VenaSeal Closure System recallZ-1564-2021 · Medtronic Inc. · initiated Mar 23, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm received increased number of reports that VenaSeal dispensing systems are not effectively advancing the adhesive forward through the delivery catheter.
FDA root cause category: Device Design
What the company is doing
A Recall notification letter titled, "URGENT: MEDICAL DEVICE CORRECTION" was sent to consignees on 03/23/2021 via UPS 2-day delivery. The consignee is requested to: -Review the VenaSeal Gun Assist Process as provided in Appendix A. -Please share this information with any other healthcare professionals in their facility who use the VenaSeal Closure System. Also share this information with any other organization where these devices may have been transferred. -Maintain a copy of this letter for the consignee's reference. -Please complete the enclosed Customer Confirmation Form and return via email [email protected] -Please notify Medtronic of any adverse events or quality problems associated with use of this product. -Complete and submit the report Online: www.fda.gov/medwatch/report.htm -Regular Mail or Fax: Download form www.fda.gov/MedWatch/getformstm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA- 0178 In other geographies, communication of the Medical Device Correction letter began on 23-Mar-2021, via regionally approved methods. Consignees will be asked to complete and return a confirmation certificate, confirming they have received the notification.
Recall details
- Recall number
- Z-1564-2021
- Status
- Terminated
- Initiated by the company
- Mar 23, 2021
- Recalling firm
- Medtronic Inc.
- Quantity
- 154,152 units
- Product code
- PJQ
- FDA submissions
- P140018
- Distribution
- Distributed nationwide to MA, NY, VT, NH, ME
Product description
VenaSeal Closure System
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
Model Number SP-101: GTIN 20643169986279, Serial Numbers: 54143, 54145, 54146, 54147, 54148, 54149, 54150, 54151, 54152, 54396, 54397, 54398, 54399, 54417, 54418, 54419, 54420, 54618, 54619, 54620, 54621, 54622, 54771, 54772, 54773, 54774, 54775, 54776, 54794, 54795, 54796, 54924, 54925, 54926, 54927, 54928, 54929, 55027, 55028, 55029, 55030, 55031, 55032, 55068, 55069, 55070, 55071, 55072, 55073, 55113, 55114, 55115, 55116, 55437, 55438, 55439, 55440, 55441, 55442, 55483, 55484, 55485, 55486, 55487, 55488, 55594, 55595, 55596, 55597, 55598, 55599, 55675, 55676, 55677, 55678, 55683, 55684, 55775, 55776, 55777, 55778, 55790, 55791, 55792, 55793, 55794, 55852, 55853, 55854, 55855, 55856, 55948, 55949, 55951, 55952, 55953, 55991, 55992, 55993, 55994, 55995, 56109, 56110, 56111, 56123, 56159, 56160, 56161, 56162, 56163, 56189, 56193, 56195, 56196, 56216, 56218, 56219, 56387, 56388, 56389, 56391, 56464, 56465, 56466, 56467, 56468, 56528, 56529, 56530, 56531, 56532, 56533, 56534, 56661, 56662, 56663, 56664, 56741, 56742, 56829, 56830, 56831, 56832, 56833, 56846, 56847, 56848, 56928, 56929, 56930, 56931, 56932, 56933, 56935, 57063, 57177, 57178, 57179, 57226, 57227, 57268, 57269, 57270, 57271, 57272, 57273, 57439, 57440, 57560, 57561, 57562, 57563, 57564, 57565, 57567, 57884, 57885, 57947, 57948, 58257, 58258, 58363, 58364, 58564, 58566, 58567, 58568, 58897, 58898, 58899, 59249, 59337, 59338, 59339, 59909, 59910, 59911, 60264, 60265, 60682, 60683, 60684, 60685, 60763, 60764, 60765, 60774, 60775, 60776, 60876, 60877, 60880, 60881, 60882, 60883, 60884, 60886, 60887, 60888, 60889, 60890, 60891, 60892, 60893, 60894, 60895, 60896, 60897, 60898, 60899, 60900, 61189, 61190, 61191, 61192, 61193, 61254, 61257, 61262, 61264, 61312, 61313, 61314, 61315, 61598, 61599, 61600, 61605, 61606, 61607, 61612, 61613, 61617, 61925, 61928, 61935, 61936, 61937, 61944, 61946, 61953, 61960, 53106, 53107, 53388, 53449, 54420, 54618, 54929, 55028 Model Number VS-402: GTIN 00613994384089, Serial Numbers: 53830, 54210, 54211, 54421, 55491 GTIN 00643169437937, Serial Numbers: 53841 GTIN 00643169931459, Serial Numbers: 53928 GTIN 20643169986262, Serial Numbers: 53821, 53834, 53835, 53842, 53843, 53845, 53894, 53896, 53900, 53926, 53927, 53929, 53938, 54141, 54212, 54214, 54215, 54216, 54217, 54218, 54220, 54221, 54222, 54287, 54288, 54289, 54291, 54292, 54318, 54320, 54321, 54323, 54325, 54360, 54362, 54363, 54367, 54371, 54372, 54393, 54395, 54422, 54549, 54550, 54552, 54555, 54558, 54559, 54567, 54568, 54695, 54696, 54697, 54698, 54777, 54779, 54780, 54791, 54792, 54793, 54830, 54831, 54834, 55076, 55077, 55110, 55111, 55112, 55227, 55357, 55443, 55444, 55489, 55601, 55602, 55679, 55680, 55686, 55687, 55779, 55782, 55796, 55797, 55860, 55861, 55955, 55959, 55962, 55963, 55971, 56113, 56115, 56116, 56164, 56165, 56166, 56167, 56197, 56198, 56199, 56200, 56201, 56220, 56221, 56222, 56384, 56385, 56386, 56469, 56470, 56471, 56472, 56473, 56535, 56536, 56537, 56641, 56736, 56737, 56739, 56740, 56818, 56819, 56820, 56821, 56822, 56823, 56824, 56825, 56826, 56827, 56840, 56841, 56842, 56843, 56844, 56845, 56937, 56938, 56939, 56940, 56941, 56942, 56943, 56944, 56945, 56946, 56947, 56948, 56949, 56950, 56951, 56952, 56953, 56954, 56955, 57077, 57082, 57180, 57181, 57182, 57183, 57184, 57185, 57186, 57228, 57229, 57230, 57231, 57232, 57233, 57234, 57235, 57236, 57237, 57274, 57275, 57276, 57277, 57278, 57279, 57280, 57281, 57282, 57283, 57284, 57285, 57286, 57287, 57288, 57289, 57290, 57291, 57296, 57297, 57298, 57299, 57568, 57569, 57570, 57571, 57572, 57573, 57574, 57575, 57576, 57577, 57578, 57579, 57580, 57581, 57582, 57583, 57584, 57586, 57587, 57588, 57589, 57590, 57591, 57592, 57593, 57594, 57595, 57596, 57597, 57598, 57599, 57600, 57601, 57602, 57603, 57604, 57605, 57606, 57607, 57608, 57886, 57887, 57888, 57889, 57949, 57950, 57951, 57952, 57953, 57954, 57955, 57962, 57963, 57964, 57965, 57966, 57967, 57968, 57969, 57970, 57971, 57972, 58057, 58058, 58059, 58060, 58061, 58062, 58081, 58082, 58083, 58084, 58085, 58101, 58102, 58103, 58104, 58105, 58107, 58108, 58109, 58110, 58111, 58112, 58113, 58193, 58194, 58195, 58259, 58260, 58261, 58262, 58356, 58357, 58358, 58359, 58360, 58378, 58379, 58380, 58381, 58382, 58383, 58384, 58385, 58386, 58387, 58388, 58389, 58415, 58416, 58417, 58418, 58419, 58420, 58421, 58422, 58423, 58424, 58425, 58426, 58427, 58428, 58429, 58430, 58431, 58432, 58433, 58434, 58435, 58436, 58437, 58438, 58439, 58440, 58441, 58442, 58569, 58570, 58571, 58572, 58573, 58602, 58603, 58604, 58605, 58609, 58610, 58611, 58612, 58613, 58614, 58615, 58616, 58617, 58618, 58619, 58620, 58621, 58622, 58623, 58624, 58625, 58626, 58627, 58628, 58629, 58630, 58631, 58632, 58633, 58634, 58635, 58636, 58637, 58638, 58639, 58640, 58641, 58642, 58643, 58644, 58645, 58646, 58647, 58839, 58901, 58902, 58903, 58904, 58905, 58906, 58907, 58908, 58988, 58989, 58990, 58991, 58992, 58993, 58994, 58995, 58996, 58997, 58998, 58999, 59000, 59001, 59002, 59003, 59004, 59005, 59006, 59007, 59008, 59009, 59010, 59011, 59012, 59013, 59014, 59015, 59016, 59017, 59018, 59019, 59020, 59021, 59022, 59023, 59151, 59234, 59251, 59252, 59253, 59254, 59255, 59256, 59257, 59258, 59259, 59260, 59261, 59262, 59263, 59264, 59265, 59266, 59267, 59268, 59269, 59270, 59271, 59272, 59273, 59274, 59275, 59276, 59277, 59278, 59279, 59280, 59281, 59282, 59283, 59284, 59285, 59286, 59343, 59344, 59345, 59349, 59356, 59357, 59358, 59359, 59360, 59361, 59362, 59363, 59364, 59365, 59366, 59367, 59368, 59584, 59585, 59586, 59690, 59912, 59913, 59914, 59915, 59916, 59954, 59955, 59956, 59957, 59958, 59959, 59960, 59961, 59962, 59963, 59964, 59965, 59966, 59967, 59968, 59969, 59970, 59971, 59972, 59973, 59974, 60124, 60266, 60267, 60268, 60269, 60270, 60271, 60272, 60273, 60274, 60275, 60276, 60277, 60278, 60279, 60280, 60281, 60282, 60283, 60687, 60688, 60689, 60690, 60691, 60692, 60693, 60694, 60695, 60766, 60767, 60768, 60777, 60778, 60779, 60780, 60781, 60782, 60783, 60784, 60785, 60786, 60787, 60788, 60793, 60794, 60795, 60796, 60797, 60798, 60799, 60800, 60801, 60812, 60813, 60814, 60815, 60816, 60818, 60819, 60820, 60821, 60822, 60823, 60824, 60825, 60826, 60828, 60829, 60830, 60831, 60832, 60833, 60834, 60878, 60879, 60901, 60902, 60903, 60904, 60905, 60906, 60907, 60908, 60909, 61187, 61188, 61246, 61247, 61248, 61249, 61250, 61251, 61252, 61253, 61258, 61259, 61260, 61261, 61265, 61266, 61267, 61268, 61269, 61270, 61271, 61272, 61273, 61274, 61275, 49967, 50226, 50229, 50431, 50547, 50651, 50816, 50832, 50833, 50972, 51017, 51076, 51077, 51178, 51195, 51307, 51310, 51461, 51509, 51590, 51657, 51659, 51660, 51752, 51755, 51757, 51768, 51794, 51795, 51856, 51859, 51872, 51908, 51909, 51910, 51958, 51961, 51982, 52028, 52029, 52128, 52129, 52130, 52138, 52139, 52217, 52222, 52224, 52225, 52226, 52274, 52275, 52293, 52294, 52297, 52372, 52373, 52412, 52413, 52414, 52415, 52416, 52417, 52420, 52426, 52427, 52429, 52438, 52441, 52445, 52447, 52448, 52449, 52454, 52483, 52485, 52487, 52488, 52489, 52490, 52491, 52504, 52505, 52506, 52599, 52600, 52601, 52602, 52639, 52640, 52641, 52642, 52646, 52647, 52648, 52649, 52706, 52707, 52708, 52718, 52719, 52720, 52721, 52722, 52723, 52728, 52729, 52737, 52738, 52739, 52740, 52741, 52790, 52791, 52792, 52793, 52794, 52795, 52796, 52797, 52801, 52802, 52803, 52804, 52805, 52806, 52815, 52816, 52817, 52818, 52819, 52820, 52821, 52823, 52829, 52832, 52833, 52834, 52835, 52836, 52837, 52838, 52839, 52895, 52896, 52897, 52898, 52899, 52900, 52901, 52902, 52903, 52904, 52905, 52910, 52911, 52912, 52913, 52914, 52915, 52916, 52917, 52918, 52938, 52939, 52940, 52941, 52942, 52943, 52944, 52945, 53025, 53026, 53042, 53043, 53044, 53051, 53052, 53053, 53054, 53055, 53056, 53057, 53058, 53060, 53061, 53062, 53063, 53064, 53098, 53099, 53100, 53101, 53170, 53171, 53172, 53173, 53174, 53175, 53176, 53186, 53187, 53188, 53189, 53190, 53197, 53198, 53199, 53200, 53201, 53202.53212, 53213, 53214, 53215, 53216, 53217, 53254, 53255, 53256, 53257, 53258, 53259, 53260, 53261, 53262, 53263, 53264, 53265, 53266, 53267, 53268, 53317, 53318, 53319, 53320, 53321, 53322, 53323, 53324, 53325, 53326, 53327, 53328, 53339, 53459, 53460, 53461, 53463, 53464, 53465, 53466, 53467, 53468, 53539, 53540, 53541, 53542, 53543, 53562, 53563, 53565, 53566, 53567, 53584, 53585, 53586, 53587, 53588, 53637, 53638, 53639, 53640, 53652, 53653, 53654, 53655, 53656, 53703, 53712, 53713, 53714, 53715, 53716, 53718, 53719, 53720, 53721, 53722, 53814, 53815, 53816, 53817, 53818, 53819, 53820, 53821, 53830, 53831, 53832, 53833, 53834, 53835, 53841, 53842, 53843, 53844, 53845, 53846, 53890, 53895, 53896, 53897, 53898, 53899, 53900, 53901, 53925, 53926, 53927, 53928, 53929, 53931, 53936, 53937, 53938, 53939, 53940, 53941, 54052, 54053, 54054, 54140, 54141, 54209, 54210, 54211, 54212, 54214, 54215, 54216, 54217, 54218, 54219, 54220, 54221, 54287, 54288, 54289, 54290, 54291, 54318, 54320, 54321, 54322, 54323, 54324, 54325, 54326, 54327, 54359, 54360, 54361, 54362, 54363, 54364, 54365, 54366, 54367, 54368, 54369, 54370, 54371, 54372, 54373, 54392, 54393, 54394, 54395, 54421, 54422, 54423, 54424, 54451, 54452, 54453, 54454, 54455, 54456, 54457, 54549, 54550, 54551, 54552, 54553, 54554, 54555, 54556, 54557, 54558, 54559, 54567, 54568, 54695, 54696, 54697, 54698, 54777, 54778, 54779, 54780, 54781, 54782, 54783, 54790, 54791, 54792, 54793, 54830, 54831, 54832, 54833, 54834, 55074, 55075, 55076, 55077, 55107, 55108, 55109, 55110, 55111, 55112, 55224, 55225, 55226, 55227, 55228, 55229, 55230, 55231, 55353, 55354, 55355, 55356, 55357, 55358, 55359, 55360, 55361, 55362, 55364, 55365, 55443, 55444, 55445, 55489, 55490, 55491, 55600, 55601, 55602, 55679, 55680, 55681, 55682, 55685, 55686, 55687, 55779, 55780, 55781, 55782, 55795, 55796, 55797, 55858, 55859, 55860, 55861, 55862, 55955, 55959, 55962, 55963, 55970, 56114, 56115, 56164, 56165, 56166, 56167, 56197, 56198, 56199, 56200, 56202, 56384, 56469, 56470, 56473 Model Number VS-404: GTIN 00643169437975, Serial Numbers: 61677 GTIN 00673978070469, Serial Numbers: 62015 GTIN 00673978070872, Serial Numbers: 62006 GTIN 00763000240776, Serial Numbers: 61475, 61673, 61674, 61678, 61696, 61933, 61934, 61938, 61939, 61940, 61941, 61943, 61947, 62013, 62022 GTIN 00763000279769, Serial Numbers: 61675 Please note that VS-403 is not approved in the U.S. and has not been distributed outside of the United States; the scope of this corrective action will include this product as it is launched in OUS regions beginning Summer 2021.
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- VenaSeal™ VS-404Medtronic, Inc. · Agent, occluding, vascular, permanent1 recall2 deaths135 reports
- VenaSeal™ VS-402Medtronic, Inc. · Agent, occluding, vascular, permanent1 recall2 deaths65 reports
- NA EC300Medtronic Xomed, Inc. · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall1 report
- Midas Rex® EC300RFMedtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall1 report
- N/A 2035UMedtronic, Inc. · Cardiac ablation percutaneous catheter1 recall1 report
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- N/A 876-474Medtronic Sofamor Danek, Inc. · TAP, BONE1 recall0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under P140018 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- VenaSeal VS-402Covidien LP · Agent, Occluding, Vascular, Permanent7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗