VERIGENE Processor SP - Product Usage: intended for… recallZ-1488-2021 · Luminex Corporation · initiated Mar 18, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Luminex Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for a false-negative result for Vibrio parahaemolyticus, Salmonella, and/or Shigella spp when used with the VERIGENE Enteric Pathogens Nucleic Acid Test Kit.
FDA root cause category: Process control
What the company is doing
A customer advisory notice dated 03/18/2021. The Customer Advisory Notice recommends customers review their QC data to ensure passing results for all VERIGENE EP test targets, and to contact Luminex if any false negatives are observed for any targets within the past twelve months, in order to arrange for a replacement of the Amplification Heater. Additionally, if a Processor SP unit is indicated to have received service 15 on the TSB Log, the Amplification Heater would have already been replaced under routine maintenance with an in specification Amplification Heater and no further action is required for that instrument.
Recall details
- Recall number
- Z-1488-2021
- Status
- Open, Classified
- Initiated by the company
- Mar 18, 2021
- Recalling firm
- Luminex Corporation
- Quantity
- 1210 Units
- Product code
- PCH
- FDA submissions
- K140083
- Distribution
- Worldwide distribution - United States (nationwide) and the countries of Austria, France, Greece, Japan, Netherlands, Turkey.
Product description
VERIGENE Processor SP - Product Usage: intended for processing and identifying nucleic acids utilizing gold nanoparticle probe technology.
Code information (models, lots, serial numbers)
Show the full code information (12k characters)
VERIGENE Processor SP: Model Number 10-0000-07, UDI 00840487101483; Serial Numbers 19072007, 18255013, 18255002, 18159005, 18159001, 17215007, 21060901, 21055901, 20310901, 18023004, 18023003, 18023002, 18023001, 19072015, 19072014, 18318010, 20261007, 20261004, 20261002, 20167003, 20153009, 20133001, 18344012, 18299005, 18299001, 18250014, 13210273, 13105142, 18235013, 18235010, 18214004, 13186249, 18214011, 17215006, 17188008, 17188007, 17188006, 17188005, 17188004, 17188003, 17188002, 17188001, 17130057, 17130055, 17130054, 17117041, 14196165, 18177009, 18177007, 21039901, 19192005, 14052037, 14142120, 14142119, 19295006, 19044007, 17262004, 14322266, 20094003, 19023004, 19023003, 19023002, 19023001, 16334080, 14218188, 14218187, 20188009, 20188007, 13133183, 19325007, 19325003, 19317010, 19233009, 17096038, 17096036, 18120006, 15208083, 15131078, 15096068, 14062046, 12324410, 12324409, 19260010, 19260005, 13018026, 13018025, 12236277, 18079009, 15096067, 16355087, 16347081, 15131075, 15131073, 14069059, 14069058, 14237219, 14237217, 16077012, 16077011, 20167006, 20167001, 20121009, 19140004, 17234005, 17234004, 14191156, 12305380, 12305379, 12303378, 12303377, 12303376, 12303375, 11213075, 11213073, 11213072, 11213071, 12236271, 12236270, 18128008, 12170169, 12170168, 19057001, 12038011, 15075059, 15075058, 20212006, 18172014, 19260006, 19192007, 20007010, 18318001, 12285355, 12285354, 17278007, 14322268, 20086010, 20086006, 12278340, 17096034, 15040037, 15040036, 17307006, 17307004, 18255007, 18255001, 13238308, 12107091, 20094005, 20036009, 18060001, 13087118, 13087117, 18333002, 18318008, 18226015, 20086009, 20086002, 19095007, 18333012, 17048020, 17048019, 13263328, 13263327, 19095014, 19084008, 17048011, 17017010, 14265244, 14265243, 14265242, 18250001, 18172007, 18172003, 18159002, 20133006, 20133004, 19031006, 19031005, 18344009, 12205218, 12205217, 12205216, 19140010, 19140007, 18214013, 20027004, 20212002, 20212001, 16277051, 15015004, 14211179, 14211178, 14211174, 13178239, 13178238, 13178237, 13178236, 13154201, 13140200, 13140199, 18079001, 18022005, 17157061, 18054003, 18054002, 12261312, 12261311, 12261310, 12261309, 12254304, 18128002, 15327108, 15327107, 18128006, 18120004, 18120001, 18099009, 18099007, 11276125, 11276124, 11276123, 11276122, 11186063, 11186062, 11152041, 14251232, 20111002, 20086008, 18177014, 18177006, 9202053, 18288014, 18288011, 18255004, 18250010, 18172008, 17017003, 13161218, 12219236, 12079064, 12079061, 11276121, 11276111, 11152052, 11152051, 10250123, 10250120, 10139083, 10139081, 10139080, 10099037, D908, D09336158, D09336157, D09325136, D09325135, D09320135, D09320134, D09320133, D09320132, D09320131, D09320130, D09320129, D09320128, D09313126, D09255073, D09135024, D0912, D0909, D0908, D0903, 21025901, 20048006, 19233003, 19095013, 19095001, 19072011, 19072009, 19044003, 18344004, 18318009, 18299012, 18299008, 18288008, 18235007, 18235005, 18150009, 18088005, 18088003, 18072005, 18060009, 18046010, 18046005, 18046004, 18046003, 18046002, 18046001, 18039004, 18037005, 18037004, 18037003, 18037002, 18026002, 17356002, 17347002, 17340004, 17278001, 17242007, 17215001, 17117044, 17096037, 17096035, 17096033, 17096032, 17062026, 16347083, 16299066, 15327109, 15208088, 15131080, 15131079, 15075056, 15033030, 15033029, 15033026, 15020016, 15020015, 14307259, 14265248, 14265247, 14230210, 14230209, 14230201, 14211173, 14211172, 14142111, 14139104, 14043026, 14006008, 13315360, 13315352, 13301349, 13301348, 13301343, 13301342, 13238309, 13227296, 13224285, 13193257, 13186241, 13170229, 13140195, 13140194, 13133184, 13133181, 13127179, 13066087, 13060074, 13051065, 13051062, 13037048, 13008016, 13008009, 13008008, 13003013, 13003012, 13003011, 13003009, 12324413, 12317405, 12310394, 12310392, 12310391, 12303373, 12303372, 12303371, 12303370, 12285357, 12248292, 12248291, 12233259, 12205214, 12205213, 12205211, 12175193, 12175192, 12156162, 12145150, 12145148, 12145147, 12145146, 12145145, 12142138, 12088816, 12086077, 12069058, 12038815, 12038016, 12038015, 12016002, 11339160, 11339159, 11339158, 11339157, 11306147, 11306146, 11306144, 11306142, 11306141, 11236093, 11236092, 11236091, 11236090, 11236087, 11213085, 11213076, 11213074, 11186066, 11186064, 11152047, 11152046, 11152044, 11101028, 11021003, 10295149, 10295148, 10295147, 10295146, 10295145, 10295144, 10295143, 10295142, 10295141, 10295140, 10295139, 10295136, 10295135, 10278163, 10278133, 10278132, 10278131, 10278128, 10276131, 10250119, 10250118, 10250117, 10210115, 10210111, 10210110, 10210109, 10210108, 10210107, 10210106, 10210105, 10182104, 10182103, 10182102, 10182101, 10162085, 10139084, 10110058, 10099033, 10099032, 10035106, 10035105, 10035017R, 10014012, 10014003, 10014001, 09350165, 09350164, 09350163, 09350161, 09350160, 09350159, 09336158, 09336157, 09336150, 09336149, 09336147, 09336146, 09336145, 09336144, 09325138, 09320127, 09313120, 09313119, 09298118, 09298117, 09298116, 09293111, 09293109, 09293106, 09293103, 09288102, 09288097, 09288096, 09288095, 09270087, 09270086, 09270085, 09270084, 09270082, 09270080, 09255080, 09255079, 09255077, 09255072, 09255071, 09230075, 09230074, 09230070, 09230066, 09230064, 09210062, 09210061, 09210059, 09210058, 09210056, 09202052, 09202047, 09163046, 09163045, 09163044, 09163043, 09163042, 09163040, 09163039, 09163038, 09163035, 09163034, 09135029, 09135028, 09135027, 09135026, 09135024, 09121022, 09121019, 09121018, 09121015, 0912, 0911, 09100014, 09100012, 09100011, 09098006, 09098005, 09098003, 09098002, 09098001, 0909, 0908, 0907, 0902, 20094009, 20094007, 20094006, 20094001, 18226006, 18226005, 13092121, 13087120, 13087119, 17017005, 17017004, 14265250, 14265249, 20058007, 19206006, 13087113, 13087112, 21006901, 21005901, 20353901, 20167009, 20027009, 20027003, 19095002, 13105146, 12285351, 12285350, 12219238, 12219233, 18046008, 18046007, 16049006, 16049005, 18177005, 18177003, 18177001, 18264001, 18172004, 16355090, 15096066, 15096064, 10110052, 10110051, 09230068, 09230067, 18023008, 18022004, 18022002, 18022001, 20133007, 20133002, 20121003, 20111005, 20111003, 19260004, 19233010, 19206007, 19084006, 19072012, 19072005, 19057004, 19057003, 18318011, 18299010, 18250015, 14118091, 14111090, 14224196, 14224195, 18039003, 18039001, 20071003, 20048010, 15040035, 15040034, 15040032, 15040031, 12088089, 12088088, 19343002, 19343001, 16319074, 16319073, 16319072, 16319071, 19282005, 19123006, 20086001, 19155009, 16214040, 16214039, 16214038, 16214037, 12268320, 12268319, 12261318, 12261317, 12261316, 10182096, 10182095, 10162094, 10110057, 19072008, 19057014, 15131077, 15131076, 13121170, 13121169, 14211171, 14196170, 14196169, 21014901, 18141003, 18088010, 18088009, 18088008, 18088007, 18088006, 18159003, 18156010, 15320092, 15320091, 19305001, 15033023, 14251236, 14251235, 19325005, 19317009, 19044008, 18318004, 18299014, 18299013, 19206008, 18079005, 19123008, 19095012, 19123010, 19123001, 10014012 Correct, 20315902, 19095009, 19095004, 19036006, 19036005, 19031002, 18333007, 18318012, 18299006, 18299004, 18264009, 18205007, 19031010, 19031008, 19031007, 14224194, 19317004, 19317001, 19305004, 14052032, 14052031, 12248290, 21062901, 19072013, 16077016, 19072010, 19036008, 17062030, 17062027, 17062024, 14218190, 14218189, 13140198, 18264004, 18264003, 12233264, 16125022, 16125021, 16077020, 15020020, 15020019, 12198195, 12198194, 18344002, 16049010, 16049009, 16049008, 19155010, 19155008, 19155003, 19155002, 19155001, 19123007, 17177009, 17177005, 18172011, 18159011, 19140005, 19123009, 19044013, 19044001, 17157064, 11250098, 11250096, 19095008, 19295008, 19295005, 12128118, 20346901, 21006902, 21021902, 21019901, 21015901, 21005902, 20363901, 20349902, 19295010, 19282004, 19282003, 18250004, 18214015, 18214012, 17296004, 20048007, 20036010, 20036005, 20036001, 19155005, 19044012, 19036004, 18205008, 18156012, 18039009, 18039008, 17117046, 16077018, 12212220, 12212219, 18205014, 18191014, 18191002, 14155125, 18191008, 14230207, 14230206, 14182148, 14182147, 14182146, 14182145, 14182144, 16334076, 15015008, 14322263, 13245317, 13245316, 19057006, 16077019, 14244227, 14244226, 14244225, 14244224, 14142118, 14142117, 13042056, 13042055, 13042053, 18046006, 13186244, 13186243, 13186242, 13003010, 18024010, 18024007, 18024006, 14307252, 14307251, 19044015, 19044014, 18333006, 14196167, 14196166, 19260007, 19260002, 19233007, 14118098, 14118096, 21011901, 13193253, 11339152, 11339151, 18255009, 18255005, 18226009, 18226008, 12061041, 12053039, 11250103, 11250101, 11250100, 11250099, 13060072, 13060071, 13051070, 13051069, 19072004, 18205011, 13224286, 13224284, 13224283, 19057011, 19057009, 19023005, 16299063, 16277060, 17353009, 17353003, 14196161, 14191160, 12069057, 12061048, 19095011, 19095003, 18172006, 18172005, 18172001, 18054005, 17062028, 17048016, 20086007, 13227299, 13227298, 12236276, 14230208, 12088082, 12088081, 12086079, 12086078, 15327110, 14090076, 13042052, 13024038, 11250095, 11250094, 10182100, 21013901, 20153003, 20153002, 18235003, 18214005, 18344015, 18344011, 18214003, 18177015, 17356003, 16355089, 16355088, 17188010, 17188009, 14196164, 14196163, 14196162, 21033901, 19044011, 19044006, 13133186, 13133185 VERIGENE Processor SP, Refurbished: Model Number 10-0000-07R, UDI 00840487101490, Serial Numbers 17307003, 14083065, D09288101, 12170178, 11276119, 11276118, 11276117, 13140192, 13127174, 13112151, 13105150, 13008013, 13238304, 13224288, 12016006, 13238306, 13238305, 13087111, 13210277, 14118093, 11040009, 12198200, 12175186, 12175180, 11101032, 11021001, 10035019, 09336153, 14069051, 13245318, 12254303, 14111087, 14111086, 11236088, 10035020, 09121016, 14251234, 13274336, 13193258, 13133188, 14083063, 13315359, 12142139, 11040012, 13178233, 12271338, 14142115, 14139103, 11213069, 13274340, 12324411, 12248294, 15131071, 15075053, 12340438, 12198198, 12175188, 09163032, 09135030, 13161214, 12317400, 12310398, 12233260, 11339155, 11276116, 13079110, 10278127, 14139110, 13203264, 12053033, 12053031, 11186065, 14069060, 13301350, 10295153, 17307001, 16242043, 13072091, 13066081, 10035014, 14177137, 12079067, 13161217, 13161215, 13245315, 12240287, 10278126, 10278125, 14244221, 11122036, 14052035, 13079102, 13066084, 10210114, 12198204, 12198203, 10043026, 13245313, 13203263, 12170167, 12128117, 12128116, 15208087, 15020013, 14251238, 14251237, 13274334, 13066090, 13051064, 12317408, 12317407, 12317406, 12248289, 12175187, 12175185, 12136133, 12128120, 12111106, 12111105, 12111104, 12111102, 12069053, 12069051, 12061049, 12038019, 12016010, 12016008, 12016007, 11339150, 11213086, 11213070, 11186059, 11186057, 11122039, 11122037, 11101029, 11068018, 11040015, 11040013, 11021004, 10350169, 10350168, 10350166, 10350165, 10336160, 10336159, 10336158, 10336157, 10336156, 10336155, 10295154, 10295152, 10295151, 10210112, 10182097, 10162086, 10119078, 10102045, 10099044, 10099043, 10099042, 10099041, 10099040, 10099039, 10099038, 10099035, 10043022, 10043021, 10035017, 10014010, 10014009, 10014008, 10014007, 09336154, 09336152, 09325142, 09325141, 09325140, 09325139, 09325137, 09293107, 09293105, 09293104, 09288100, 09255076, 09202049, 09163033, 09100008, 13072092, 11021008, 15020017, 14251233, 14083064, 11306138, 10295150, 10278134, 10162091, 10162090, 13008011, 12310393, 14307258, 10350171, 10350170, 14211180, 09293108, 09277087, 14322261, 14083069, 17296006, 11250093, 11250091, 12248297, 11306134, 10336161, 13087115, 09336143, 10119077, 10035018, 09288098, 12271336, 11021005, 12053032, 12219229, 12198197, 12198196, 12128121, 14191158, 14191157, 12198205, 10014011, 19036001, 09350166, 09270083, 14230204, 14083066, 13224290, 12261315, 12261313, 12136132, 12136131, 13003003, 12303369, 13170224, 12268327, 12229250, 12198201, 12045024, 11306139, 11250092, 15020012, 14251239, 12128114, 12128113, 12038020, 12305381, 12271329, 12292363, 12268326, 10035013, 13186248, 13186247, 13178240, 12111101, 10336164, 10336163, 13224287, 13018024, 12278344, 15055044, 15033025, 14083067, 14069053, 13072093, 12278343, 12271337, 14177134, 13301341, 10278129, 13127180, 11101030, 11021006, 10278130, 10099036, 10099030, 13161216, 12142140, 13315358, 12198202, 11101031, 12271330, 12229255
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Verigene® Processor SP 10-0000-07Luminex Corporation · Instrumentation for clinical multiplex test systems3 recalls15 reports
- Verigene® Processor SP (Refurbished) 10-0000-07RLuminex Corporation · Instrumentation for clinical multiplex test systems3 recalls6 reports
Other devices under the same FDA submission
2 devices cleared under K140083 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Enteric Pathogens (EP) Stool Sample Preparation Kit 30-002-023Luminex Corporation · Gastrointestinal pathogen panel multiplex nucleic acid-based assay system2 recalls1 report
- Enteric Pathogens (EP) Nucleic Acid Test Cartridge 20-006-023Nanosphere, Inc. · Gastrointestinal Bacterial Panel Multiplex Nucleic Acid-Based Assay System1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗