Venovo recallZ-1898-2021 · Bard Peripheral Vascular Inc · initiated Dec 14, 2020

Class II Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.

FDA root cause category: Under Investigation by firm

What the company is doing

On 01/13/2021, the firm sent an "URGENT MEDICAL DEVICE SAFETY NOTICE" via FedEx to customers informing them that the Recalling Firm has received reports regarding the venous stent system 9F where the proximal end of the stent does not immediately expand upon deployment, and remains connected to the delivery system, and that over-manipulation or forcing the catheter delivery system in attempts to assist the stent's expansion, could potentially have a varying degree of harm associated with it. potential harm ranges from prolonging the procedure, damaging or deformity of the stent, potential vascular injury and / or hemodynamic disruption affecting the blood flow and / or a thrombotic event. The Recalling Firm is informing customers that the affected product can continue to be used, but are instructing them to: 1. Share this communication with all physicians using the Venovo" Venous Stent Systems product within their organization to ensure that they are aware of this Safety Notice. 2. Continue the use of these Venovo" Venous Stent Systems 9F, 14mm products following the Safety Notice guidance. 3. Report any adverse health consequences experienced with the use of this product to the Recalling Firm. A list of the affected Product Codes and Lot/Serial Numbers are included as Appendix 1 attached to the Recall Notification Letter. For questions/inquiries/assistance contact the Recalling Firm's Customer/Technical Support at 1-888-237-2762 (Opt 4, Opt 2) Monday thru Friday 8:00 a.m. - 5:00 p.m. CST Updated Information for Recall Expansion (dated 05/11/2021): On 05/07/2021, the firm sent additional " "URGENT MEDICAL DEVICE SAFETY NOTICE" via FedEx to customers informing them that the scope of the recall (initiated on 01/13/2021) was being expanded to include all sizes and lots within expiry date. Customer are instructed to: 1. Check all inventory locations within their institution for the affected Venous Stent System listed in Attachment A and immediately

Recall details

Recall number
Z-1898-2021
Recall class
Class II
Status
Open, Classified
Initiated by the company
Dec 14, 2020
Recalling firm
Bard Peripheral Vascular Inc
Quantity
Total Devices=12,258 (U.S.= 7,234; O.U.S.=5,024)
Product code
QAN
FDA submissions
P180037
Distribution
Worldwide distribution. US Nationwide, ALGERIA, ANDORRA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BRUNEI, BULGARIA, CHILE, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, DEUTSCHLAND, EGYPT, FINLAND, FRANCE, GEORGIA, GREECE, HONG KONG, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, MALAYSIA, MEXICO, NETHERLANDS, NORWAY, OMAN, PANAMA, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDIA ARABIA, SINGAPORE, SLOVAKIA (SLOVAK REPUBLIC), SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TURKEY, UNITED ARAB EMIRATES (UAE), UNITED KINGDOM, VIETNAM

Product description

VENOVO Venous Stent System 8F Product Description(REF)/dimension: VENEL10040 /Venovo 8F 10/40/1200mm OUS; VENEL10060 /Venovo 8F 10/60/1200mm OUS; VENEL10080 /Venovo 8F 10/80/1200mm OUS; VENEL10100 /Venovo 8F 10/100/1200mm OUS; VENEL10120 /Venovo 8F 10/120/1200mm OUS; VENEL10140 /Venovo 8F 10/140/1200mm OUS; VENEL10160 /Venovo 8F10/160/1200mm OUS; VENEL12040 /Venovo 8F 12/40/1200mm OUS; VENEL12060 /Venovo 8F 12/60/1200mm OUS; VENEL12080 /Venovo 8F 12/80/1200mm OUS; VENEL12100 /Venovo 8F 12/100/1200mm OUS; VENEL12120 /Venovo 8F 12/120/1200mm OUS; VENEL12140 /Venovo 8F 12/140/1200mm OUS; VENEL12160 /Venovo 8F 12/160/1200mm OUS; VENEM10040 /Venovo 8F 10/40/800mm OUS; VENEM10060 /Venovo 8F 10/60/800mm OUS; VENEM10080 /Venovo 8F 10/80/800mm OUS; VENEM10100 /Venovo 8F 10/100/800mm OUS; VENEM10120 /Venovo 8F 10/120/800mm OUS; VENEM10140 /Venovo 8F 10/140/800mm OUS; VENEM10160 /Venovo 8F 10/160/800mm OUS; VENEM12040 /Venovo 8F 12/40/800mm OUS; VENEM12060 /Venovo 8F 12/60/800mm OUS; VENEM12080 /Venovo 8F 12/80/800mm OUS; VENEM12100 /Venovo 8F 12/100/800mm OUS; VENEM12120 /Venovo 8F 12/120/800mm OUS; VENEM12140 /Venovo 8F 12/140/800mm OUS; VENEM12160 /Venovo 8F 12/160/800mm OUS; VENUL10040 /Venovo 8F 10/40/1200mm US; VENUL10060 /Venovo 8F 10/60/1200mm US; VENUL10080 /Venovo 8F 10/80/1200mm US; VENUL10100 /Venovo 8F 10/100/1200mm US; VENUL10120 /Venovo 8F 10/120/1200mm US; VENUL10140 /Venovo 8F 10/140/1200mm US; VENUL10160 /Venovo 8F 10/160/1200mm US; VENUL12040 /Venovo 8F 12/40/1200mm US; VENUL12060 /Venovo 8F 12/60/1200mm US; VENUL12080 /Venovo 8F 12/80/1200mm US; VENUL12100 /Venovo 8F 12/100/1200mm US; VENUL12120 /Venovo 8F 12/120/1200mm US; VENUL12140 /Venovo 8F 12/140/1200mm US; VENUL12160 /Venovo 8F 12/160/1200mm US; VENUM10040 /Venovo 8F 10/40/800mm US; VENUM10060 /Venovo 8F 10/60/800mm US; VENUM10080 /Venovo 8F 10/80/800mm US; VENUM10100 /Venovo 8F 10/100/800mm US; VENUM10120 /Venovo 8F 10/120/800mm US; VENUM10140 /Venovo 8F 10/140/800mm US; VENUM10160 /Venovo 8F 10/160/800mm US; VENUM12040 /Venovo 8F 12/40/800mm US; VENUM12060 /Venovo 8F 12/60/800mm US; VENUM12080 /Venovo 8F 12/80/800mm US; VENUM12100 /Venovo 8F 12/100/800mm US; VENUM12120 /Venovo 8F 12/120/800mm US; VENUM12140 /Venovo 8F 12/140/800mm US; VENUM12160 /Venovo 8F 12/160/800mm US;

Code information (models, lots, serial numbers)

Show the full code information (37k characters)

Product Code/GTIN: VENEL10040 /00801741102004; VENEL10060 /00801741104664; VENEL10080 /00801741104671; VENEL10100 /00801741102035; VENEL10120 /00801741102042; VENEL10140 /00801741102059; VENEL10160 /00801741102066; VENEL12040 /00801741102141; VENEL12060 /00801741102158; VENEL12080 /00801741102172; VENEL12100 /00801741102189; VENEL12120 /00801741102196; VENEL12140 /00801741102202; VENEL12160 /00801741102219; VENEM10040 /00801741101939; VENEM10060 /00801741101946; VENEM10080 /00801741101953; VENEM10100 /00801741101960; VENEM10120 /00801741101977; VENEM10140 /00801741101984; VENEM10160 /00801741101991; VENEM12040 /00801741102073; VENEM12060 /00801741102080; VENEM12080 /00801741102097; VENEM12100 /00801741102103; VENEM12120 /00801741102110; VENEM12140 /00801741102127; VENEM12160 /00801741102134; VENUL10040 /00801741103209; VENUL10060 /00801741103216; VENUL10080 /00801741103223; VENUL10100 /00801741103230; VENUL10120 /00801741103247; VENUL10140 /00801741103254; VENUL10160 /00801741103261; VENUL12040 /00801741103346; VENUL12060 /00801741103353; VENUL12080 /00801741103360; VENUL12100 /00801741103377; VENUL12120 /00801741103384; VENUL12140 /00801741103391; VENUL12160 /00801741103407; VENUM10040 /00801741103131; VENUM10060 /00801741103148; VENUM10080 /00801741103155; VENUM10100 /00801741103162; VENUM10120 /00801741103179; VENUM10140 /00801741103186; VENUM10160 /00801741103193; VENUM12040 /00801741103278; VENUM12060 /00801741103285; VENUM12080 /00801741103292; VENUM12100 /00801741103308; VENUM12120 /00801741103315; VENUM12140 /00801741103322; VENUM12160 /00801741103339;  U.S. Device; Serial Numbers: ANEQ4079; ANEQ4090; ANEQ4091; ANEQ4101; ANEQ4104; ANEQ4105; ANEQ4109; ANEQ4978; ANEQ4979; ANEQ4980; ANEQ4996; ANEQ4997; ANEQ5023; ANER0477; ANER0478; ANER0490; ANER0491; ANER0510; ANER0515; ANER0532; ANER1502; ANER1528; ANER1529; ANER1530; ANER1533; ANER2795; ANER2796; ANER2797; ANER2812; ANER2813; ANER2814; ANER2815; ANER2816; ANER2838; ANER2841; ANER2855; ANER4217; ANER4228; ANER4236; ANER4238; ANER4239; ANER5126; ANER5127; ANER5136; ANER5141; ANER5142; ANER5153; ANER5159; ANER5160; ANER5169; ANER5175; ANES2223; ANES2224; ANES2233; ANES2234; ANES2235; ANES2236; ANES3045; ANES3050; ANES3067; ANES4102; ANES4103; ANES4111; ANES4112; ANES4113; ANES4114; ANES4115; ANES4116; ANES4117; ANES4118; ANET0561; ANET0566; ANET0567; ANET0568; ANET1490; ANET1508; ANET1509; ANET2533; ANET2534; ANET2535; ANET2536; ANET2537; ANET2540; ANET2541; ANET3797; ANET3798; ANET3799; ANET3800; ANET3814; ANET3815; ANET3816; ANET3817; ANET3818; ANET3877; ANET3878; ANEU0748; ANEU1594; ANEU1601; ANEU1602; ANEU1603; ANEU1604; ANEU1605; ANEU1606; ANEU2442; ANEU3275; ANEU3276; ANEU3277; ANEU3278; ANEU3284; ANEU3285; ANEU3286; ANEU3287; ANEU3288; ANEU3289; ANEU3290; ANEU3306; ANEU4231; ANEU4232; ANEU4238; ANEU4239; ANEU4240; ANEU4246; ANEU4253; ANEU4256; ANEU4260; ANEU4261; ANEU4264; ANEU4265; ANEV0586; ANEV0587; ANEV0591; ANEV0594; ANEV0598; ANEV0599; ANEV0603; ANEV0606; ANEV0611; ANEV0614; ANEV1714; ANEV1715; ANEV1716; ANEV1717; ANEV1718; ANEV1737; ANEV1738; ANEV1739; ANEV1740; ANEV3078; ANEV3079; ANEV3080; ANEV3081; ANEV3082; ANEV3494; ANEV4019; ANEV4020; ANEV4021; ANEV4022; ANEV4023; ANEV4024; ANEV4025; ANEV4043; ANEV4044; ANEV4045; ANEV4046; ANEV4071; ANEV4076; ANEW0553; ANEW0554; ANEW0560; ANEW0561; ANEW0562; ANEW0563; ANEW0564; ANEW0565; ANEW0566; ANEW0599; ANEW0600; ANEW2005; ANEW2006; ANEW2007; ANEW2028; ANEW2029; ANEW2030; ANEW2031; ANEW2032; ANEW2033; ANEW2034; ANEW2239; ANEW2772; ANEW2773; ANEW3378; ANEW3379; ANEW3380; ANEW3381; ANEW3394; ANEW3395; ANEW3396; ANEW3397; ANEW3422; ANEW4229; ANEW4230; ANEW4231; ANEW4232; ANEW4233; ANEW4242; ANEW4243; ANEW4244; ANEW4245; ANEW4246; ANEW4247; ANEW4276; ANEW4280; ANEW4282; ANEW4291; ANEW4297; ANEW4298; ANEW4300; ANEX0704; ANEX0711; ANEX0713; ANEX0720; ANEX0729; ANEX0732; ANEX0735; ANEX0736; ANEX0739; ANEX0740; ANEX0741; ANEX0742; ANEX0743; ANEX0744; ANEX0745; ANEX1737; ANEX1743; ANEX1744; ANEX1746; ANEX1747; ANEX1750; ANEX1752; ANEX1753; ANEX1754; ANEX1755; ANEX3120; ANEX3121; ANEX3123; ANEX3131; ANEX3135; ANEX3136; ANEX3137; ANEX3144; ANEX3145; ANEX3147; ANEX3148; ANEX3149; ANEX3151; ANEX3153; ANEX3154; ANEX3156; ANEX4737; ANEX4740; ANEX4744; ANEX4745; ANEX4751; ANEX4753; ANEX4761; ANEX4762; ANEX4764; ANEX4773; ANEX4779; ANEX4780; ANEX4782; ANEX4783; ANEX4784; ANEX4785; ANEX4786; ANEX4787; ANEX4788; ANEX4789; ANEX6042; ANEX6046; ANEX6049; ANEX6056; ANEX6057; ANEX6065; ANEX6066; ANEX6068; ANEY1246; ANEY1247; ANEY1251; ANEY1257; ANEY1258; ANEY1259; ANEY1262; ANEY1263; ANEY1264; ANEY1265; ANEY1266; ANEY1267; ANEY1268; ANEY1269; ANEY1278; ANEY1279; ANEY1280; ANEY2049; ANEY2050; ANEY2051; ANEY2052; ANEY2053; ANEY2054; ANEY2055; ANEY2056; ANEY3241; ANEY3243; ANEY3245; ANEY3255; ANEY3256; ANEY3257; ANEY3258; ANEY3261; ANEY3537; ANEY3538; ANEY4346; ANEY4354; ANEY4355; ANEY4356; ANEY4357; ANEY4358; ANEY4562; ANEZ0957; ANEZ0958; ANEZ0959; ANEZ0967; ANEZ0972; ANEZ0973; ANEZ0974; ANEZ0980; ANEZ0996; ANEZ0997; ANEZ1000; ANEZ1001; ANEZ1002; ANEZ1003; ANEZ1004; ANEZ2107; ANEZ2112; ANEZ2113; ANEZ2121; ANEZ2122; ANEZ2123; ANEZ2124; ANEZ2137; ANEZ2138; ANEZ2139; ANEZ2140; ANEZ3453; ANEZ3454; ANEZ3455; ANEZ3502; ANEZ3939; ANEZ3949; ANFN1482; ANFN1483; ANFN1484; ANFN1485; ANFN1486; ANFN1487; ANFN1488; ANFN1490; ANFN1492; ANFN1493; ANFN2809; ANFN2817; ANFN2821; ANFN2822; ANFN2824; ANFN2825; ANFN2835; ANFN2836; ANFN2837; ANFN2838; ANFN2839; ANFN2840; ANFN2841; ANFN2842; ANFN2843; ANFN3989; ANFN3990; ANFN3991; ANFN3992; ANFN3993; ANFN3994; ANFN3999; ANFN4000; ANFN4001; ANFN4002; ANFP0891; ANFP0896; ANFP0897; ANFP0915; ANFP0923; ANDS0730; ANDS0722; ANDS0729; ANDS1275; ANDS1312; ANDS1302; ANDS2454; ANDS4842; ANDS4851; ANDS4854; ANDS4852; ANDT1073; ANDT1713; ANDU2241; ANDT1721; ANDS4853; ANDT2908; ANDT2907; ANDT2906; ANDT2905; ANDU0121; ANDU1126; ANDU1125; ANDU0123; ANDU0122; ANDV0714; ANDU1894; ANDU1906; ANDU1861; ANDU1862; ANDU1890; ANDU1856; ANDU1884; ANDU1873; ANDU1124; ANDU1114; ANDU1905; ANDU1878; ANDU1867; ANDU3084; ANDU3144; ANDU3145; ANDU3139; ANDU3140; ANDU3104; ANDU3098; ANDU3128; ANDU3103; ANDU3127; ANDU3089; ANDV3781; ANDU3090; ANDV1215; ANDV1227; ANDU4041; ANDV1261; ANDU3111; ANDV3603; ANDV3602; ANDV3606; ANDV3576; ANDV3604; ANDV3601; ANDV4566; ANDV4554; ANDV4551; ANDV4567; ANDV4546; ANDV4552; ANDV4549; ANDV4550; ANDV4553; ANDV4568; ANDV4545; ANDV3600; ANDV4555; ANDV1234; ANDV3605; ANDV4547; ANDW2044; ANDW2042; ANDW2028; ANDW2043; ANDW2046; ANDW2047; ANDW2045; ANDX1371; ANDX1388; ANDX1429; ANDX1428; ANDW3626; ANDX1427; ANDX1423; ANDX1426; ANDV4569; ANDX1425; ANDW3625; ANDW4360; ANDW4367; ANDW4383; ANDX1430; ANDW4369; ANDW4400; ANDX1424; ANDW4340; ANDX0708; ANDW4419; ANDX0725; ANDX0720; ANDW4402; ANDW4353; ANDX2191; ANDX2172; ANDX2171; ANDX2192; ANDX2190; ANDX3082; ANDX3103; ANDX3105; ANDX3102; ANDX3101; ANDX3086; ANDX3081; ANDX2173; ANDX3100; ANDX3085; ANDX3083; ANDX2193; ANDX3104; ANDX3099; ANDX3084; ANDX3106; ANDX3107; ANDX5400; ANDX5399; ANDY1247; ANDZ0784; ANDY2488; ANDY2486; ANDY2487; ANDY2470; ANDZ2797; ANDZ2796; ANDZ1376; ANDZ2803; ANDY3323; ANDY3310; ANDZ1255; ANDY3311; ANDY3324; ANDZ1243; ANDZ1254; ANDZ1372; ANDZ1357; ANDZ1373; ANDY3325; ANDZ1358; ANDZ1375; ANDY3326; ANDZ1256; ANDZ1244; ANDZ1377; ANDZ1258; ANDZ1374; ANDZ2326; ANDZ1257; ANDZ2325; ANDZ3877; ANDZ3844; ANDZ3870; ANDZ3861; ANEN1090; ANEN1107; ANEN1112; ANEN1088; ANEN1111; ANEN1093; ANEN1108; ANEN1109; ANEN1094; ANEN1089; ANEN1110; ANEN2194; ANEN2193; ANEN2189; ANEN2190; ANEN2185; ANEN2182; ANEN2186; ANEN2187; ANEN3444; ANEN3421; ANEN3438; ANEN2188; ANEN3430; ANEN3436; ANEN3422; ANEN3441; ANEN3433; ANEN2180; ANEN3423; ANEN3431; ANEN4928; ANEN4916; ANEP0914; ANEP0919; ANEP0915; ANEP0912; ANEP0906; ANEP0918; ANEP2441; ANEP2455; ANEP2450; ANEP2442; ANEP2461; ANEQ3077; ANEP2443; ANEP3957; ANEP3963; ANEP3944; ANEP3974; ANEP3967; ANEP3984; ANEP3969; ANEP3942; ANEP3949; ANEQ3078; ANEP3985; ANEP5356; ANEP5334; ANEP5340; ANEP5386; ANEP5359; ANEP5358; ANEP5338; ANEP5365; ANEP5380; ANEP5385; ANEP5361; ANEP5374; ANEP5341; ANEP5335; ANEP5352; ANEQ1378; ANEQ1397; ANEQ1379; ANEQ1399; ANEQ1398; ANEQ1376; ANEQ1377; ANEQ1401; ANEQ1396; ANEQ1380; ANEQ1375; ANEQ2773; ANEQ2774; ANEQ1400; ANEQ2776; ANEQ2775; ANEQ2756; ANEQ2757; ANEQ2755; ANEQ4078; ANFQ2793;  O.U.S. Device; Serial Numbers: ANDU2240; ANES2208; ANDU2240; ANDY3304; ANEP2429; ANEP5328; ANEP2429; ANDZ3827; ANDN0820; ANDU2240; ANEP2429; ANDU2240; ANEP2429; ANEP2429; ANDU2240; ANEP5328; ANEP2429; ANES1042; ANDZ2288; ANEP5328; ANDU2240; ANDY3304; ANDZ3827; ANEP2429; ANDN0820; ANES1042; ANDZ2288; ANEP2429; ANEP5328; ANEP2429; ANEP5328; ANES1042; ANDU2240; ANDY3304; ANDZ2288; ANDZ3827; ANDZ3827; ANCV0493; ANDU2240; ANDU2240; ANES1042; ANES1042; ANEP5328; ANEP2429; ANEP2429; ANDN0820; ANDU2240; ANEP5328; ANDY3304; ANDN0820; ANDZ2288; ANDN0820; ANEP5328; ANCV0493; ANDZ3827; ANEP2429; ANEP5328; ANES2208; ANEP5328; ANES1042; ANES2208; ANEY3214; ANEY4323; ANFP2666; ANFP2666; ANDU2240; ANEY3214; ANEY3214; ANEY3214; ANEY3214; ANEP5328; ANEP2429; ANDR1257; ANDR2270; ANDR2270; ANDU1102; ANDV1222; ANDY2458; ANDY2458; ANDY3301; ANDY3301; ANDZ1338; ANDZ3819; ANEN1079; ANEN3405; ANER5167; ANEX0679; ANEZ2095; ANEZ3410; ANFN3972; ANCR2676; ANDR2270; ANDV1222; ANDV2517; ANDY2458; ANEZ3410; ANEZ3410; ANEZ2095; ANDX5352; ANEQ4065; ANEQ4961; ANEN3405; ANDV1222; ANDX5352; ANDZ1338; ANDV2517; ANDV1222; ANFN3972; ANEN3405; ANFN3972; ANFN3972; ANFN3972; ANEN3405; ANDV2517; ANDY2458; ANDR2270; ANEQ4065; ANDV2517; ANEQ4065; ANEN3405; ANEQ4961; ANFN3972; ANDV2517; ANEZ2095; ANCV1579; ANEN3405; ANDR2270; ANEN3405; ANDX5352; ANDZ3819; ANEN1079; ANEQ4961; ANER5167; ANEX0679; ANEX4720; ANEY4324; ANEX4720; ANEY4324; ANEZ0921; ANEZ0921; ANEZ2095; ANEZ2095; ANEZ3410; ANDR2270; ANEQ4065; ANEQ4961; ANDR2270; ANEQ4961; ANEQ4961; ANEQ4961; ANEQ4065; ANEQ4065; ANDR1257; ANCV1579; ANDZ1338; ANEQ4065; ANER5167; ANDR1257; ANDV1222; ANDV2517; ANDX5352; ANDX5352; ANDY2458; ANDR1257; ANDR1257; ANDR1257; ANDZ1338; ANDZ1338; ANCV1579; ANDY2458; ANEQ4065; ANFN3972; ANEX4720; ANEZ0921; ANFN3972; ANEQ4961; ANDV2517; ANDY2458; ANDQ3035; ANDQ3035; ANDQ3035; ANDQ3877; ANDU1105; ANDW4412; ANDX0715; ANDX5308; ANDZ2292; ANDZ3826; ANEN1080; ANEQ4066; ANES2209; ANET2531; ANET2531; ANEX1699; ANEX1699; ANEY4325; ANEZ0982; ANBY2401; ANCT0397; ANCT0397; ANDZ1330; ANES2209; ANDX5308; ANEY4325; ANDW4412; ANCW4225; ANDV0287; ANEQ2730; ANEQ2730; ANEQ2730; ANEQ4066; ANEX1699; ANEZ0982; ANEX1699; ANFP0878; ANFP0878; ANET2531; ANDQ3035; ANDQ3035; ANDQ3877; ANEQ4066; ANET2531; ANEN1080; ANEY4325; ANDX5308; ANDX5308; ANDQ3035; ANEQ2730; ANCW3270; ANES2209; ANDX5308; ANDZ1330; ANDZ2292; ANDZ3826; ANEN1080; ANEX1699; ANEY3212; ANEY4325; ANEZ0982; ANFP0878; ANET2531; ANEX1699; ANEX1699; ANEY3212; ANEY4325; ANEZ0982; ANDQ3035; ANDQ3035; ANEZ0982; ANEY4325; ANCW3270; ANCW4225; ANEY4325; ANET2531; ANES2209; ANET2531; ANDQ3035; ANDQ3877; ANDZ3826; ANDQ3877; ANES2209; ANET2531; ANDV0287; ANDX0715; ANDX5308; ANDZ1330; ANDZ2292; ANDQ3877; ANDX5308; ANEN1080; ANEY4325; ANDQ3877; ANEQ4066; ANEZ0982; ANCW3270; ANEQ4066; ANES2209; ANEQ2730; ANEY4325; ANDQ3041; ANCU2726; ANCX3902; ANDQ3880; ANCX3902; ANDQ3041; ANDQ3041; ANEP5323; ANEP5323; ANEQ2731; ANEP5323; ANEP5323; ANEQ4067; ANDQ3880; ANDY3305; ANDQ3880; ANCX3902; ANDY3305; ANEP0856; ANEP0856; ANEP0856; ANEP5323; ANEQ2731; ANES2210; ANES2210; ANEZ0922; ANFN3973; ANFN3973; ANEZ0922; ANFN1442; ANFN3973; ANFN3973; ANDQ3041; ANCX3902; ANDY3305; ANDY3305; ANEQ2731; ANDY3305; ANEP5323; ANDQ3880; ANDY3305; ANCX3902; ANDQ3041; ANDQ3041; ANDY3305; ANDQ3880; ANDQ3880; ANDQ3880; ANEQ2731; ANER0516; ANCX3902; ANDQ3880; ANEP5323; ANEP5323; ANEP5323; ANDQ3880; ANEP0856; ANEP5323; ANEZ0922; ANFN3973; ANDQ3023; ANDT1058; ANDT1058; ANDT1058; ANDU4075; ANDU4075; ANDZ1348; ANEN1081; ANEN1081; ANEP0857; ANEP0857; ANEP3924; ANET3794; ANEX0680; ANEX0680; ANEX0680; ANEZ2096; ANEP3924; ANEZ2096; ANEP3924; ANEP3924; ANDQ3023; ANDQ3023; ANEP3924; ANEP3924; ANDU4075; ANDQ3023; ANDU4075; ANDZ1348; ANCU3706; ANDZ2290; ANCU3706; ANDZ2290; ANDZ2290; ANDZ2290; ANDZ2296; ANEP5332; ANDZ2296; ANCU3716; ANEP5332; ANDZ2296; ANDZ2296; ANDR1253; ANDU2239; ANEQ2716; ANDU2239; ANES2187; ANCU0847; ANDX5374; ANDW4406; ANDW4406; ANES2187; ANDU2239; ANCU0847; ANDR1253; ANEQ2716; ANDR1253; ANDW4406; ANES2187; ANCY3323; ANDU2239; ANDU2239; ANDW4406; ANDU2239; ANES2187; ANDW4406; ANEP3938; ANDQ3029; ANDQ3029; ANDW4406; ANDX5374; ANDQ3029; ANDR1253; ANDR1253; ANDX5374; ANES2187; ANDR1253; ANDR1253; ANEQ2716; ANDW4406; ANCU0847; ANCU0847; ANDR1253; ANDU2239; ANDU2239; ANDW4406; ANDW4406; ANEQ2716; ANCU0847; ANCY3323; ANDR1253; ANDR1253; ANDX5374; ANDX5374; ANDU2239; ANEQ2716; ANES2187; ANDR1253; ANDU2239; ANDW4406; ANEP3938; ANDR1253; ANDU2239; ANDQ3037; ANDQ3037; ANDT1057; ANDT1057; ANDW4338; ANDW4338; ANDX5309; ANDX5309; ANDZ2308; ANEP0852; ANEP0852; ANER5149; ANET3785; ANEZ0906; ANDT1057; ANDQ3037; ANDX5309; ANEQ4945; ANDT1057; ANDT1057; ANDX5309; ANCW3260; ANDT1057; ANEP3936; ANEP3936; ANDX5309; ANEQ4945; ANDT1057; ANEX3092; ANEQ4036; ANCY3332; ANDU4074; ANDX5309; ANDZ2308; ANEQ4036; ANEZ0906; ANDZ2308; ANEP3936; ANDX5309; ANEP0852; ANEQ4945; ANEQ4945; ANET3785; ANEZ0906; ANFP2368; ANEQ4945; ANET3785; ANEX0671; ANEX0671; ANEX3092; ANEX3092; ANEZ0906; ANDQ3037; ANDX5309; ANDQ3037; ANDT1057; ANDX5309; ANDX5309; ANDZ2308; ANEP3936; ANEP3936; ANEP3936; ANEQ4036; ANEQ4036; ANEQ4036; ANDT1057; ANDT1057; ANEZ0906; ANEQ4945; ANDT1057; ANEQ4036; ANDW4338; ANCR1284; ANDZ2308; ANER5149; ANET3785; ANEX0671; ANER5149; ANDX5309; ANEX3092; ANDX5309; ANER5149; ANEP3936; ANDQ3037; ANEQ4945; ANEQ4945; ANDX5309; ANDU4074; ANEZ3430; ANEZ0906; ANCY3303; ANDP3952; ANDP3952; ANDP3952; ANDU4027; ANDU4027; ANDV4526; ANDW4339; ANDW4339; ANDX5324; ANDX5385; ANDZ3818; ANDZ3818; ANEN1086; ANEQ2717; ANEQ4946; ANEQ4946; ANER5165; ANET2530; ANEU0741; ANEU3303; ANEX0749; ANEZ0907; ANEZ0907; ANCQ1427; ANCQ1427; ANET2530; ANFN3944; ANDZ3818; ANEQ2717; ANEQ4037; ANCQ1427; ANEQ2717; ANEQ2717; ANER5165; ANCQ1427; ANDW4339; ANDX5385; ANCQ1427; ANCU1844; ANEX0749; ANDZ3818; ANEQ2717; ANEQ2717; ANEQ2717; ANET2530; ANFN3944; ANEU0741; ANEU3303; ANEX0749; ANEX0749; ANEZ0907; ANDP3952; ANDQ3871; ANDQ3871; ANDX5324; ANDX5324; ANDX5324; ANDX5385; ANDZ3818; ANEQ4037; ANEQ4037; ANER5165; ANER5165; ANEZ0907; ANFN3944; ANCY3303; ANEZ0907; ANEU0741; ANEU3303; ANDU4027; ANDX5324; ANDX5324; ANEN1086; ANDX5385; ANDZ3818; ANDX5385; ANEQ2717; ANET2530; ANEQ4037; ANDX5324; ANEQ4037; ANEQ2717; ANER5165; ANFN3944; ANER5165; ANEX0749; ANDX5324; ANDZ3818; ANEQ2717; ANEQ4037; ANDT0505; ANDT0505; ANDT1054; ANDT1054; ANDY2469; ANDY2469; ANDY4963; ANDY4963; ANDY4963; ANDY4963; ANDZ1345; ANEW2047; ANEW4217; ANEY4322; ANEY4322; ANFN1431; ANDT1054; ANEP2417; ANDZ1345; ANDT1054; ANDT1054; ANDT0505; ANDT1054; ANDT1054; ANDT1054; ANDY2469; ANEP2417; ANDY4963; ANEP0862; ANDY2469; ANEP2417; ANEW2047; ANEW4217; ANEY3209; ANEY3209; ANEY4322; ANDN0807; ANDQ0320; ANDQ1722; ANDQ3030; ANDQ3030; ANDT0505; ANDT0505; ANDT1054; ANDT1054; ANDT1054; ANDY2469; ANDY4963; ANEP0862; ANEP2417; ANDT1054; ANDT1054; ANEY4322; ANDT1054; ANEP0862; ANEP0862; ANEW4217; ANDT0505; ANDT1054; ANCU3703; ANDT1054; ANDY4963; ANDY4963; ANEP0862; ANFN1431; ANDT0505; ANCR2675; ANDQ0320; ANDQ1722; ANDT0505; ANEP0862; ANEP0862; ANEP0862; ANEP2417; ANFN1431; ANDN0808; ANDQ1726; ANEQ1347; ANEQ1347; ANDQ3031; ANEQ1347; ANES2188; ANDQ1726; ANDQ3031; ANDN0808; ANDQ1726; ANDQ3031; ANEQ1347; ANEQ1347; ANEQ1347; ANDQ3031; ANDQ3031; ANDQ3031; ANEQ1347; ANEQ1347; ANDQ3031; ANDQ3031; ANDN0808; ANES2188; ANDQ1726; ANEQ1347; ANDU1108; ANDY3280; ANCY1571; ANDU1108; ANEN3385; ANEN3385; ANEQ1348; ANEN3385; ANCY1571; ANEN3385; ANCY1571; ANCY1571; ANCY1571; ANEN3385; ANDU1108; ANEN3385; ANDW1996; ANCW3261; ANET1855; ANDN0606; ANDR1266; ANDR1266; ANDW1996; ANDW1996; ANDW1996; ANDY2448; ANDW1996; ANET1855; ANDW1996; ANDY2448; ANCW3261; ANCW3261; ANDY2448; ANET1855; ANDY2448; ANDS4772; ANDS4772; ANDS4772; ANDS4772; ANDT0508; ANDW4345; ANDY4968; ANDY4968; ANDY4968; ANEQ2732; ANEZ3411; ANDQ3024; ANDR1268; ANDS4772; ANDT0508; ANDX5315; ANDX5384; ANDY4968; ANDZ3817; ANDZ3817; ANEN4877; ANEN4877; ANEP0861; ANER5152; ANET0555; ANET3795; ANEW3410; ANEZ3411; ANFN3974; ANDT0508; ANCT4323; ANCY2200; ANER5152; ANDZ3817; ANDX5384; ANEQ2732; ANET0555; ANFN1443; ANFN3974; ANDT0508; ANEQ2732; ANCS1161; ANDS4772; ANEW3410; ANEY1223; ANDS4772; ANDX5315; ANEQ2732; ANDT0508; ANDZ3817; ANDX5315; ANEQ2732; ANEQ2732; ANDX5384; ANDX5384; ANDX5315; ANFN1443; ANEQ2732; ANDZ3817; ANEP0861; ANEQ2732; ANEW3410; ANEY1223; ANFP0911; ANDZ3817; ANFP2377; ANFP0911; ANEQ2732; ANER5152; ANEQ2732; ANDZ3817; ANDY4968; ANEQ2732; ANCY3322; ANDS4772; ANDQ3024; ANEQ2732; ANET0555; ANER5152; ANER5152; ANEY1223; ANFP0911; ANDQ3024; ANER5152; ANEW3410; ANEQ2732; ANDX5384; ANER5152; ANEY1223; ANER5152; ANCS1161; ANEY1223; ANDQ3024; ANDS4772; ANDX5315; ANEQ2732; ANEQ2732; ANEQ2732; ANER5152; ANEW3410; ANEQ2732; ANFN3974; ANFP2377; ANDX5315; ANFP0911; ANDX5384; ANEQ2732; ANEP0861; ANDZ3817; ANFN1443; ANDQ3024; ANDQ3024; ANDR1268; ANDT0508; ANDT0508; ANDX5315; ANDX5315; ANDX5384; ANEN4877; ANER5152; ANET0555; ANEY1223; ANEZ3411; ANEZ3411; ANFN1443; ANFP0911; ANDQ1716; ANDX5384; ANDX5384; ANET0555; ANET0555; ANET3795; ANEY1223; ANEZ3411; ANDT0508; ANDT0508; ANER5152; ANEN4877; ANEQ2732; ANEQ2732; ANEW3410; ANEZ3411; ANCS1161; ANDS4772; ANDX5315; ANDX5384; ANER5152; ANFN1443; ANCY3322; ANEQ2732; ANEQ2732; ANER5152; ANER5152; ANDX5315; ANEW3410; ANDY4968; ANDR1268; ANDW4345; ANDX5315; ANDX5384; ANDX5384; ANDY4968; ANDZ2301; ANDZ3817; ANDS4772; ANFP2377; ANDX5315; ANCY2200; ANER5152; ANEY1223; ANFN1443; ANDT0508; ANDZ3817; ANEW3410; ANEZ3411; ANEP0861; ANDX5384; ANDX5384; ANDQ3024; ANEP0861; ANDQ1716; ANEQ2732; ANFP2377; ANEQ2732; ANEW3410; ANDS3973; ANDS3973; ANDS4774; ANDS4774; ANDS4774; ANDS4774; ANDT2946; ANDW1997; ANDW4361; ANDX0760; ANDX0760; ANDX0760; ANDY1234; ANEP2420; ANDQ5251; ANDS2451; ANDS3973; ANDS4774; ANDW4361; ANDX2782; ANDZ3820; ANEN1082; ANEN2140; ANER1507; ANEZ0923; ANFN1444; ANFN1444; ANFN2844; ANDQ5251; ANDS3973; ANFN3975; ANDZ3820; ANEN4896; ANER1507; ANER1507; ANDQ5251; ANDQ3872; ANDX5373; ANEQ4962; ANER1507; ANDQ3025; ANDZ2291; ANDX5373; ANEQ2733; ANER1507; ANCN1543; ANEQ1353; ANDT2946; ANDT2946; ANDX2782; ANDX5373; ANEQ2733; ANDT2946; ANDX3060; ANDX5373; ANEQ1353; ANEQ1353; ANEQ4962; ANER1507; ANDS3973; ANDW1997; ANEQ2733; ANFN1444; ANFN2844; ANDT2946; ANEQ2733; ANER0517; ANDS3973; ANDS4774; ANER1507; ANER1507; ANDZ2291; ANFP2318; ANES4086; ANDS3973; ANDX0760; ANDX5373; ANER1507; ANFN1444; ANES2211; ANER1507; ANDS3973; ANFP2318; ANDT2946; ANDZ3820; ANEQ2733; ANFN2844; ANEN4896; ANER1507; ANDT2946; ANDW4361; ANDP1356; ANEN4896; ANFN1444; ANEQ2733; ANER1507; ANDZ3820; ANER1507; ANDT2946; ANFN3975; ANDX2782; ANDX5373; ANDW4361; ANEZ2098; ANFN1444; ANDZ2291; ANER1507; ANEQ1353; ANDQ5251; ANFN1444; ANDX5373; ANER1507; ANER1507; ANER1507; ANER1507; ANFN3975; ANDS3973; ANEN4896; ANDQ3872; ANDW4361; ANDX2782; ANEQ4962; ANDT2946; ANER1507; ANDQ5251; ANDX5373; ANDY1234; ANDX5373; ANER1507; ANDS4774; ANER1507; ANDS4774; ANDX0760; ANDY1234; ANER1507; ANER1507; ANDW1997; ANDX0760; ANDX5373; ANDZ3820; ANER1507; ANER1507; ANDV0687; ANDY1234; ANER1507; ANFN2844; ANFP2318; ANEZ2098; ANDU1877; ANDW1997; ANDX5373; ANDW4361; ANDS3973; ANEN4896; ANDW4361; ANDS2451; ANDP1356; ANDW4361; ANER1507; ANEQ4962; ANDQ3025; ANDS2451; ANES2211; ANES4086; ANES4086; ANEZ0923; ANEZ2098; ANER0517; ANFN1444; ANDQ3025; ANDQ3872; ANDS2451; ANDS2451; ANDS4774; ANDT2946; ANDT2946; ANDV0687; ANDW1997; ANDW3561; ANDW4361; ANDX0760; ANDX0760; ANDX2782; ANDX2782; ANDX2782; ANDX3060; ANDX5373; ANDY1234; ANDZ2291; ANDZ3820; ANDZ3820; ANEN2140; ANEQ1353; ANER1507; ANES1043; ANES1043; ANES2211; ANEZ0923; ANFN1444; ANFN1444; ANFN2844; ANFN3975; ANFN3975; ANFP0879; ANEN2140; ANEN4896; ANEP2420; ANEQ1353; ANEQ2733; ANES2211; ANES2211; ANEZ2098; ANEZ2098; ANEZ3412; ANFN1444; ANFP0879; ANFP2318; ANDQ3025; ANDQ5251; ANDS3973; ANDS4774; ANDT2946; ANDX0760; ANDX2782; ANDX3060; ANER1507; ANER1507; ANER1507; ANER1507; ANES2211; ANDW4361; ANDS3973; ANDV0687; ANES4086; ANDX5373; ANEP2420; ANES4086; ANDQ3872; ANER1507; ANDX5373; ANDS2451; ANDT2946; ANEN4896; ANER1507; ANDQ3025; ANDS4774; ANER1507; ANDQ3025; ANEP2420; ANER1507; ANDQ3025; ANDS2451; ANEN4896; ANEQ1353; ANEQ4962; ANER1507; ANFN1444; ANDS2451; ANER0517; ANER1507; ANDT2946; ANER1507; ANER1507; ANFP0879; ANDT2946; ANDY1234; ANDQ3025; ANDU1877; ANDV0687; ANDX5373; ANDY1234; ANES4086; ANEN2140; ANFP2317; ANFP2318; ANDS4774; ANDT2946; ANDW1997; ANEQ1353; ANDT2946; ANDX5373; ANCW1737; ANDQ3025; ANER1507; ANDT2946; ANDS4774; ANFN1444; ANDX3060; ANDY1234; ANDX5373; ANDT2946; ANDY1234; ANER0517; ANDP0508; ANDP0508; ANDP1356; ANDP1356; ANEP2420; ANEQ1353; ANEQ2733; ANEQ4962; ANER0517; ANER1507; ANEZ2098; ANFP2318; ANER0517; ANDT2946; ANDT2946; ANES4086; ANCX3883; ANCX3883; ANDV0702; ANDV2519; ANDV2519; ANDX0727; ANDX0727; ANDX2426; ANDX2426; ANDX3063; ANDX3063; ANDY3291; ANDY3291; ANDZ1331; ANDQ3033; ANDS0753; ANDS0753; ANDT4130; ANDV0702; ANDX2032; ANDY3291; ANDZ1331; ANDZ2297; ANDZ3824; ANDZ3824; ANEN4886; ANEP0860; ANER5168; ANEZ2099; ANEZ3952; ANFN3976; ANDW4378; ANEP0860; ANDV0702; ANDZ3824; ANEQ1354; ANFN3976; ANDX5307; ANDY1221; ANEQ1354; ANEP3934; ANDX2426; ANDY3291; ANEN2145; ANEP3934; ANEQ1354; ANDQ0322; ANDS0753; ANDX2426; ANEQ4069; ANEZ3952; ANEQ4069; ANFP2336; ANDZ2297; ANEQ4069; ANDV1233; ANDY1221; ANEP0860; ANEQ1354; ANFN3976; ANEQ1354; ANEP0860; ANFN3976; ANFN3976; ANEQ4069; ANFN3976; ANDT4130; ANFP0912; ANEQ4069; ANEP3934; ANEP3934; ANEQ4069; ANEP0860; ANEZ3952; ANDQ3033; ANEQ1354; ANCS3348; ANDT4130; ANEP0860; ANEP3934; ANEQ4069; ANDV0702; ANDY2464; ANEP0860; ANDV1233; ANEQ4069; ANCS3348; ANDV0702; ANDV1233; ANDX2426; ANDY3291; ANEN2145; ANEN4886; ANEP0860; ANDW4378; ANDS0753; ANEP0860; ANEQ1354; ANEQ1354; ANDS0753; ANDV0702; ANDX2426; ANDX2426; ANEN2145; ANEP0860; ANEP0860; ANEQ1354; ANER5168; ANDQ0322; ANDQ3033; ANDS0753; ANDS0753; ANDU3110; ANDV0300; ANDW4378; ANDW4378; ANDX0727; ANDX2426; ANDX5307; ANDX5307; ANDY1221; ANDY2464; ANDZ1331; ANDZ2297; ANEN2145; ANEN3389; ANEN4886; ANEP0860; ANEP3934; ANER5168; ANES2212; ANES2212; ANEZ0924; ANEZ2099; ANEZ2099; ANEZ2099; ANEZ3952; ANEZ3952; ANFN2787; ANFN3976; ANFN3976; ANFP0912; ANEZ0924; ANEZ3952; ANFP0912; ANDQ3033; ANDS0753; ANDT4130; ANDV1233; ANDY1221; ANEP3934; ANEP3934; ANEQ1354; ANEQ1354; ANEQ4069; ANEQ4069; ANEZ3952; ANDZ2297; ANEQ4069; ANEQ4069; ANDS0753; ANDT4130; ANDV0702; ANDX2032; ANDX5307; ANEN2145; ANEN3389; ANEQ4069; ANDQ0322; ANDS0753; ANDT4130; ANDU3110; ANEQ1354; ANEQ1354; ANER5168; ANDX5307; ANEQ1354; ANES2212; ANEN2145; ANEQ4069; ANDV2519; ANDX2426; ANDX5307; ANDY2464; ANDZ1331; ANDZ1331; ANDZ3824; ANEN1064; ANEN2145; ANEN2145; ANEQ1354; ANDZ1331; ANER5168; ANFP0912; ANFP2336; ANDV2519; ANEQ1354; ANES2212; ANDT4130; ANCX3883; ANDV0702; ANDY2464; ANER5168; ANDZ2297; ANEQ1354; ANEN3389; ANDT4130; ANDT4130; ANES2212; ANDQ0322; ANDT4130; ANDZ3824; ANEN3389; ANDV0702; ANDV1233; ANEZ0924; ANEZ3952; ANFN3976; ANCT4331; ANDQ0322; ANDU3110; ANEQ4069; ANEP0860; ANEQ4069; ANER5168; ANDV2519; ANEQ1354; ANDY1221; ANEQ1354; ANCX3884; ANDR0754; ANDS2461; ANDS3976; ANDS3976; ANDS3976; ANDS4839; ANDT4131; ANDY1231; ANDY3284; ANDY3284; ANDY3284; ANDR0754; ANDS2461; ANDS3976; ANDX2031; ANDX3062; ANDX5363; ANDX5363; ANDY1231; ANDZ3823; ANEN2147; ANEN3398; ANER1508; ANET0556; ANEZ3429; ANCS0655; ANEQ1355; ANER0518; ANEQ1355; ANDS3976; ANDS3976; ANDS4839; ANDX5323; ANDX5363; ANDX5363; ANER0518; ANDX5363; ANER0518; ANDX5363; ANER0518; ANER0518; ANET0556; ANER0518; ANER1508; ANFP2365; ANDS4839; ANDX5323; ANDX5323; ANDX5363; ANEQ1355; ANDU1889; ANDX2031; ANER1508; ANER1508; ANET0556; ANER0518; ANEQ1355; ANDX5363; ANDX5363; ANDT4131; ANER0518; ANDX5323; ANER1508; ANDS3976; ANDQ0321; ANDW1999; ANET0556; ANDS2461; ANDR0754; ANEQ1355; ANER1508; ANES4087; ANEZ2100; ANDQ0321; ANDQ3876; ANDS2461; ANDS2461; ANDS3976; ANDT4131; ANDT4131; ANDU1889; ANDW1999; ANDW1999; ANDX0739; ANDX0739; ANDX3062; ANDX5323; ANDX5323; ANDX5323; ANDX5363; ANDY3284; ANDY4974; ANDZ3823; ANEN2147; ANEN2147; ANEQ1355; ANEQ2734; ANEQ4963; ANER1508; ANES4087; ANES4087; ANET0556; ANET0556; ANEY2031; ANEY2031; ANEZ3429; ANEZ3429; ANFN1445; ANFN1445; ANFN3977; ANFN3977; ANFN3977; ANFP2365; ANEN3398; ANEQ1355; ANEQ4963; ANER1508; ANET0556; ANEY2031; ANEZ2100; ANFP2365; ANDQ3876; ANDX2031; ANDX5323; ANER0518; ANER0518; ANER0518; ANER1508; ANER1508; ANDU1889; ANDT4131; ANER0518; ANER0518; ANDQ0321; ANDS3976; ANDS4839; ANER1508; ANDX5363; ANDX5363; ANDS3976; ANDX5323; ANDQ3876; ANDT4131; ANDU1889; ANDW1999; ANDW4395; ANDW4395; ANDX5363; ANDX5363; ANDY3284; ANDY4974; ANDZ3823; ANET0556; ANDY3284; ANDR0754; ANDU1889; ANEN3398; ANER1508; ANET0556; ANFP2365; ANDS4839; ANDX2031; ANET0556; ANER0518; ANDX5363; ANDR0754; ANDX2031; ANDS4839; ANDS3976; ANDR0754; ANDX2031; ANDX5323; ANFN1445; ANDZ3823; ANDR0754; ANDS4839; ANDU1889; ANEN3398; ANER1508; ANDU1889; ANDS4839; ANDX5363; ANEN2147; ANER0518; ANDR0754; ANDR0754; ANDS4839; ANDZ3823; ANEN3398; ANEQ4963; ANER0475; ANER0518; ANER1508; ANER0518; ANDU4028; ANDV0275; ANDV0275; ANDW4413; ANDW4413; ANDX5301; ANDX5301; ANDY4972; ANDY4972; ANDU4028; ANDV0275; ANEN1083; ANEN1083; ANEQ2735; ANDW4413; ANEQ2735; ANER5120; ANEQ2735; ANDQ3028; ANES4088; ANEQ2735; ANES4088; ANER5120; ANES4088; ANES4088; ANDV0275; ANDQ3028; ANDU4028; ANEP5329; ANES4088; ANES4088; ANES4088; ANFP0880; ANFP0880; ANDQ3028; ANDW4413; ANEP5329; ANEP5329; ANES4088; ANES4088; ANDX5301; ANEN1083; ANEQ2735; ANEQ2735; ANES4088; ANDX5301; ANDV0275; ANDQ3028; ANDW4413; ANDW4413; ANDX0751; ANDX5301; ANDX5301; ANDY4972; ANER5120; ANDU4028; ANDV0275; ANDX5301; ANDX5301; ANER5120; ANER5120; ANES4088; ANER5120; ANDU4028; ANDX5301; ANDV0275; ANDV0275; ANDV0275; ANEQ2735; ANEQ2735; ANEN1083; ANEN1083; ANEN4883; ANEN4883; ANEP5329; ANEQ2735; ANEQ2735; ANFP0880; ANFP0880; ANFP0880; ANFP2331; ANER5120; ANDQ3027; ANDQ3027; ANDU4040; ANDU4040; ANDX5327; ANDX5327; ANDX5389; ANDX5389; ANEP5309; ANEP5309; ANEP5309; ANEQ1356; ANEQ2736; ANEP5309; ANEP5309; ANER0519; ANDQ3874; ANDX5327; ANER1535; ANEP5309; ANDQ3027; ANDX5327; ANEP5309; ANEQ1356; ANDX5327; ANDQ3874; ANDU1103; ANDU4040; ANDX5389; ANDZ1326; ANEP5309; ANER1535; ANER1535; ANER1535; ANDQ3874; ANDQ3874; ANDU1103; ANER1535; ANDU1103; ANCX3885; ANEP5309; ANEP5309; ANEQ1356; ANDU1103; ANEP5309; ANEP5309; ANEQ2736; ANDU1104; ANDY2452; ANDY2452; ANDY3299; ANEN4894; ANDQ3034; ANDQ3034; ANEN1084; ANEN1084; ANEQ4964; ANEN1084; ANEN1084; ANDY2452; ANDZ3825; ANEN1084; ANCQ4294; ANDQ3034; ANDU1104; ANDY2452; ANDZ3825; ANEN1084; ANER0520; ANER0520; ANER0520; ANER0520; ANCV1583; ANER0520; ANDU1104; ANEN1084; ANDT1055; ANDR1267; ANDT1055; ANDW1998; ANDW1998; ANDW1998; ANDX1402; ANDY4964; ANDY4964; ANES2189; ANET0542; ANEW4218; ANEX1697; ANEZ2081; ANEZ3398; ANFN3946; ANFP0857; ANDR1267; ANDT1055; ANEN2153; ANEP2424; ANDU4039; ANDT1055; ANDT0506; ANEZ3398; ANCW3262; ANDT0506; ANDU2242; ANEZ3398; ANDZ2307; ANDR1267; ANDW1998; ANEN2153; ANDU2242; ANDZ2307; ANEX6010; ANEZ0908; ANDU4039; ANFN3946; ANDZ2307; ANDT0506; ANDT1055; ANFP0857; ANDU4039; ANDT1055; ANDZ2307; ANEZ2081; ANEX6010; ANDT1055; ANEW4218; ANDT1055; ANDW1998; ANDT0506; ANDU2242; ANDU2242; ANDZ2307; ANEN2153; ANCS1158; ANFP0857; ANDT1055; ANEZ3398; ANDW1998; ANEP2424; ANDW1998; ANDY3289; ANDY3289; ANDY4964; ANDZ2307; ANET0542; ANEW3366; ANEX1697; ANEX6010; ANEX6010; ANEZ0908; ANEZ3398; ANFP0857; ANDT0506; ANDX1402; ANDY3289; ANDR1267; ANDT0506; ANDU2242; ANDU4039; ANEP2424; ANEP2424; ANEW4218; ANEX6010; ANEX6010; ANEZ2081; ANEZ3398; ANFP0857; ANDU4039; ANDT0506; ANDT1055; ANDW1998; ANEW4218; ANDU4039; ANDZ2307; ANEN2153; ANDZ2307; ANES2189; ANDU2242; ANDU2242; ANES2189; ANDP1357; ANDR1267; ANEN2153; ANEN2153; ANDR1267; ANDZ2307; ANFN3946; ANDY3289; ANDZ2307; ANDX1402; ANDZ2307; ANDR1267; ANDT1055; ANDX1402; ANEN2153; ANEN2153; ANDU4039; ANDU4039; ANEP2424; ANET0542; ANFP0857; ANCW3262; ANDT0506; ANDU2242; ANDU4039; ANDU4039; ANDZ2307; ANDU2242; ANDT0506; ANDU4039; ANDR1267; ANEZ2081; ANDR1267; ANDX1402; ANEP2424; ANES2189; ANDP2690; ANDU2242; ANCR1285; ANDY4964; ANEX6010; ANDU4039; ANEP2424; ANEX1697; ANEP2421; ANDQ3868; ANDS0752; ANDT1059; ANDX1401; ANDX2425; ANDX5362; ANDY1230; ANDZ1322; ANDZ2286; ANEN4897; ANEN4897; ANEP0867; ANEP2421; ANEV1711; ANEW4219; ANFN1432; ANDN0810; ANDS0752; ANDS0752; ANDV4534; ANDX5362; ANDZ1322; ANDZ2286; ANEQ4039; ANEN4897; ANEQ2718; ANCW1424; ANDT0507; ANDX1401; ANDQ3022; ANDV4534; ANFP0858; ANFN3947; ANFN3947; ANEW4219; ANEP3930; ANES4062 ANDX2425; ANDQ1713; ANEP2421; ANEP3930; ANEZ2082; ANDQ3022; ANFP2358; ANDV4534; ANEQ4039; ANER1514; ANDT0507; ANDV4534; ANDY1230; ANDV4534; ANDT1059; ANEP3930; ANEW4219; ANCW1424; ANER1514; ANCX1867; ANER1514; ANDX2425; ANFN3947; ANDY1230; ANDY1230; ANDZ1322; ANDY1230; ANEP0867; ANEQ4039 ANES2190; ANDZ1322; ANEP0867; ANEW3367; ANEW4219; ANES4062; ANES4062; ANEV1711; ANEW3367; ANEW4219; ANEX0746; ANEX1708; ANEY1228; ANFN1432; ANFN3947; ANFP0858; ANDQ1713; ANDQ3022; ANDQ3022; ANDQ3868; ANDX3057; ANDX5362; ANCV0494; ANDN0810; ANDT0507; ANDT0507; ANDX1401; ANDX1401; ANDX2425; ANDX5362; ANDY1230; ANDZ2286; ANEN4897; ANEP0867; ANEP3930; ANEQ2718; ANEQ4039; ANEQ4039; ANEQ4039; ANER1514; ANER1514; ANER1514; ANES2190; ANES4062; ANEX1708; ANEX3091; ANDT1059; ANDQ3868; ANEP3930; ANCU0848; ANDX1401; ANER1514; ANDQ3868; ANDX2425; ANDY1230; ANEQ4039; ANER1514; ANES4062; ANEY1228; ANDT0507; ANDV4534; ANDY1230; ANDN0810; ANDT0507; ANDT1059; ANDX2425; ANDZ1322; ANER1514; ANFP0858; ANDT0507; ANEW4219; ANEZ2082; ANER1514; ANEX3091; ANEW4219; ANCU0848; ANDX3057; ANDZ1322; ANER1514; ANES2190; ANDS0752; ANER1514; ANDT0507; ANDQ1713; ANDS0752; ANDX2425; ANDX3057; ANER1514; ANEX3091; ANFN3947; ANDX2425; ANDX2425; ANER1514; ANEX3091; ANDT0507; ANDX1401; ANDX2425; ANDV4534 ANEP3930; ANES4062; ANFP0858; ANDS0752; ANDV4534; ANDX1401; ANEQ4039; ANEQ4039; ANDX3057; ANEZ2082; ANES4062; ANEQ4039; ANDT0507; ANDX2425; ANES4062; ANEW4219; ANEP3930; ANEQ2718; ANCV0494; ANDQ1713; ANDQ1713; ANFN3947; ANDV4534; ANDX2425; ANEQ4039; ANFN3947; ANDQ1713; ANDT1059; ANEQ4039; ANER1514; ANDS0752; ANDS0752; ANDX3057; ANES2190; ANDZ1322; ANEP3930; ANES4062; ANFN3947; ANEQ4039; ANEX3091; ANER1514; ANES2190; ANDX2425; ANEQ4039; ANEP3930; ANES4062; ANDX3057; ANDX3057; ANDT0507; ANEX1708; ANEP5317; ANDR2271; ANDR2271; ANDV3567; ANDX1405; ANDX2034; ANDY4960; ANDZ1332; ANEP3939; ANES2191; ANET0543; ANEW4260; ANFN2783; ANFN3948; ANCW4216; ANDN0821; ANDN0821; ANDV4528; ANDZ2304; ANEN1070; ANEN3402; ANEQ4040; ANBY2393; ANCQ2599; ANDV4528; ANEQ4040; ANFN3948; ANFN3948; ANCW3263; ANEQ4040; ANEQ4040; ANDR2271; ANFN3948; ANFN3948; ANDP3957; ANEX3093; ANCW3263; ANEP3939; ANFN3948; ANDV4528; ANDV4528; ANDV4528; ANEQ4040; ANES2191; ANDV4528; ANDR2271; ANDV4528; ANCW4216; ANEP5317; ANCT1573; ANDV4528; ANDZ1332; ANEP5317; ANDZ2304; ANDZ2304; ANDZ2304; ANEP5317; ANEP5317; ANFN3948; ANDQ3039; ANDV4528; ANDZ1332; ANDZ1332; ANEP5317; ANET0543; ANET3786; ANET3786; ANEW4260; ANEW4260; ANEX3093; ANEX3093; ANEY1221; ANEY3210; ANEY3210; ANEY3210; ANFN2783; ANFN3948; ANDP3957; ANDQ1717; ANDR2271; ANCT4314; ANCV0496; ANDP3957; ANDQ3039; ANDV4528; ANDX1374; ANDY1238; ANDY4960; ANDZ1332; ANEP3939; ANEQ2719; ANEQ4040; ANEQ4040; ANEQ4040; ANEQ4040; ANER5118; ANER5118; ANES2191; ANEX3093; ANEX6157; ANEY1221; ANFN3948; ANDZ2304; ANDY1238; ANEQ2719; ANDX1374; ANDV4528; ANEQ4040; ANDP3957; ANDV4528; ANDX1374; ANDY1238; ANDP3987; ANDV3567; ANEP3939; ANER5118; ANEY3210; ANDZ2304; ANDV4528; ANEY3210; ANEZ3431; ANDZ2304; ANDV4528; ANDP3957; ANES2191; ANEX6157; ANEZ3431; ANER5118; ANDQ1717; ANET3786; ANEY1221; ANEX6157; ANFN3948; ANDV4528; ANCW4216; ANDQ1717; ANDX1405; ANDQ3039; ANDX1374; ANDX2034; ANDY1238; ANDZ1332; ANEN1070; ANER5118; ANEQ4040; ANDR2271; ANDQ1717; ANDR2271; ANDV4528; ANDV4528; ANEQ4040; ANEX6157; ANEX6157; ANET3786; ANEY1221; ANEP5317; ANEQ2719; ANEQ4040; ANDP3957; ANEZ3431; ANDQ3039; ANDX1405; ANEP5317; ANDX1374; ANDX1374; ANEQ4040; ANEQ4040; ANFN3948; ANEQ4040; ANCW3263; ANEQ2719; ANEQ4040; ANER5118; ANDQ1717; ANEP5317; ANEQ4040; ANES2191; ANDR2271; ANDV3567; ANEP5317; ANEQ4040; ANDV3567; ANDX2034; ANEN3402; ANDR2271; ANET3786; ANEX6157; ANET3786; ANDY1238; ANDZ2304; ANET3786; ANFN3948; ANEP2435; ANDQ1723; ANDQ3040; ANDU2243; ANDW4355; ANDW4355; ANDW4355; ANDX1406; ANDX2035; ANDX5316; ANDX5390; ANDZ2283; ANEN1071; ANEN4895; ANER1515; ANES4063; ANEX0672; ANEX6011; ANEZ3399; ANFN1433; ANCX0889; ANDP3959; ANDP3959; ANDP3959; ANDP3959; ANDX2035; ANDZ2283; ANEN1071; ANEQ2720; ANEQ4041; ANDW4355; ANEX0672; ANFN1433; ANDQ3040; ANDV1244; ANDX2035; ANDX2035; ANET3787; ANDQ3040; ANEQ2720; ANER1515; ANEY3211; ANEQ2720; ANDX5316; ANFN1433; ANCX0889; ANCX3891; ANDQ3879; ANDU2243; ANDU2243; ANDV1244; ANDX2035; ANDU2243; ANDX2035; ANDV1244; ANES4063; ANDQ1723; ANDV1244; ANEP2435; ANEQ2720; ANEX6011; ANDV1244; ANDQ1723; ANDQ3040; ANDZ2283; ANEQ2720; ANDQ5218; ANEN4895; ANEX3081; ANER1515; ANES4063; ANET3787; ANEW3412; ANEX3081; ANEX6011; ANEY3211; ANEZ3399; ANFN1433; ANFN3949; ANDQ0326; ANDQ1723; ANDQ3040; ANDQ3879; ANDP3959; ANDQ0326; ANDX5316; ANDX5316; ANDX5316; ANDX5390; ANDY3308; ANDZ2283; ANEQ2720; ANEQ4947; ANEQ4947; ANER1515; ANES4063; ANEX3081; ANEX6011; ANDU2243; ANDU2243; ANEQ4041; ANCX3891; ANDW4355; ANEN4895; ANDQ3879; ANDQ5218; ANDU2243; ANDV1244; ANDX5390; ANDZ2283; ANEN4895; ANEP2435; ANEQ4947; ANER1515; ANER1515; ANER1515; ANES4063; ANFN3949; ANDQ1723; ANDQ5218; ANDV1244; ANCQ1429; ANDZ2283; ANEQ4947; ANER1515; ANEY3211; ANDX1406; ANDX2035; ANDQ3040; ANCQ1429; ANDX5390; ANDX5390; ANEX6011; ANEY3211; ANDU2243; ANDX2035; ANER1515; ANDQ1723; ANDW4355; ANDX1406; ANER1515; ANEQ2720; ANCQ2600; ANDX1406; ANEQ4947; ANEQ4947; ANCN3578; ANDX2035; ANDP3959; ANDQ1723; ANDU2243; ANDX2035; ANDX2035; ANEQ2720; ANEQ4041; ANEQ4041; ANEQ4041; ANER1515; ANEX6011; ANES4063; ANEX6011; ANDZ2283; ANEQ2720; ANDX2035; ANEP2435; ANDX5316; ANEQ4041; ANEX3081; ANDW4355; ANER1515; ANEX6011; ANFN1433; ANDX5390; ANDQ3879; ANDX5390; ANEN1071; ANES4063; ANEX0672; ANEY3211; ANEQ4947; ANEX6011; ANDX1406; ANES4063; ANFN1433; ANEX6011; ANEZ3399; ANFN3949; ANEN2150; ANDS0751; ANDS0751; ANDS4788; ANDU2244; ANDV4533; ANDX3056; ANDX5361; ANDY3283; ANDY4965; ANDZ1327; ANDZ2287; ANEW4220; ANEW4220; ANEX1707; ANFN1434; ANFN2784; ANDP3950; ANDP3950; ANDX1373; ANDY4965; ANEN4880; ANDS4788; ANDS4823; ANEQ4042; ANCQ1430; ANDN0809; ANDS4788; ANDW4372; ANEQ4042; ANDY3283; ANEQ4042; ANER1516; ANEN2150; ANEN4880; ANCS1563; ANCW4217; ANDS4823; ANEN2150; ANFN2784; ANDS4788; ANDY4965; ANEP0877; ANER1516; ANDS4823; ANDS4788; ANDY3283; ANFN2784; ANDW4372; ANDZ2287; ANDZ2287; ANEX1707; ANDP3950; ANDX1373; ANDX3056; ANDY1229; ANDY2456; ANCW4217; ANDS0751; ANDS4788; ANDS4823; ANDU2244; ANDV4533; ANDW4372; ANDW4372; ANDW4372; ANDY2456; ANDY2456; ANEP0877; ANEQ4042; ANER1516; ANEX1707; ANDY3283; ANDS4788; ANEN2150; ANDX1373; ANEW4220; ANDS4788; ANEX1707; ANEQ4042; ANDV4533; ANDS4823; ANER1516; ANER1516; ANEX1707; ANDS4788; ANDU2244; ANEW4220; ANEW4220; ANCS3345; ANCS3345; ANDS4788; ANDX1373; ANDZ2287; ANDU2244; ANDU2244; ANEW4220; ANER1516; ANER1516; ANFN2784; ANER1516; ANFN2784; ANER1516; ANDW4372; ANDS4823; ANEP0877; ANDV4533; ANEN4880; ANEX1707; ANEN2150; ANER1516; ANCS3345; ANDN0809; ANEN4880; ANDY3283; ANEN4880; ANEN4880; ANEW4220; ANEW4220; ANEN2150; ANDS4823; ANEP0877; ANEQ4042; ANEX1707; ANDS0751; ANDU2244; ANER1516; ANDS4823; ANDU2244; ANDS4788; ANDZ2287; ANEP0877; ANDY4965; ANDZ2287; ANER1516; ANEW4220; ANEX1707; ANEX1707; ANEX1707; ANCQ1430; ANER1516; ANFN1434; ANEN3387; ANEN1072; ANEP2430; ANBZ1288; ANDX5376; ANEP2430; ANCS3433; ANEN1072; ANER1517; ANDS1282; ANDS1282; ANDU2245; ANEP2430; ANDW2002; ANDW4389; ANDX5376; ANDY3288; ANCS3433; ANDU2245; ANDU2245; ANDU2245; ANDW2002; ANDY3288; ANDY3288; ANEN1072; ANEP2430; ANEQ4043; ANER1517; ANDS1282; ANEQ4043; ANEQ4043; ANER1517; ANDU2245; ANDU2245; ANDW2002; ANDW2002; ANEN1072; ANEN3387; ANBZ1288; ANDS1282; ANDW4389; ANDS1282; ANDS1282; ANEP2430; ANDX5376; ANER1517; ANDS1282; ANEP2430; ANEN3387; ANEQ4043; ANDX5376; ANEN1072; ANEN3387; ANDU2245; ANEN3387; ANDX5376; ANEN1072; ANEN3387; ANFN3952; ANEN1072; ANDY3287; ANDX5375; ANDY3287; ANEP2412; ANDV4536; ANDY3287; ANEQ4948; ANES4065; ANDS4776; ANDS4776; ANDV4536; ANDX5375; ANDY3287; ANEP2412; ANEP2412; ANEP2412; ANEQ4948; ANDY3287; ANEP2412; ANDV4536; ANDY3287; ANDX1403; ANDX5375; ANCY3335; ANDV4536; ANDX1403; ANDX1403; ANDX5375; ANDS4776; ANDY3287; ANEQ4948; ANES4065; ANEQ4948; ANES4065; ANDS4776; ANDT1686; ANEZ3427; ANEZ3427; ANCY3335; ANDS4776; ANDT1686; ANEQ4948; ANDX1403; ANCU1841; ANEQ4948; ANEQ4948; ANEP2412; ANDY3287; ANES4065; ANES4065; ANEZ3427; ANEZ3427;

Devices named in this recall

28 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 87208).