Surgiwand II recallZ-1411-2021 · Covidien · initiated Mar 5, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.
FDA root cause category: Process design
What the company is doing
On 3/10/2021, Medtronic issued an Urgent Medical Device Recall notices to customers. All consignees were notified by letter delivered via Federal Express and/or certified mail. The letter informs customers of the recall of specific item codes and production lots of Covidien Surgiwand" II Suction and Irrigation Device following reports from customers of foreign particles in the device tubing.
Recall details
- Recall number
- Z-1411-2021
- Status
- Open, Classified
- Initiated by the company
- Mar 5, 2021
- Recalling firm
- Covidien
- Quantity
- 59,828 devices
- Product code
- GEI
- FDA submissions
- K961771
- Distribution
- Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
Product description
Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094
Code information (models, lots, serial numbers)
UDI-Device Identifier (GTIN/UPN): 10884523000887, 20884523000884 UDI-Device Identifier (GTIN/UPN): 10884523000900, 20884523000907 UDI-Device Identifier (GTIN/UPN): 10884523000917, 20884523000914
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Surgiwand II 178093Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall83 reports
- Surgiwand II 178094Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall10 reports
- Surgiwand II 178083Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall5 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K961771 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Surgiwand II 178091Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES84 reports
- Surgiwand II 178099Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES11 reports
- Surgiwand II 178081Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page