Arcos Modular Revision Hip System recallZ-2273-2021 · Biomet, Inc. · initiated Jun 2, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention
FDA root cause category: Process control
What the company is doing
Zimmer Biomet issued Urgent Medical Device Notification Letter on 6/2/21 to Distributors/ Risk Managers/Surgeons, stated reason for recall, health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please ca
Recall details
- Recall number
- Z-2273-2021
- Status
- Terminated
- Initiated by the company
- Jun 2, 2021
- Recalling firm
- Biomet, Inc.
- Quantity
- 12 units
- Product code
- KWA
- FDA submissions
- K090757
- Distribution
- US Nationwide distribution.
Product description
Arcos Modular Revision Hip System Standard Cone Prox Body, Size B, 80 MM Item Number: 11-301342
Code information (models, lots, serial numbers)
Lot Numbers/UDI : 391390 (01)00880304474208(17)310323(10)391390
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Arcos Modular Revision Hip System 11-301342Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish1 recall4 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K090757 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARCOS MODULAR REVISION SYSTEM 31-301852Biomet Orthopedics, LLC · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer46 reports
- Arcos® Modular Revision System 31-301852Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented11 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88304).
- Z-2259-2021 · TerminatedArcos Modular Revision Hip System Standard Cone Prox Body, Size A, 50 MM Item Number: 11-301300
- Z-2260-2021 · TerminatedArcos Modular Revision Hip System Standard Cone Prox Body, Size A, 60 MM Item Number: 11-301301
- Z-2261-2021 · TerminatedArcos Modular Revision Hip System Standard Cone Prox Body, Size B, 60 MM Item Number: 11-301302
- Z-2262-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size A, 50 MM Item Number: 11-301310
- Z-2263-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size A, 60 MM Item Number: 11-301311
- Z-2264-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size B, 60 MM Item Number: 11-301312
- Z-2265-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size D, 60 MM Item Number: 11-301314
- Z-2266-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size E, 60 MM Item Number: 11-301315
- Z-2267-2021 · TerminatedArcos Modular Revision Hip System Standard Cone Prox Body, Size A, 70 MM Item Number: 11-301321
- Z-2268-2021 · TerminatedArcos Modular Revision Hip System Standard Cone Prox Body, Size B, 70 MM Item Number: 11-301322
- Z-2269-2021 · TerminatedArcos Modular Revision Hip System Standard Cone Prox Body, Size E, 70 MM Item Number:11-301325
- Z-2270-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size A, 70 MM Item Number: 11-301331
- Z-2271-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size C, 70 MM Item Number: 11-301333
- Z-2272-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size E, 70 MM Item Number: 11-301335
- Z-2274-2021 · TerminatedArcos Modular Revision Hip System Standard Cone Prox Body, Size C, 80 MM Item Number: 11-301343
- Z-2275-2021 · TerminatedArcos Modular Revision Hip System High Offset Cone Prox Body, Size D, 80 MM Item Number: 11-301354
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page