Medtronic LINQ II Model Number # LNQ22 Insertable Cardiac… recallZ-2076-2021 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated May 14, 2021

Class II Open, Classified All recalls posted in July 2021

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device fail to detect Brady episodes, Pause (Asystole) episodes and PVC's after a ride-through POR reset.

FDA root cause category: Software design

What the company is doing

On June 1 2021 Medtronic sent a letter notifying the consignees the following: -LINQ II insertable cardiac monitors (ICM) that undergo a partial electrical reset appear to be programmed ON, but are no longer able to detect and report Brady, Pause, and PVC events to clinicians. Medtronic estimates that 0.21% of LINQ II ICMs have experienced a partial electrical reset resulting in the inability to detect Brady, Pause, and PVC events. -A correction for currently implanted LINQ II ICMs is not available. -We are requesting that hospitals quarantine all LINQ II ICMs on hospital shelves. Physicians should cease implanting any remaining LINQ II ICMs that may remain in shelf stock and return any unused product to Medtronic. -There will be an update for future manufactured LINQ II ICMs, which is anticipated to be available in the U.S. July 2021. UPDATE: On June 16 2021, the firm sent an updated letter to inform their consignees of the availability of LINQ II Insertable Cardiac Monitors (ICMs) with an FDA cleared update addressing the issue referenced in the June 2021 letter. Updated LINQ II ICMs will not be susceptible to the potential for disabled Brady, Pause, and PVC detection after a partial electrical reset. Note that: - Initial supply of the updated LINQ II ICMs may be limited. - LINQ II ICMs manufactured prior to the approval of this update cannot be corrected in the field. ---- LINQ II ICMs implanted prior to the release of this update will continue to be susceptible to this issue. - Please continue to return all unused, affected LINQ II ICMs in your inventory to Medtronic. Contact Customer Service at 1-800-848-9300, or your local Medtronic Representative to assist you with a product return. UPDATE: On 08/02/2023, Medtronic sent another notification advising consignees of the availability of a software update for identified LINQ II ICMs.

Recall details

Recall number
Z-2076-2021
Recall class
Class II
Status
Open, Classified
Initiated by the company
May 14, 2021
Posted by the FDA
Jul 14, 2021
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
21,685
Product code
MXD
FDA submissions
K200795
Distribution
Worldwide distribution - US Nationwide in the states of AK, AL, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY. The countries of Australia, Austria, Canada, Denmark, Finland, Germany, Ireland, Italy, Netherlands, Netherlands Antilles, Sweden, Switzerland, and United Kingdom.

Product description

Medtronic LINQ II Model Number # LNQ22 Insertable Cardiac Monitor Sterile EO Rx Only

Code information (models, lots, serial numbers)

Model Number: LNQ22

Other devices under the same FDA submission

1 device cleared under K200795 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.