Monaco recallZ-2406-2021 · Elekta Inc · initiated Jul 26, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.

FDA root cause category: Under Investigation by firm

What the company is doing

Elekta will send an Important Field Safety Notice 382-01-MON-019 to all affected customers beginning August 10, 2021. The notice informs users of the specific product and version numbers affected by the issue, and any workaround that can be used to avoid the issue. A copy of the Important Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. Customers are asked to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible, but no later than 30 days after receipt. This issue is expected to be addressed in a future Monaco release, tentatively scheduled for the end of June 2022.

Recall details

Recall number
Z-2406-2021
Status
Open, Classified
Initiated by the company
Jul 26, 2021
Recalling firm
Elekta Inc
Product code
MUJ
FDA submissions
K183037, K190178
Distribution
Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom

Product description

Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Code information (models, lots, serial numbers)

Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10 UDI: (01)00858164002190(10) 5.40.00 (01)00858164002190(10) 5.40.01 (01)00858164002190(10) 5.40.02 (01)00858164002268(10) 5.51.10