Skip to content

RayStation 10.0Recalls & reported problems

Raysearch Laboratories AB (Publ) · Class 2

5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
9

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

9 FDA recall records name this device's identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 16, 2025Open, Classified

    Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

    Z-1187-2025 · Root cause: Under Investigation by firm · FDA record ↗

  2. Mar 28, 2024Open, Classified

    Potential for reported SSD to be too high.

    Z-1746-2024 · Root cause: Software design · FDA record ↗

  3. Feb 23, 2024Open, Classified

    A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.

    Z-1482-2024 · Root cause: Under Investigation by firm · FDA record ↗

  4. Feb 11, 2022Open, Classified

    Incorrect source to skin or surface distance (SSD) calculation may result in dose error when using RayStation system.

    Z-0850-2022 · Root cause: Under Investigation by firm · FDA record ↗

  5. Sep 22, 2021Open, Classified

    An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be both under- and overestimated

    Z-0849-2022 · Root cause: Device Design · FDA record ↗

  6. Sep 9, 2021Terminated

    If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning

    Z-0848-2022 · Root cause: Under Investigation by firm · FDA record ↗

  7. Jun 8, 2021Terminated

    Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.

    Z-2138-2021 · Root cause: Under Investigation by firm · FDA record ↗

  8. Feb 22, 2021Terminated

    For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. In some cases, when merging two beams where the first beam has clinical dose and the second beam has approximate dose, the dose of the resulting beam will be labeled as Clinical, although dose for some of the beam segments is still calculated with the SVD dose engine and it should be labeled Approximate: Mixed dose. The difference between approximate and final Clinical dose is in most cases small, but there can be body sites such as lung where the difference can be significant.

    Z-1597-2021 · Root cause: Software design · FDA record ↗

  9. Jan 22, 2021Terminated

    A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an incorrectly created ROI are used to determine if a plan is acceptable for treatment, this could potentially lead to an inappropriate plan being approved. The error could lead to a more conservative or more liberal plan than intended.

    Z-2039-2023 · Root cause: Software design · FDA record ↗

Other System, Planning, Radiation Therapy Treatment devices

All System, Planning, Radiation Therapy Treatment devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.