7 open recalls name at least one Monaco model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
45
Deaths reported
0
Injuries reported
1
Recalls
8

We found 5 models of Monaco by Elekta, Inc. in the FDA's device registry (GUDID). Together they are named in 45 adverse event reports in the FDA's MAUDE database, received between Dec 8, 2016 and Oct 9, 2024 — classified as 1 injury, 44 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (45 reports).

8 FDA recall records name at least one of these models by its device identifier, 7 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Monaco recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 28, 2023Open, Classified

    Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.

    Z-1233-2023 · Root cause: Software design · Recall details · FDA record ↗

  2. Sep 14, 2021Open, Classified

    Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

    Z-2563-2021 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  3. Jul 26, 2021Open, Classified

    If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.

    Z-2406-2021 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  4. Apr 16, 2020Open, Classified

    The Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose.

    Z-1825-2020 · Root cause: Software design · Recall details · FDA record ↗

  5. Nov 22, 2019Open, Classified

    Monaco is using the incorrect energy when optimizing and calculating dose.

    Z-0670-2020 · Root cause: Software Design Change · names 3 of these models · Recall details · FDA record ↗

  6. Sep 4, 2019Open, Classified

    It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

    Z-2561-2019 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  7. Sep 4, 2019Open, Classified

    It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

    Z-2560-2019 · Root cause: Software design · Recall details · FDA record ↗

  8. Aug 23, 2019Terminated

    Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.

    Z-2409-2019 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

Most reported Monaco problems

Device problems

  • Computer Software Problem45
  • Misassembled During Installation1

Patient problems

  • No Consequences Or Impact To Patient16
  • No Clinical Signs, Symptoms or Conditions12
  • No Patient Involvement10
  • No Information7
  • Insufficient Information6
  • Radiation Overdose4
  • Radiation Underdose1

Monaco models

5 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Monaco adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Elekta, Inc. products

All Elekta, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page