Medical device recalls: July 2021
The FDA posted 62 medical device recall records in July 2021, from 12 recall events by 8 companies. 15 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 3 events, 16 records
- Boston Scientific Corporation · 2 events, 12 records, 6 Class I
- Philips Respironics, Inc. · 2 events, 6 records, 6 Class I
- Exactech, Inc. · 1 event, 22 records
- Magellan Diagnostics, Inc. · 1 event, 3 records, 3 Class I
- Implantcast GMBH · 1 event, 1 record
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · 1 event, 1 record
- Vivorte, Inc. · 1 event, 1 record
FDA root cause
- Unknown/Undetermined by firm · 22
- Software design · 15
- Device Design · 13
- Nonconforming Material/Component · 9
- Software Manufacturing/Software Deployment · 1
- Radiation Control for Health and Safety Act · 1
- Labeling Change Control · 1
Class I recalls (15)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1956-2021 · Philips Respironics, Inc. · posted Jul 15, 2021 · Open, Classified
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
- Z-1957-2021 · Philips Respironics, Inc. · posted Jul 15, 2021 · Open, Classified
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
- Z-1958-2021 · Philips Respironics, Inc. · posted Jul 15, 2021 · Open, Classified
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
- Z-1972-2021 · Philips Respironics, Inc. · posted Jul 13, 2021 · Open, Classified
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the device s the air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
- Z-1973-2021 · Philips Respironics, Inc. · posted Jul 13, 2021 · Open, Classified
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the device s the air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
- Z-1974-2021 · Philips Respironics, Inc. · posted Jul 13, 2021 · Open, Classified
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the device s the air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
- Z-2015-2021 · Boston Scientific Corporation · posted Jul 13, 2021 · Open, Classified
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal…
- Z-2016-2021 · Boston Scientific Corporation · posted Jul 13, 2021 · Open, Classified
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal…
- Z-2017-2021 · Boston Scientific Corporation · posted Jul 13, 2021 · Open, Classified
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal…
- INVIVE CRT-P, Model Numbers:Class IZ-2018-2021 · Boston Scientific Corporation · posted Jul 13, 2021 · Open, Classified
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal…
- INTUA CRT-P, Model Numbers:Class IZ-2019-2021 · Boston Scientific Corporation · posted Jul 13, 2021 · Open, Classified
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal…
- Z-2020-2021 · Boston Scientific Corporation · posted Jul 13, 2021 · Open, Classified
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal…
- Z-1952-2021 · Magellan Diagnostics, Inc. · posted Jul 1, 2021 · Terminated
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
- Z-1953-2021 · Magellan Diagnostics, Inc. · posted Jul 1, 2021 · Terminated
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
- Z-1954-2021 · Magellan Diagnostics, Inc. · posted Jul 1, 2021 · Terminated
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
Class II recalls (47)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2139-2021 · GE Healthcare, LLC · posted Jul 23, 2021 · Open, Classified
Masimo SpO2 Saturation Values can become frozen after an extended length of use without a power down.
- Z-2112-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2113-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2114-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2115-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2116-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2117-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2118-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2119-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2120-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2121-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2122-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2123-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2124-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2125-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2126-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2127-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2128-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2129-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2130-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2131-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2132-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2133-2021 · Exactech, Inc. · posted Jul 22, 2021 · Open, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Z-2076-2021 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Jul 14, 2021 · Open, Classified
Device fail to detect Brady episodes, Pause (Asystole) episodes and PVC's after a ride-through POR reset.
- Z-2052-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2053-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2054-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2055-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2056-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2057-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2058-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2059-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2060-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2061-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2062-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2063-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2064-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- Z-2065-2021 · GE Healthcare, LLC · posted Jul 12, 2021 · Open, Classified
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation…
- EcoFit cup cementless:Class IIZ-2050-2021 · Implantcast GMBH · posted Jul 9, 2021 · Terminated
A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique…
- Revolution CTClass IIZ-2004-2021 · GE Healthcare, LLC · posted Jul 8, 2021 · Completed
GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its…
- Z-2024-2021 · Vivorte, Inc. · posted Jul 7, 2021 · Terminated
incorrect expiration date- product is listed with 02-15-2019 the correct kit expiration date is 12-18-2018
- Essentio pacemaker, Models:Class IIZ-1998-2021 · Boston Scientific Corporation · posted Jul 2, 2021 · Open, Classified
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery…
- Proponent pacemaker, Models:Class IIZ-1999-2021 · Boston Scientific Corporation · posted Jul 2, 2021 · Open, Classified
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery…
- Accolade pacemaker, Models:Class IIZ-2000-2021 · Boston Scientific Corporation · posted Jul 2, 2021 · Open, Classified
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery…
- Altrua 2 pacemaker, Models: ↗Class IIZ-2001-2021 · Boston Scientific Corporation · posted Jul 2, 2021 · Open, Classified
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery…
- Z-2002-2021 · Boston Scientific Corporation · posted Jul 2, 2021 · Open, Classified
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery…
- Z-2003-2021 · Boston Scientific Corporation · posted Jul 2, 2021 · Open, Classified
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.