specialized Orthopedics Inc · 1 event, 3 records
- Datascope Corp. · 1 event, 2 records
- Tenacore LLC · 1 event, 2 records, 2 Class I
- Alcon Research, LLC · 1 event, 1 record
- Arrow International Inc · 1 event, 1 record
- Howmedica Osteonics Corp. · 1 event, 1 record
- Integra LifeSciences Corp. · 1 event, 1 record
FDA root cause
- Process control · 12
- Software design · 3
- Device Design · 3
- Component design/selection · 2
- Vendor change control · 1
- Under Investigation by firm · 1
- Process design · 1
- Nonconforming Material/Component · 1
Class I recalls (2)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1331-2021 · Tenacore LLC · posted Apr 2, 2021 · Terminated
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free…
- Z-1332-2021 · Tenacore LLC · posted Apr 2, 2021 · Terminated
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free…
Class II recalls (21)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1518-2021 · Omnia SRL · posted Apr 29, 2021 · Terminated
Omnia SRL received a notification from the contract manufacturer informing them of inadequate sterilization processes which have been taking place at their contracted sterilizing facility since 2016.
- LogiCal Monitoring Set, Model Number DPS564898. for blood pressure monitoring and blood sampling. ↗Class IIZ-1470-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Mark 910 LogiCal, Model Number DPS567201. for blood pressure monitoring and blood sampling. ↗Class IIZ-1471-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1472-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1473-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1474-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1475-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Manifold rotator (Component): ↗Class IIZ-1476-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Stopcock rotator (Component): ↗Class IIZ-1477-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1478-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1479-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1480-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1481-2021 · Smiths Medical ASD Inc. · posted Apr 21, 2021 · Terminated
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring…
- Z-1434-2021 · Siemens Medical Solutions USA, Inc · posted Apr 19, 2021 · Terminated
Due to the configuration of certain Windows Service Permissions within the operating systems of the Sensis/ Sensis Vibe computer, there is a risk for exposure of sensitive information, manipulation of data, or Denial of Service attacks and could result…
- Z-1379-2021 · Howmedica Osteonics Corp. · posted Apr 9, 2021 · Terminated
The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction…
- Z-1374-2021 · Arrow International Inc · posted Apr 7, 2021 · Open, Classified
Difficulty inserting the biopsy ejector rod into the bone lesion biopsy cannula, preventing the biopsy sample from being collected. Use of a defective device is likely to require immediate repeat biopsy, leading to a prolonged procedure
- Z-1349-2021 · Nuvasive Specialized Orthopedics Inc · posted Apr 5, 2021 · Open, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
- Z-1350-2021 · Nuvasive Specialized Orthopedics Inc · posted Apr 5, 2021 · Open, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
- NUVASIVE FREEDOM:Class IIZ-1351-2021 · Nuvasive Specialized Orthopedics Inc · posted Apr 5, 2021 · Open, Classified
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
- Z-1354-2021 · Integra LifeSciences Corp. · posted Apr 5, 2021 · Terminated
Incorrect labeling, Catheter did not have the required holes and markings
- Z-1328-2021 · Alcon Research, LLC · posted Apr 1, 2021 · Terminated
Reports have been received concerning mold found on gowns included in custom ophthalmic surgery packages.
Class III recalls (3)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1506-2021 · Datascope Corp. · posted Apr 28, 2021 · Open, Classified
There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
- Z-1507-2021 · Datascope Corp. · posted Apr 28, 2021 · Open, Classified
There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
- Z-1409-2021 · OriGen Biomedical, Inc. · posted Apr 10, 2021 · Terminated
It was identified that tissue culture bags were incorrectly labeled as free of phthalates.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.