CARESCAPE PDM-Masimo SpO2. Physiological data monitor recallZ-2139-2021 · GE Healthcare, LLC · initiated May 21, 2021
Class II Open, Classified All recalls posted in July 2021
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Masimo SpO2 Saturation Values can become frozen after an extended length of use without a power down.
FDA root cause category: Software Manufacturing/Software Deployment
What the company is doing
On May 21st the firm sent a letter to its customers with the following instructions: You can continue to use the CARESCAPE PDM with Masimo SET technology by completing the power down instructions below annually. 1. If the PDM is on a patient, provide alternate monitoring if needed, during the power down process. 2. Disconnect the PDM from the display device by removing the PDM from the mounting dock or by disconnecting the communication cable from the back of the PDM. 3. Open the battery door compartment and remove the battery (if present), for approximately 10 seconds. 4. Reinsert the battery and reconnect the PDM to the display device
Recall details
- Recall number
- Z-2139-2021
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- May 21, 2021
- Posted by the FDA
- Jul 23, 2021
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 111,728 units
- Product code
- MHX
- FDA submissions
- K071073
- Distribution
- worldwide distribution US nationwide, Albania, Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Fiji, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, State of Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Product description
CARESCAPE PDM-Masimo SpO2. Physiological data monitor.
Code information (models, lots, serial numbers)
CARESCAPE PDM: product codes SA3, SPX; model numbers 2042084-001, 2094504-001, 2016793-002, 2016793-003; Software kits, field replaceable units, and upgrade kits: part numbers 2034826-001, 2034826-002, 2034826-003, 2034826-004, 2034826-005, 2034826-006, 2034826-007, 2034826-008, 2034826-009, 2034826-010, 2034826-011, 2034826-012, 2034826-013, 2031069-010, 2045825-001, 2045825-002, 2045825-004, 2045825-005
Other devices under the same FDA submission
1 device cleared under K071073 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- N/A Patient Data ModuleGE Medical Systems Information Technologies, Inc. · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page