Sharesource Connectivity Platform recallZ-2173-2021 · Baxter Healthcare Corporation · initiated Jun 30, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm notified its consignees by letter on 06/30/2021. The letter explained the issue with the instructions for use, specifically, That the AMIA Sharesource User Guide describes the function of the UF (ultrafiltration) Limit setting as the ending criteria of the cycler's Extra Last Drain Mode. Baxter will be sending a Follow-Up communication once the AMIA Sharesource User Guide has been updated on the portal.

Recall details

Recall number
Z-2173-2021
Status
Terminated
Initiated by the company
Jun 30, 2021
Recalling firm
Baxter Healthcare Corporation
Quantity
1 unit
Product code
FKX
FDA submissions
K151525
Distribution
Worldwide distribution - US Nationwide distribution and the country of Japan.

Product description

Sharesource Connectivity Platform for Use with the Amia Automated PD System

Code information (models, lots, serial numbers)

Product Code 5C9400, UDI: 45413765565692, 55413765565699

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K151525 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 88219).