RayStation (Radiation Treatment Planning System) recallZ-2138-2021 · Raysearch Laboratories AB · initiated Jun 8, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.
FDA root cause category: Under Investigation by firm
What the company is doing
RaySearch issued Urgent Field Safety Notice, Medical Device Correction #80230 via email on 06/08/21. Letter states reason for recall, health risk and action to take: TO BE TAKEN BY THE USER " Do not create setup beams with the localization point as isocenter (option 2). " Always perform secondary verification of new treatment positions. RayTreat will notify the user when the treatment positions have been updated. " Always perform imaging before treating the patient. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. SOLUTION This issue will be resolved in the next version of RayStation, scheduled for market release in June 2021 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact [email protected].
Recall details
- Recall number
- Z-2138-2021
- Status
- Terminated
- Initiated by the company
- Jun 8, 2021
- Recalling firm
- Raysearch Laboratories AB
- Quantity
- 5 units
- Product code
- MUJ
- FDA submissions
- K200487, K200569
- Distribution
- US distribution to TN; and Belgium
Product description
RayStation (Radiation Treatment Planning System) : RayStation 9A, 9B, 9B SP1, 10A, 10A SP1, 10B and 11A, in combination with RayCare Model Numbers/ UDI: RayStation 9.0 (9.0.0.113) 07350002010174; RayStation 9.1 (9.1.0.933) 07350002010266; RayStation 9.1 Service Pack 1 (9.2.0.483) 07350002010297; RayStation 10.0 (10.0.0.1154) 07350002010303; RayStation 10.0 Service Pack 1 (10.0.1.52) 07350002010365; RayStation 10.1 (10.1.0.613) 07350002010310; RayStation 11.0 (11.0.0.951) 07350002010389;
Code information (models, lots, serial numbers)
Product name (build number) UDI-DI RayStation 9.0 (9.0.0.113) 07350002010174 RayStation 9.1 (9.1.0.933) 07350002010266 RayStation 9.1 Service Pack 1 (9.2.0.483) 07350002010297 RayStation 10.0 (10.0.0.1154) 07350002010303 RayStation 10.0 Service Pack 1 (10.0.1.52) 07350002010365 RayStation 10.1 (10.1.0.613) 07350002010310 RayStation 11.0 (11.0.0.951) 07350002010389
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- RayStation 9Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment12 recalls0 reports
- RayStation 9B SP1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment11 recalls0 reports
- RayStation 10.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment10 recalls1 report
- RayStation 10.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 10.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 11.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment8 recalls0 reports
Other devices under the same FDA submission
1 device cleared under K200487, K200569 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RayCare 4.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page