Skip to content

Venovo® VENUM12060Recalls & reported problems

Angiomed GmbH & Co. Medizintechnik KG · Venovo® Venous Stent System 12 mm x 60 mm (80 cm delivery catheter) · Class 3 · Implant

All models of Venovo (84)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741103285), received between Aug 6, 2024 and Aug 6, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 14, 2020Class IIOpen, Classified

    Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.

    Z-1898-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2024Failure to Advance; Detachment of Device or Device Component; Failure to Advance; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE PLASTIC PART OVER SHAFT OF THE DEVICE WAS ALLEGEDLY LOOSENED. IT WAS FURTHER REPORTED THAT THERE WAS ALLEGEDLY LIMITED ADVANCEMENT OF DEVICE AND UNABLE TO DEPLOY THE STENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19920915 ↗

Other Stent, iliac vein devices

All Stent, iliac vein devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page