Novation recallZ-2125-2021 · Exactech, Inc. · initiated Jun 29, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
The firm issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 6/28/2021 via email on 6/29/2021 that were addressed to the attention of Exactech Agents, Representatives, and Distributors in possession of affected products.
Recall details
- Recall number
- Z-2125-2021
- Status
- Open, Classified
- Initiated by the company
- Jun 29, 2021
- Posted by the FDA
- Jul 22, 2021
- Recalling firm
- Exactech, Inc.
- Quantity
- 89,050 total
- Product code
- LPH
- FDA submissions
- K100269
- Distribution
- Distribution US nationwide.
Product description
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-22-XX, Novation GXL Liner, +5mm Lateralized, G00, 22mm ID. Orthopedic hip implant component.
Code information (models, lots, serial numbers)
All serial numbers. UDI-DI number: 10885862024190
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Novation 136-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
Product lines
Other devices under the same FDA submission
12 devices cleared under K100269 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Novation 130-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls19 reports
- Novation 132-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls2 reports
- Novation 130-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls0 reports
- Novation 132-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls0 reports
- NA 181-02-01Exactech, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate1 recall1 report
- Novation 136-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
- Novation 138-22-70Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
- Novation 138-28-50Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 recall0 reports
- NA 121-01-02Exactech, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented4 reports
- Novation 180-01-44Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 reports
- NA 181-02-00Exactech, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- Novation 180-01-46Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 88126).
- Z-2112-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-28-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip…
- Z-2113-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-32-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip…
- Z-2114-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-36-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip…
- Z-2115-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-22-XX, Novation GXL Neutral Liner, G00, 22mm ID. Orthopedic hip implant co…
- Z-2116-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-28-XX, Acumatch GXL 0 Degree Liner, 28mm; Novation GXL Liner, Neutral, 28m…
- Z-2117-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral, …
- Z-2118-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral, …
- Z-2119-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-40-XX, Novation GXL Liner, Neutral, 40mm. Orthopedic hip implant component…
- Z-2120-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-22-XX, Novation GXL Liner, G00, 22mm ID; and Novation GXL Liner, Lipped Ant…
- Z-2121-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-28-XX, Acumatch GXL 15 Degree Liner, 28mm; Novation GXL Liner, G0, 28mm ID;…
- Z-2122-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-36-XX, Acumatch GXL 15 Degree Liner, 36mm, and Novation GXL Liner, Lipped A…
- Z-2123-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40…
- Z-2124-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 134-28-XX, Acumatch GXL, Ext Cov Liner, 28mm. Orthopedic hip implant component…
- Z-2126-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-28-XX, Novation GXL Liner, +5mm Lateralized, 28mm; and Novation GXL Liner, …
- Z-2127-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-32-XX, Novation GXL Liner, +5mm Lateralized, 32mm. Orthopedic hip implant …
- Z-2128-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-36-XX, Novation GXL Liner, +5mm Lateralized, 36mm. Orthopedic hip implant …
- Z-2129-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-40-XX, Novation GXL Liner, +5mm Lateralized, 40mm. Orthopedic hip implant …
- Z-2130-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-22-XX, Novation GXL 10 deg Liner, G00, 22mm ID. Orthopedic hip implant com…
- Z-2131-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Line…
- Z-2132-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL a…
- Z-2133-2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page