Enteric Pathogens (EP) Stool Sample Preparation Kit 30-002-023Recalls & reported problems
Luminex Corporation · The Verigene® Enteric Pathogens Nucleic Acid Test (EP) is a multiplexed, qualitative test for simultaneous detection and identification of common pathogenic enteric bacteria, viruses, and genetic virulence markers from liquid or soft stool preserved in Cary-Blair medium, collected from individuals with signs and symptoms of gastrointestinal infection. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840487101568), received between May 2, 2023 and May 2, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Chemical Problem” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K140083); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jun 5, 2023Class IIOpen, Classified
It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.
Z-2023-2023 · Root cause: Other · Recall details · FDA record ↗
- Jun 17, 2022Class IIOpen, Classified
There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.
Z-1335-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Chemical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2023Chemical Problem; Chemical Problem
ON (B)(6) 2023, (B)(6) HOSPITAL NOTIFIED LUMINEX THAT THEY OBSERVED THEIR EP STOOL PREP KIT NOT ABSORBING THE CARY BLAIR BUFFER PROPERLY. THE CUSTOMER WAS UNAWARE OF WHICH SPECIFIC KIT THE SWABS ORIGINATED FROM. LOT OF SWAB PROVIDED: 247P17 - LN:202699 EXP 4/30/2025; EP KIT LOT NOT KNOWN AS STOOL PREP KITS WERE SEPARATED FROM SHIPMENT. NO PATIENT SAMPLES WERE AFFECTED BY THE SWAB CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16854218 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140083. One submission often covers many models, so check the product and lot details in each record.
- Z-2451-2021 · Aug 3, 2021 · Open, ClassifiedVerigene EP Amplification Reagent Kit Test
- Z-1488-2021 · Mar 18, 2021 · Open, ClassifiedVERIGENE Processor SP - Product Usage: intended for processing and identifying nucleic acids utilizing gold nanoparticle probe technology.
- Z-2009-2021 · Oct 30, 2020 · Open, ClassifiedVerigene Enteric Pathogens Nucleic Acid Test (EP) Model Number: 20-006-023
- Z-2685-2014 · Oct 10, 2013 · TerminatedThe Verigene Enteric Pathogens Nucleic Acid Test (EP) is a multiplexed, qualitative test for simultaneous detection and identification of common pathogenic enteric bacteria and gen
Other Gastrointestinal pathogen panel multiplex nucleic acid-based assay system devices
- FilmArray® Gastrointestinal (GI) Panel RFIT-ASY-0116Biofire Diagnostics, LLC · Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System6 recalls2 deaths24 reports
- BD MAX™ Enteric Bacterial Panel 442963Becton, Dickinson And Company · Gastrointestinal pathogen panel multiplex nucleic acid-based assay system2 recalls102 reports
- BD MAX™ Extended Enteric Bacterial Panel 443812Becton, Dickinson And Company · Gastrointestinal pathogen panel multiplex nucleic acid-based assay system1 recall114 reports
- BD MAX™ Enteric Parasite Panel 442960Becton, Dickinson And Company · Gastrointestinal pathogen panel multiplex nucleic acid-based assay system1 recall70 reports
- xTAG® Gastrointestinal Pathogen Panel (GPP) I032C0316Luminex Molecular Diagnostics, Inc · Gastrointestinal pathogen panel multiplex nucleic acid-based assay system1 recall0 reports
- FilmArray® Gastrointestinal (GI) Panel RFIT-ASY-0104Biofire Diagnostics, LLC · Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page