DAR recallZ-1866-2021 · Covidien, LP · initiated Apr 29, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The company was notified by their supplier of sterilization services for airway products in Italy of potential deviations from validated parameters for ethylene oxide sterilization.

FDA root cause category: Material/Component Contamination

What the company is doing

An Urgent Medical Device Field Safety Notice dated 4/29/21 was sent to customers. Actions you should take: 1. Immediately quarantine and discontinue use of item codes with associated lot numbers listed in Attachment A. 2. If you have distributed the devices listed in Attachment A, please promptly forward the information from this letter to those recipients. Actions being taken by Medtronic: 1. Medtronic is investigating the sterilization records for affected production lots to determine whether sterilization has been achieved despite the deviations. 2. Medtronic will communicate with affected customers regarding the finding of our investigation once complete and will provide direction on what customers should do with the devices quarantined because of this Field Safety Notice. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2. An URGENT MEDICAL DEVICE WITHDRAWAL notification letter dated 7/12/21 was sent to customers. Medtronic has concluded its investigation of potential deviations in the ethylene oxide sterilization processes performed by Steril Milano, the former supplier of our sterilization services for the DAR" airway products. Medtronic analyzed the available sterilization data and conducted validation tests on production lots where data was available. For products where sterilization process data was available and testing could be conducted, we determined that sterilization of those devices was achieved. However, because there was not a consistent controlled process at the sterilization facility, we are issuing a withdrawal of the production lots of the DAR" airway products which are currently under quarantine at your facility. Required Actions: 1. Please return affected product listed on Attachment A as indicated below. All unu

Recall details

Recall number
Z-1866-2021
Status
Open, Classified
Initiated by the company
Apr 29, 2021
Posted by the FDA
Jun 16, 2021
Recalling firm
Covidien, LP
Quantity
32625
Product code
CAH
FDA submissions
K941676
Distribution
Worldwide distribution - US Nationwide distribution in the states of DE, KY, NY, TX, VA, VT, and WA. Global Distribution.

Product description

Electrostatic Filter Angled, Item Code 350S19006 - Product Usage: indicated for single use on anesthetized patients and respiratory care patients who require a breathing circuit with ISO standard 15mm or 22 mm connectors.

Code information (models, lots, serial numbers)

GTIN Primary Package: 40884522012981 GTIN Secondary Package: 30884522012984  Lot Numbers: 20G1304FAX 20G1305FAX 20L0026FAX 20L0027FAX 20L0410FAX 20L0411FAX 20L0412FAX

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 87925).

Sources and updates