Ingenia 1.5T S-a Magnetic Resonance Medical Electrical… recallZ-2413-2021 · Philips North America Llc · initiated Aug 11, 2021
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
FDA root cause category: Device Design
What the company is doing
Philips notified U.S. customers an Urgent Medical Device Correction letter on 8/11/21. Letter states reason for recall, health risk and action to take: Philips advises that customers continue to utilize the primary motorized function for the MR system since the identified issue with the HDU does not occur in the motorized mode. If you utilize the manual function and it is not functioning properly, then you must perform table movement using the motorized function. A Philips Field Service Engineer (FSE) will perform a HDU during the next scheduled preventive maintenance visit. If the issue is detected in the HDU, the Philips FSE will replace the necessary mechanical component(s) to resolve the issue. If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377) and reference FCO78100527.
Recall details
- Recall number
- Z-2413-2021
- Status
- Completed
- Initiated by the company
- Aug 11, 2021
- Recalling firm
- Philips North America Llc
- Quantity
- 144 units
- Product code
- LNH
- FDA submissions
- K153324
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.
Product description
Ingenia 1.5T S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Model: 781347
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Model: 78137 Serial Number Unique Device Identifier (UDI#) 81045 N/A 81075 (01)00884838068421(21)81075 81076 (01)00884838068421(21)81076 81077 (01)00884838068421(21)81077 81096 (01)00884838068421(21)81096 81124 (01)00884838068421(21)81124 81114 (01)00884838068421(21)81114 81126 (01)00884838068421(21)81126 81125 (01)00884838068421(21)81125 81144 (01)00884838068421(21)81144 81145 (01)00884838068421(21)81145 81159 (01)00884838068421(21)81159 81003 N/A 81111 (01)00884838068421(21)81111 81050 N/A 81051 (01)00884838068421(21)81051 81090 (01)00884838068421(21)81090 81021 N/A 81038 N/A 81185 (01)00884838068421(21)81185 81179 (01)00884838068421(21)81179 81005 N/A 81006 N/A 81011 N/A 81009 N/A 81019 N/A 81020 N/A 81016 N/A 81023 N/A 81017 N/A 81024 N/A 81014 N/A 81018 N/A 81004 N/A 81041 N/A 81057 (01)00884838068421(21)81057 81055 (01)00884838068421(21)81055 81037 (01)00884838068421(21)81037 81049 (01)00884838068421(21)81049 81060 (01)00884838068421(21)81060 81056 (01)00884838068421(21)81056 81064 (01)00884838068421(21)81064 81062 (01)00884838068421(21)81062 81070 (01)00884838068421(21)81070 81078 (01)00884838068421(21)81078 81073 (01)00884838068421(21)81073 81074 (01)00884838068421(21)81074 81112 (01)00884838068421(21)81112 81115 (01)00884838068421(21)81115 81122 (01)00884838068421(21)81122 81147 (01)00884838068421(21)81147 81146 (01)00884838068421(21)81146 81153 (01)00884838068421(21)81153 81156 (01)00884838068421(21)81156 81158 (01)00884838068421(21)81158 81093 (01)00884838068421(21)81093 81027 N/A 81035 N/A 81010 (01)00884838068421(21)81010 81063 (01)00884838068421(21)81063 81028 (01)00884838068421(21)81028 81030 (01)00884838068421(21)81030 81106 (01)00884838068421(21)81106 81052 N/A 81067 (01)00884838068421(21)81067 81105 (01)00884838068421(21)81105 81116 (01)00884838068421(21)81116 81013 N/A 81007 N/A 81008 N/A 81040 N/A 81031 (01)00884838068421(21)81031 81152 (01)00884838068421(21)81152 81044 N/A 81167 (01)00884838068421(21)81167 81170 (01)00884838068421(21)81170 81171 (01)00884838068421(21)81171 81047 (01)00884838068421(21)81047 81059 (01)00884838068421(21)81059 81108 (01)00884838068421(21)81108 81148 (01)00884838068421(21)81148 81133 (01)00884838068421(21)81133 81001 N/A 81086 (01)00884838068421(21)81086 81118 (01)00884838068421(21)81118 81139 (01)00884838068421(21)81139 81140 (01)00884838068421(21)81140 81138 (01)00884838068421(21)81138 81151 (01)00884838068421(21)81151 81175 N/A 81184 N/A 81079 (01)00884838068421(21)81079 81069 (01)00884838068421(21)81069 81061 (01)00884838068421(21)81061 81065 (01)00884838068421(21)81065 81098 (01)00884838068421(21)81098 81085 (01)00884838068421(21)81085 81092 (01)00884838068421(21)81092 81091 (01)00884838068421(21)81091 81032 N/A 81043 (01)00884838068421(21)81043 81026 N/A 81129 (01)00884838068421(21)81129 81131 (01)00884838068421(21)81131 81149 (01)00884838068421(21)81149 81155 (01)00884838068421(21)81155 81162 (01)00884838068421(21)81162 81172 (01)00884838068421(21)81172 81165 (01)00884838068421(21)81165 81180 (01)00884838068421(21)81180 81084 (01)00884838068421(21)81084 81087 (01)00884838068421(21)81087 81088 (01)00884838068421(21)81088 81095 (01)00884838068421(21)81095 81094 (01)00884838068421(21)81094 81100 (01)00884838068421(21)81100 81127 (01)00884838068421(21)81127 81132 (01)00884838068421(21)81132 81174 (01)00884838068421(21)81174 81039 N/A 81071 (01)00884838068421(21)81071 81157 (01)00884838068421(21)81157 81176 (01)00884838068421(21)81176 81117 (01)00884838068421(21)81117 81173 (01)00884838068421(21)81173 81025 N/A 81097 (01)00884838068421(21)81097 81154 (01)00884838068421(21)81154 81015 N/A 81029 N/A 81046 N/A 81033 N/A 81034 N/A 81048 N/A 81042 N/A 81058 (01)00884838068421(21)81058 81080 (01)00884838068421(21)81080 81072 (01)00884838068421(21)81072 81103 (01)00884838068421(21)81103 81113 (01)00884838068421(21)81113 81123 (01)00884838068421(21)81123 81128 (01)00884838068421(21)81128 81137 (01)00884838068421(21)81137 81178 (01)00884838068421(21)81178
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia 1.5T S 781347Philips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging7 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 88491).
- Z-2411-2021 · CompletedIngenia 1.5T- a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or s…
- Z-2412-2021 · CompletedIngenia 1.5T Evolution-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images…
- Z-2414-2021 · CompletedIngenia 1.5T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or…
- Z-2415-2021 · CompletedSmartPath to dStream for 1.5T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic…
- Z-2416-2021 · CompletedIngenia Ambition S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and…
- Z-2417-2021 · CompletedIngenia 3.0T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or sp…
- Z-2418-2021 · CompletedIngenia Elition S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/…
- Z-2419-2021 · CompletedIngenia Elition X-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/…
- Z-2420-2021 · CompletedIngenia 3.0T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or…
- Z-2421-2021 · CompletedSmartPath to dStream for XR and 3.0T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectr…
- Z-2422-2021 · CompletedIngenia Ambition X- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.