Surgiwand II 178099Recalls & reported problems
Covidien LP · Suction and Irrigation Device with Spatula Tip · Class 2
All models of Surgiwand II (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10884523000931), received between Jan 18, 2018 and Mar 7, 2024 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K961771); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Detachment of Device or Device Component4
- Device Contamination with Chemical or Other Material2
- Melted2
- Device Slipped1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement2
- Insufficient Information1
- No Code Available1
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, THE SHAFT OF THE DEVICE WAS SEPARATED. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18859176 ↗MalfunctionMay 23, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, THE FRONT TIP PART OF THE DEVICE FELL OFF, NOT INTO THE PATIENT CAVITY. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16988890 ↗MalfunctionSep 28, 2022Melted; Melted
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE TIP OF THE DEVICE MELTED. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15498112 ↗MalfunctionAug 3, 2022Break; Break
ACCORDING TO THE REPORTER, DURING THE LAPAROSCOPIC ANTERIOR RESECTION, WHEN SUCTIONING OF FLUID, THE DISTAL TIP OF THE BLACK SHEATH GOT DAMAGED. A NEW PIECE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15161608 ↗MalfunctionMar 25, 2022Sparking; Sparking
ACCORDING TO THE REPORTER, THE ENERGY LEAKED FROM THE TIP.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 13900971 ↗MalfunctionApr 13, 2021Break; Break
ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE DEVICE'S DISTAL TIP OF THE BLACK SHEATH DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11661021 ↗InjuryMay 14, 2020Detachment of Device or Device Component; Device Contamination with Chemical or Other Material; Detachment of Device or Device Component; Device Contamination with Chemical or Other Material
ACCORDING TO THE REPORTER, DURING IRRIGATION OF ABDOMINAL CAVITY ON A LAPAROSCOPIC APPENDECTOMY, IT WAS NOTED THAT THERE WAS A FOREIGN OBJECT THAT CAME FROM THE DEVICE THAT FELL INTO PATIENT'S CAVITY.
Patient problems: No Code Available
MAUDE report 10058293 ↗MalfunctionMay 13, 2020Detachment of Device or Device Component; Device Contamination with Chemical or Other Material; Detachment of Device or Device Component; Device Contamination with Chemical or Other Material
ACCORDING TO THE REPORTER, DURING IRRIGATION OF ABDOMINAL CAVITY ON A LAPAROSCOPIC CHOLECYSTECTOMY, THERE WAS A FOREIGN UNIDENTIFIED MATERIAL FOUND ON THE DEVICE THAT FELL INTO THE PATIENT'S CAVITY AND WAS RETRIEVED BY A GRASPER. SURGEON USED ANOTHER DEVICE TO RESOLVE THE ISSUE.
Patient problems: No Information
MAUDE report 10053356 ↗MalfunctionAug 8, 2019Melted; Melted
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE TIP PART OF THE DEVICE WAS CHIPPED OR MELTED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8874470 ↗MalfunctionNov 20, 2018Break; Break
ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE DEVICE SHAFT WAS BURNT DOWN ABOUT 3CM.
Patient problems: No Patient Involvement
MAUDE report 8088865 ↗MalfunctionJan 18, 2018Break; Device Slipped; Break; Device Slipped
ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE TIP OF THE DEVICE WAS BURNED DURING ACTIVATING MONOPOLAR FUNCTION. THE BLACK SHEATH COVER LOOSE FROM THE SHAFT AND THE DISTAL TIP OF THE BLACK SHEATH WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7203364 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K961771. One submission often covers many models, so check the product and lot details in each record.
- Z-1411-2021 · Mar 5, 2021 · Open, ClassifiedCovidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page