Centricity recallZ-1533-2021 · GE Healthcare, LLC · initiated Mar 19, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When the Cross Enterprise Display option is enabled on the Centricity Universal Viewer and Workflow Manager option is used as the PACS worklist launching the viewer, the patient jacket may potentially include a different patient s comparison exam(s).
FDA root cause category: Under Investigation by firm
What the company is doing
A Recall notification letter titled, "URGENT MEDICAL DEVICE CORRECTION" was sent to the consignees on 03/19/2021 via letter. The letter instructs the consignee that they can continue to use the device by following the below safety instructions: "1. Visually verify that the patient demographic information for all comparison exams when hung in the Image Viewer match that of the primary patient under review. 2. If you encounter comparison exams that do not belong to the primary patient under review, close the primary exam and reopen the exam from Workflow Manager." The consignee is also told to destroy the installation media for the affected software once a GE Healthcare representative has updated the system.
Recall details
- Recall number
- Z-1533-2021
- Status
- Terminated
- Initiated by the company
- Mar 19, 2021
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 112 units
- Product code
- LLZ
- FDA submissions
- K182419
- Distribution
- Distributed nationwide to AR, CA, CO, DE, FL, IL, IN, LA, ME, NE, NJ, NV, NY, PA, SC, TX and internationally to Canada.
Product description
Centricity Universal Viewer - Product Usage: intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
UDI: 00840682145794; Serial Numbers: 105906-2-Centricity Universal Viewer-02312613; 6100002127; 4049-1-Centricity Universal Viewer-00104520; 4049-1-Centricity Universal Viewer-00104521; 4049-1-Centricity Universal Viewer-00104522; 6100003887; 9619-c23f-0fae-4eb2-95f8-1529-7e0f-cb89; 103835-2-Centricity Universal Viewer-01370706; 103835-2-Centricity Universal Viewer-01370707; 103835-2-Centricity Universal Viewer-14549671; 103835-2-Centricity Universal Viewer-19676013; 103835-2-Centricity Universal Viewer-19676014; 103835-2-Centricity Universal Viewer-19676015; 103835-2-Centricity Universal Viewer-19676016 103835-2-Centricity Universal Viewer-19735426; 1ec2-1d6c-8e6b-4261-a436-2190-812d-b5df; 10388-1-Centricity Universal Viewer-14762267; 10388-1-Centricity Universal Viewer-17744828; 10388-1-Centricity Universal Viewer-17744829; 10388-1-Centricity Universal Viewer-17744830; 10388-1-Centricity Universal Viewer-17744831; 10388-1-Centricity Universal Viewer-19661410; 6100003912; 4642-6-Centricity Universal Viewer-02255109; 4642-6-Centricity Universal Viewer-02255132; 4642-6-Centricity Universal Viewer-02255133; 4642-6-Centricity Universal Viewer-14580266; 4642-6-Centricity Universal Viewer-14580267; 2457-1-Centricity Universal Viewer-00108168; 3791-1-Centricity Universal Viewer-18836815; 3791-9-Centricity Universal Viewer-15291373; 3791-9-Centricity Universal Viewer-19172098; 6100003882; b1aa-a5a8-dacb-4a74-90ca-41af-c9ed-83dd; 6100003897; 3545-3-Centricity Universal Viewer-00096349; 3545-3-Centricity Universal Viewer-00096350; 3545-3-Centricity Universal Viewer-00158590; 3545-3-Centricity Universal Viewer-00158591; 3545-3-Centricity Universal Viewer-00158592; 3545-3-Centricity Universal Viewer-00158593; 3545-3-Centricity Universal Viewer-00158594; 3545-3-Centricity Universal Viewer-01546163; 3545-3-Centricity Universal Viewer-01764841; 3545-3-Centricity Universal Viewer-01800418; 1198-18-Centricity Universal Viewer-00116885; 1198-18-Centricity Universal Viewer-00116886; 1198-18-Centricity Universal Viewer-00116887; 1198-18-Centricity Universal Viewer-00116888; 1198-18-Centricity Universal Viewer-00116889; 1198-18-Centricity Universal Viewer-00116890; 1198-18-Centricity Universal Viewer-00116891; 1198-18-Centricity Universal Viewer-00116892; 1198-18-Centricity Universal Viewer-00116894; 1198-18-Centricity Universal Viewer-00116895; 1198-18-Centricity Universal Viewer-00116896; 1198-18-Centricity Universal Viewer-00116897; 1198-18-Centricity Universal Viewer-00116898; 1198-18-Centricity Universal Viewer-00116908; 1198-18-Centricity Universal Viewer-01370310; 41a1-fc8a-46c3-48c0-a90d-86fc-4d2e-5676; 100866-1-Centricity Universal Viewer-01287648; 100866-1-Centricity Universal Viewer-01287649; 100866-1-Centricity Universal Viewer-19611328; 100866-1-Centricity Universal Viewer-19611337; 100866-1-Centricity Universal Viewer-19611338; 100866-1-Centricity Universal Viewer-20021255; 100866-1-Centricity Universal Viewer-20021256; 100866-10-Centricity Universal Viewer-19611339; 100866-10-Centricity Universal Viewer-19611340; 100866-10-Centricity Universal Viewer-19611341; 100866-4-Centricity Universal Viewer-19660158; 4314-1-Centricity Universal Viewer-14640637; 4314-1-Centricity Universal Viewer-14776705; 4314-1-Centricity Universal Viewer-14776706; 4314-1-Centricity Universal Viewer-14776707; 4314-1-Centricity Universal Viewer-14776708; 4314-1-Centricity Universal Viewer-14776724; 6bc8-ba1a-6f59-4e1c-a6b8-98b2-7262-75dc; 103915-1-Centricity Universal Viewer-19669927; 6901-11-Centricity Universal Viewer-01974089; 6901-11-Centricity Universal Viewer-02354909; 6901-11-Centricity Universal Viewer-02354910; 6901-11-Centricity Universal Viewer-02354911; 6901-11-Centricity Universal Viewer-02354912; 6901-11-Centricity Universal Viewer-02354913; 7275-45-Centricity Universal Viewer-01891908; 7275-45-Centricity Universal Viewer-19665236; 7275-45-Centricity Universal Viewer-19665240; 7275-45-Centricity Universal Viewer-19665241; 7275-45-Centricity Universal Viewer-19665242; 7275-45-Centricity Universal Viewer-19665243; 7275-45-Centricity Universal Viewer-19665244; 6901-12-Centricity Universal Viewer-02354849; 6901-12-Centricity Universal Viewer-02354850; 6901-12-Centricity Universal Viewer-02354851; 6901-12-Centricity Universal Viewer-02354852; 6901-12-Centricity Universal Viewer-02354853; 6901-12-Centricity Universal Viewer-02354872; 6901-12-Centricity Universal Viewer-02354874; 6901-12-Centricity Universal Viewer-02354875; 6901-12-Centricity Universal Viewer-02354876; 6901-12-Centricity Universal Viewer-02354877; 3760-1-Centricity Universal Viewer-19605860; 3760-1-Centricity Universal Viewer-19727283; 3760-1-Centricity Universal Viewer-19727284; 283043-Centricity Universal Viewer-02285955; 283043-Centricity Universal Viewer-02339571; 283043-Centricity Universal Viewer-14766564
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page